A polystyrene phantom was developed following the guidance of the International Atomic Energy Association (IAEA) for gamma knife (GK) quality assurance. Its performance was assessed by measuring the absorbed dose rate to water and dose distributions. The phantom was made of polystyrene, which has an electron density (1.0156) similar to that of water. The phantom included one outer phantom and four inner phantoms. Two inner phantoms held PTW T31010 and Exradin A16 ion chambers. One inner phantom held a film in the XY plane of the Leksell coordinate system, and another inner phantom held a film in the YZ or ZX planes. The absorbed dose rate to water and beam profiles of the machine-specific reference (msr) field, namely, the 16 mm collimator field of a GK Perfexion™ or Icon™, were measured at seven GK sites. The measured results were compared to those of an IAEA-recommended solid water (SW) phantom. The radius of the polystyrene phantom was determined to be 7.88 cm by converting the electron density of the plastic, considering a water depth of 8 g/cm2 . The absorbed dose rates to water measured in both phantoms differed from the treatment planning program by less than 1.1%. Before msr correction, the PTW T31010 dose rates (PTW Freiberg GmbH, New York, NY, USA) in the polystyrene phantom were 0.70 (0.29)% higher on average than those in the SW phantom. The Exradin A16 (Standard Imaging, Middleton, WI, USA) dose rates were 0.76 (0.32)% higher in the polystyrene phantom. After msr correction factors were applied, there were no statistically significant differences in the A16 dose rates measured in the two phantoms; however, the T31010 dose rates were 0.72 (0.29)% higher in the polystyrene phantom. When the full widths at half maximum and penumbras of the msr field were compared, no significant differences between the two phantoms were observed, except for the penumbra in the Y-axis. However, the difference in the penumbra was smaller than variations among different sites. A polystyrene phantom developed for gamma knife dosimetry showed dosimetric performance comparable to that of a commercial SW phantom. In addition to its cost effectiveness, the polystyrene phantom removes air space around the detector. Additional simulations of the msr correction factors of the polystyrene phantom should be performed.
Abstract
A polystyrene phantom was developed following the guidance of the International Atomic Energy Association (IAEA) for gamma knife (GK) quality assurance. Its performance was assessed by measuring the absorbed dose rate to water and dose distributions. The phantom was made of polystyrene, which has an electron density (1.0156) similar to that of water. The phantom included one outer phantom and four inner phantoms. Two inner phantoms held PTW T31010 and Exradin A16 ion chambers. One inner phantom held a film in the XY plane of the Leksell coordinate system, and another inner phantom held a film in the YZ or ZX planes. The absorbed dose rate to water and beam profiles of the machine-specific reference (msr) field, namely, the 16 mm collimator field of a GK Perfexion™ or Icon™, were measured at seven GK sites. The measured results were compared to those of an IAEA-recommended solid water (SW) phantom. The radius of the polystyrene phantom was determined to be 7.88 cm by converting the electron density of the plastic, considering a water depth of 8 g/cm2 . The absorbed dose rates to water measured in both phantoms differed from the treatment planning program by less than 1.1%. Before msr correction, the PTW T31010 dose rates (PTW Freiberg GmbH, New York, NY, USA) in the polystyrene phantom were 0.70 (0.29)% higher on average than those in the SW phantom. The Exradin A16 (Standard Imaging, Middleton, WI, USA) dose rates were 0.76 (0.32)% higher in the polystyrene phantom. After msr correction factors were applied, there were no statistically significant differences in the A16 dose rates measured in the two phantoms; however, the T31010 dose rates were 0.72 (0.29)% higher in the polystyrene phantom. When the full widths at half maximum and penumbras of the msr field were compared, no significant differences between the two phantoms were observed, except for the penumbra in the Y-axis. However, the difference in the penumbra was smaller than variations among different sites. A polystyrene phantom developed for gamma knife dosimetry showed dosimetric performance comparable to that of a commercial SW phantom. In addition to its cost effectiveness, the polystyrene phantom removes air space around the detector. Additional simulations of the msr correction factors of the polystyrene phantom should be performed.
참고문헌 (16)
- Christer Lindquist, "THE LEKSELL GAMMA KNIFE PERFEXION AND COMPARISONS WITH ITS PREDECESSORS", Operative Neurosurgery, 61(3), pp.130~141 (2007), Ovid Technologies (Wolters Kluwer Health)[DOI]
- Michele Zeverino, "Commissioning of the Leksell gamma Knife® Icon", Medical Physics, 44(2), pp.355~363 (2017), Wiley[DOI]
- Hyun-Tai Chung, "Determination of the Absorbed Dose Rate to Water for the 18-mm Helmet of a Gamma Knife", International Journal of Radiation Oncology*Biology*Physics, 79(5), pp.1580~1587 (2011), Elsevier BV[DOI]
- Hugo Palmans, "Dosimetry of small static fields used in external photon beam radiotherapy: Summary of TRS‐483, the IAEA–AAPM international Code of Practice for reference and relative dose determination", Medical Physics, 45(11), pp.e1123~ (2018), Wiley[DOI]
- Paula L. Petti, "Recommendations on the practice of calibration, dosimetry, and quality assurance for gamma stereotactic radiosurgery: Report of AAPM Task Group 178", Medical Physics, 48(7), pp.e733~e770 (2021), Wiley[DOI]
- Jagdish P. Bhatnagar, "Unintended attenuation in the Leksell Gamma Knife® Perfexion™ calibration‐phantom adaptor and its effect on dose calibration", Medical Physics, 36(4), pp.1208~1211 (2009), Wiley[DOI]
- Jeong-Hoon Park, "Application of the gamma evaluation method in Gamma Knife film dosimetry", Medical Physics, 38(10), pp.5778~5787 (2011), Wiley[DOI]
- V. Sarkar, "Head to head comparison of two commercial phantoms used for SRS QA", J. Radiosurg. SBRT, 4(3), pp.213~223 (2016)
- Jae Pil Chung, "Development of a PMMA phantom as a practical alternative for quality control of gamma knife® dosimetry", Radiation Oncology, 13(1), pp.176~ (2018), Springer Science and Business Media LLC[DOI]
- J. Seco, "Assessing the effect of electron density in photon dose calculations", Medical Physics, 33(2), pp.540~552 (2006), Wiley[DOI]
- Yona Choi, "Consistency of dose rates after applying machine‐specific reference correction factors for the gamma knife 16 mm collimator field", Medical Physics, 50(6), pp.3816~3824 (2023), Wiley[DOI]
- Lalageh Mirzakhanian, "Determination of factors for ion chambers used in the calibration of Leksell Gamma Knife Perfexion model using EGSnrc and PENELOPE Monte Carlo codes", Medical Physics, 45(4), pp.1748~1757 (2018), Wiley[DOI]
- E Zoros, "Monte Carlo and experimental determination of correction factors for gamma knife perfexion small field dosimetry measurements", Physics in Medicine & Biology, 62(18), pp.7532~7555 (2017), IOP Publishing[DOI]
- Schaarschmidt Thomas, "GEANT4-based Monte Carlo Simulation of Beam Quality Correction Factors for the Leksell Gamma Knifer PerfexionTM", Journal of the Korean Physical Society, 73(12), pp.1814~1820 (2018), 한국물리학회
- T. Schaarschmidt, "GEANT4-based Monte Carlo Simulation of Beam Quality Correction Factors for the Leksell Gamma Knife® Perfexion™", Doctor, Philosophy, pp.2019
- Azam Niroomand‐Rad, "Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55", Medical Physics, 25(11), pp.2093~2115 (1998), Wiley[DOI]