Evaluating polyester resin as a viable substitute for PMMA in computed tomography dosimetry phantoms

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류원자력공학

초록

The current study aimed to evaluate the suitability of polyester resin as an alternative material to polymethyl methacrylate (PMMA) for computed tomography (CT) dosimetry phantoms using the GEANT4/GATE Monte Carlo simulation platform. Cylindrical phantoms (32 cm diameter) constructed of polyester resin and PMMA were simulated and compared in terms of atomic composition, effective atomic number, electron density, mass density, and photon interaction mechanisms. Weighted CT dose index (CTDIw) values were calculated for each phantom at 80, 110, and 130 kVp tube voltages based on measurements of CTDI100,c and CTDI100,p. Results demonstrated that the physical properties of polyester closely matched those of PMMA, and the polyester phantom displayed equivalent dosimetric behavior to the PMMA phantom at all tube voltages tested. CTDIw values from the polyester phantom were within 1.4 % of the PMMA phantom across all tube voltages. Conversion coefficients were derived to equate polyester CTDIw values to PMMA dose equivalents. This study found that a polyester resin phantom exhibited radiation dosimetry commensurate with the standard PMMA phantom for CT dose assessment. Consequently, polyester resin represents a viable substitute material when PMMA is unavailable for construction of CT dosimetry phantoms

Abstract

The current study aimed to evaluate the suitability of polyester resin as an alternative material to polymethyl methacrylate (PMMA) for computed tomography (CT) dosimetry phantoms using the GEANT4/GATE Monte Carlo simulation platform. Cylindrical phantoms (32 cm diameter) constructed of polyester resin and PMMA were simulated and compared in terms of atomic composition, effective atomic number, electron density, mass density, and photon interaction mechanisms. Weighted CT dose index (CTDIw) values were calculated for each phantom at 80, 110, and 130 kVp tube voltages based on measurements of CTDI100,c and CTDI100,p. Results demonstrated that the physical properties of polyester closely matched those of PMMA, and the polyester phantom displayed equivalent dosimetric behavior to the PMMA phantom at all tube voltages tested. CTDIw values from the polyester phantom were within 1.4 % of the PMMA phantom across all tube voltages. Conversion coefficients were derived to equate polyester CTDIw values to PMMA dose equivalents. This study found that a polyester resin phantom exhibited radiation dosimetry commensurate with the standard PMMA phantom for CT dose assessment. Consequently, polyester resin represents a viable substitute material when PMMA is unavailable for construction of CT dosimetry phantoms

참고문헌 (26)

  1. T. Kim, "Comparison of X-ray computed tomography and magnetic resonance imaging to detect pestinfested fruits : a pilot study", Nucl. Eng. Technol, 54, pp.514~522 (2022)[DOI]
  2. C. Lee, "Organ dose conversion coefficients in CT scans for Korean adult males and females", Nucl. Eng. Technol, 54, pp.681~688 (2022)[DOI]
  3. A. Albngali, "Dose equilibrium and CTDI quality assurance in CT and the implications for the effective dose estimation in patients", Phys. Med, 67, pp.207~ (2019)[DOI]
  4. F. Guerra Paiva, "Dosimetric study of pediatric PET", CT scan using different phantoms, 2020, pp.173, 108887~
  5. D. Pennington, "Pediatric Imaging for the Emergency Provider", pp.6~7 (2022), Elsevier
  6. L. N. Fatichah, "Identification of the computed tomography dose index for tube voltage variations in a polyester-resin phantom", Appl. Radiat. Isot, 192, pp.110605~ (2023)[DOI]
  7. Y. Choi, "Development of a polystyrene phantom for quality assurance of a Gamma Knife", Nucl. Eng. Technol, 55, pp.2935~2940 (2023)[DOI]
  8. T. Cui, "Monte Carlo simulation and study of REE", PET composites with wide γ-ray protection, 2023, pp.55, 2919~2926
  9. K. -P. Chang, "A Comparison of physical and dosimetric properties of lung substitute materials : comparison of cork and others as lung substitutes", Med. Phys, 39, pp.2013~2020 (2012)[DOI]
  10. P. Sopapan, "Determination of Xray and gamma-ray shielding capabilities of recycled glass derived from deteriorated silica gel", Nucl. Eng. Technol, 55, pp.3441~3449 (2023)[DOI]
  11. L. Gerward, "WinXCom—a program for calculating X-ray attenuation coefficients", Radiat. Phys. Chem, 71, pp.653~654 (2004)[DOI]
  12. F. Smekens, "Split exponential track length estimator for Monte-Carlo simulations of small-animal radiation therapy", Phys. Med. Biol, 59, pp.7703~7715 (2014)[DOI]
  13. P. Papadimitroulas, "Dosimetry applications in GATE Monte Carlo toolkit", Phys. Med, 41, pp.136~140 (2017)[DOI]
  14. I. Fathi, "GATE", GEANT4 simulation of radiation risk induced cancer from mammographic screening, 2022, pp.193, 109929~
  15. T. Kostou, "Size-specific dose estimations for pediatric chest, abdomen", pelvis and head CT scans with the use of GATE, 2019, pp.65, 181~190
  16. I. J. Chetty, "Report of the AAPM Task Group No. 105 : issues associated with clinical implementation of Monte Carlo-based photon and electron external beam treatment planning", Med. Phys, 34, pp.4818~4853 (2007)[DOI]
  17. J. M. Boone, "Method for evaluating bow tie filter angle-dependent attenuation in CT : theory and simulation results : bow tie filter characterization in CT", Med. Phys, 37, pp.40~48 (2009)[DOI]
  18. J. J. DeMarco, "A Monte Carlo based method to estimate radiation dose from multidetector CT(MDCT) : cylindrical and anthropomorphic phantoms", Phys. Med. Biol, 50, pp.3989~4004 (2005)[DOI]
  19. M. Alshipli, "Measurement of attenuation coefficients and CT numbers of epoxy resin and epoxy-based Rhizophora spp particleboards in computed tomography energy range", Radiat. Phys. Chem, 149, pp.41~48 (2018)
  20. R. H. Asiah, "Development of in-house head computed tomography dose index phantoms based on polyester-resin materials", Iranian Journal of Medical Physics, 18 (2021)
  21. K. Sato, "Accuracy of virtual monochromatic images generated by the decomposition of photoelectric absorption and Compton scatter in dual-energy computed tomography", Phys Eng Sci Med, 45, pp.239~249 (2022)[DOI]
  22. C. Lee, "Organ dose conversion coefficients in CT scans for Korean adult males and females", Nucl. Eng. Technol, 54, pp.681~688 (2022)[DOI]
  23. M. El Mansouri, "Impact of tube voltage on radiation dose(CTDI)and image quality at chest CT examination", At. Indones, 47, pp.105~ (2021)[DOI]
  24. I. Amini, "Construction and verification of a physical chest phantom from suitable tissue equivalent materials for computed tomography examinations", Radiat. Phys. Chem, 150, pp.51~57 (2018)
  25. C. Anam, "Comparison of central, peripheral, and weighted size-specific dose in CT", XST, 28, pp.695~708 (2020)[DOI]
  26. P. Dio, "Evaluation of radiation dose accuracy calculated using IndoseCT software with direct measurement on polyester-resin phantoms", Radiat. Phys. Chem, 201, pp.110473~ (2022)[DOI]

관련 논문AI추천

Radiation properties and dosimetric insights of Nylon CT phantoms: Advanced computational analysis using Phy-X, XCOM, and GATE/GEANT4El rhazouani O., Halimi A., Khallouqi A., Sekkat H.한국원자력학회 · Nuclear Engineering and Technology2025 · 제57권 제12호Evaluating backscatter factor variations in diagnostic radiology using different phantom materialsEl rhazouani O., Halimi A., Khallouqi A., Madkouri Y., Sekkat H.한국원자력학회 · Nuclear Engineering and Technology2025 · 제57권 제8호Evaluating the Kerma coefficients relevant to various water-equivalent materials at different photon energiesAbdi Zahra, Baghani Hamid Reza, Shamsabadi Reza한국원자력학회 · Nuclear Engineering and Technology2025 · 제57권 제2호Assessing radiological properties of novel tissue equivalent materials for heterogeneous pediatric head phantom: Experimental insights and Monte Carlo simulationsEl rhazouani Omar, Halimi Abdellah, Khallouqi Abdellah, Madkouri Youssef, Sekkat Hamza한국원자력학회 · Nuclear Engineering and Technology2025 · 제57권 제7호다양한 팬텀 모양 및 재질에 따른 전산화단층촬영장치 선량 평가이창래대한방사선과학회 · Journal of Radiological Science and Technology2017 · 제40권 제1호Investigation of radiological properties and water equivalence of polymer gel dosimeters in the energy range 0.1–10 keVBüyükyıldız Mehmet, Kaur Parminder, Levet Aytaç, Parsaei Sara, Rafiei Mustafa Mohammad, Thakur Sonika한국원자력학회 · Nuclear Engineering and Technology2025 · 제57권 제6호

이 간행물의 다른 논문 (49건)