Positional correction of a 3D position-sensitive virtual Frisch-grid CZT detector for gamma spectroscopy and imaging based on a theoretical assumption

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류원자력공학

초록

The virtual Frisch-grid method for room-temperature radiation detectors has been widely used because of its simplicity and high performance. Recently, side electrodes were separately attached to each surface of the detectors instead of covering the entire detector surface with a single electrode. The side-electrode structure enables the measurement of the three-dimensional (3D) gamma-ray interaction in the detector. The positional information of the interaction can then be utilized to precisely calibrate the response of the detector for gamma-ray spectroscopy and imaging. In this study, we developed a 3D position-sensitive 5  5  12 mm3 cadmium-zinc-telluride (CZT) detector and applied a flattening method to correct detector responses. Collimated gamma-rays incident on the surface of the detector were scanned to evaluate the positional accuracy of the detection system. Positional distributions of the radiation interactions with the detector were imaged for quantitative and qualitative evaluation. The energy spectra of various radioisotopes were measured and improved by the detector response calibration according to the calculated positional information. The energy spectra ranged from 59.5 keV (emitted by 241Am) to 1332 keV (emitted by 60Co). The best energy resolution was 1.06% at 662 keV when the CZT detector was voxelized to 20  20  10.

Abstract

The virtual Frisch-grid method for room-temperature radiation detectors has been widely used because of its simplicity and high performance. Recently, side electrodes were separately attached to each surface of the detectors instead of covering the entire detector surface with a single electrode. The side-electrode structure enables the measurement of the three-dimensional (3D) gamma-ray interaction in the detector. The positional information of the interaction can then be utilized to precisely calibrate the response of the detector for gamma-ray spectroscopy and imaging. In this study, we developed a 3D position-sensitive 5  5  12 mm3 cadmium-zinc-telluride (CZT) detector and applied a flattening method to correct detector responses. Collimated gamma-rays incident on the surface of the detector were scanned to evaluate the positional accuracy of the detection system. Positional distributions of the radiation interactions with the detector were imaged for quantitative and qualitative evaluation. The energy spectra of various radioisotopes were measured and improved by the detector response calibration according to the calculated positional information. The energy spectra ranged from 59.5 keV (emitted by 241Am) to 1332 keV (emitted by 60Co). The best energy resolution was 1.06% at 662 keV when the CZT detector was voxelized to 20  20  10.

참고문헌 (30)

  1. Stefano Del Sordo, "Progress in the Development of CdTe and CdZnTe Semiconductor Radiation Detectors for Astrophysical and Medical Applications", Sensors, 9(5), pp.3491~3526 (2009), MDPI AG[DOI]
  2. Qiushi Zhang, "Progress in the Development of CdZnTe Unipolar Detectors for Different Anode Geometries and Data Corrections", Sensors, 13(2), pp.2447~2474 (2013), MDPI AG[DOI]
  3. Z. He, "3-D position sensitive CdZnTe gamma-ray spectrometers", Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 422(1-3), pp.173~178 (1999), Elsevier BV[DOI]
  4. P.N. Luke, "Unipolar charge sensing with coplanar electrodes-application to semiconductor detectors", IEEE Transactions on Nuclear Science, 42(4), pp.207~213 (1995), Institute of Electrical and Electronics Engineers (IEEE)[DOI]
  5. Zhong He, "1-D position sensitive single carrier semiconductor detectors", Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 380(1-2), pp.228~231 (1996), Elsevier BV[DOI]
  6. D. S. McGregor, "Single charge carrier type sensing with a parallel strip pseudo-Frisch-grid CdZnTe semiconductor radiation detector", Applied Physics Letters, 72(7), pp.792~794 (1998), AIP Publishing[DOI]
  7. V. Ivanov, "Development of Large Volume Hemispheric CdZnTe Detectors for Use in Safeguards Applications", Proceedings of ESARDA 19th Annual Symposium on Safeguards and Nuclear Material Management, pp.447~453 (1997)
  8. K Parnham, "Performance improvement of CdZnTe detectors using modified two-terminal electrode geometry", Journal of Crystal Growth, 214-215, pp.1152~1154 (2000), Elsevier BV[DOI]
  9. G. Montemont, "A capacitive Frisch grid structure for CdZnTe detectors", IEEE Transactions on Nuclear Science, 48(3), pp.278~281 (2001), Institute of Electrical and Electronics Engineers (IEEE)[DOI]
  10. W. J. McNeil, "Single-charge-carrier-type sensing with an insulated Frisch ring CdZnTe semiconductor radiation detector", Applied Physics Letters, 84(11), pp.1988~1990 (2004), AIP Publishing[DOI]
  11. A.E. Bolotnikov, "Performance characteristics of Frisch-ring CdZnTe detectors", IEEE Transactions on Nuclear Science, 53(2), pp.607~614 (2006), Institute of Electrical and Electronics Engineers (IEEE)[DOI]
  12. A. E. Bolotnikov, "Array of Virtual Frisch-Grid CZT Detectors With Common Cathode Readout for Correcting Charge Signals and Rejection of Incomplete Charge-Collection Events", IEEE Transactions on Nuclear Science, 59(4), pp.1544~1551 (2012), Institute of Electrical and Electronics Engineers (IEEE)[DOI]
  13. A. E. Bolotnikov, "An array of virtual Frisch-grid CdZnTe detectors and a front-end application-specific integrated circuit for large-area position-sensitive gamma-ray cameras", Review of Scientific Instruments, 86(7), pp.073114~ (2015), AIP Publishing[DOI]
  14. Wonho Lee, "Mini Compton Camera Based on an Array of Virtual Frisch-Grid CdZnTe Detectors", IEEE Transactions on Nuclear Science, 63(1), pp.259~265 (2016), Institute of Electrical and Electronics Engineers (IEEE)[DOI]
  15. 김영학, "Development of a Virtual Frisch-Grid CZT Detector Based on the Array Structure", 방사선방어학회지, 45(1), pp.35~44 (2020), 테스트학회
  16. A. E. Bolotnikov, "Use of high-granularity position sensing to correct response non-uniformities of CdZnTe detectors", Applied Physics Letters, 104(26), pp.263503-1~263503-5 (2014), AIP Publishing[DOI]
  17. Zhong He, "Review of the Shockley–Ramo theorem and its application in semiconductor gamma-ray detectors", Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 463(1-2), pp.250~267 (2001), Elsevier BV[DOI]
  18. J.K. Polack, "Variation of electric shielding on virtual Frisch-grid detectors", Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 621(1-3), pp.424~430 (2010), Elsevier BV[DOI]
  19. A.E. Bolotnikov, "Rejecting incomplete charge-collection events in CdZnTe and other semiconductor detectors", Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 664(1), pp.317~323 (2012), Elsevier BV[DOI]
  20. L. Ocampo Giraldo, "A linear array of position-sensitive virtual Frisch-grid CdZnTe for low-energy gamma rays", Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 903, pp.204~214 (2018), Elsevier BV[DOI]
  21. Aleksey E. Bolotnikov, "Positionsensitive Frisch-grid CdZnTe detectors for gamma-ray spectroscopy and imaging", Proc. SPIE, 11114, pp.111140S~ (2019)[DOI]
  22. I. Kuvvetli, "A 3D CZT high resolution detector for x- and gamma-ray astronomy", Proc. SPIE, 9154, pp.91540X~ (2014), SPIE[DOI]
  23. Carl Budtz-Jorgensen, "New Position Algorithms for the 3D CZT Drift Detector", IEEE Transactions on Nuclear Science, 64, pp.1611~1618 (2017), Institute of Electrical and Electronics Engineers (IEEE)[DOI]
  24. 김영학, "Radiation Measurement and Imaging using 3D Position Sensitive Pixelated CZT Detector", Nuclear Engineering and Technology, 51(5), pp.1417~1427 (2019), 한국원자력학회[DOI]
  25. "110-R2 Charge Sensitive Preamplifier" (2018)[URL]
  26. "X Gaussian Shaping Amplifier Instrument" (2018)[URL]
  27. A.E. Bolotnikov, "A 4X4Array module of position-sensitive virtual Frisch-grid CdZnTe detectors for gamma-ray imaging spectrometers", Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 954, pp.161036~ (2020), Elsevier BV[DOI]
  28. A.E. Bolotnikov, "Performance of 8x8x32 and 10x10x32 mm3 CdZnTe position-sensitive virtual Frisch-grid detectors for high-energy gamma ray cameras", Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 969, pp.164005~ (2020), Elsevier BV[DOI]
  29. E Kalemci, "Investigation of charge sharing among electrode strips for a CdZnTe detector", Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 478(3), pp.527~537 (2002), Elsevier BV[DOI]
  30. A. E. Bolotnikov, "Optimization of virtual Frisch-grid CdZnTe detector designs for imaging and spectroscopy of gamma rays", SPIE Proceedings, 6706, pp.670603~ (2007), SPIE[DOI]

관련 논문AI추천

이 간행물의 다른 논문 (41건)