During root canal preparation procedures, the potential for instrument breakage is always present. When instrument breakage occurs, it leads to anxiety of the clinician and as well as a metallic obstruction of the canal which hinders further cleaning and shaping. Separated instruments must always be attempted for retrieval and if retrieval is not possible bypass should be tried. With the increased use of nickel-titanium (NiTi)instruments the incidence of separated instruments has increased. A considerable amount of research has been done to understand the various factors related to the fracture of NiTi instruments to minimize its occurrence. This paper presents a review of the literature regarding the fracture of NiTi instruments and also describes a case report showing the use of a modified 18-guage needle and cyanoacrylate glue to retrieve a separated NiTi instrument from the mesiolingual canal of a mandibular first molar. (Restor Dent Endod 2013;38(2):93-97)

Abstract

During root canal preparation procedures, the potential for instrument breakage is always present. When instrument breakage occurs, it leads to anxiety of the clinician and as well as a metallic obstruction of the canal which hinders further cleaning and shaping. Separated instruments must always be attempted for retrieval and if retrieval is not possible bypass should be tried. With the increased use of nickel-titanium (NiTi)instruments the incidence of separated instruments has increased. A considerable amount of research has been done to understand the various factors related to the fracture of NiTi instruments to minimize its occurrence. This paper presents a review of the literature regarding the fracture of NiTi instruments and also describes a case report showing the use of a modified 18-guage needle and cyanoacrylate glue to retrieve a separated NiTi instrument from the mesiolingual canal of a mandibular first molar. (Restor Dent Endod 2013;38(2):93-97)

참고문헌 (18)

  1. Hülsmann M, "Removal of silver cones and fractured instruments using the canal finder system", J Endod, 16, pp.596~600 (1990)
  2. Fors UG, "Endodontic treatment of root canals obstructed by foreign objects", Int Endod J, 19, pp.2~10 (1986)
  3. Vaudt J, "Ex vivo study on root canal instrumentation of two rotary nickel-titanium systems in comparison to stainless steel hand instruments", Int Endod J, 42, pp.22~33 (2009)
  4. Bonaccorso A, "Shaping ability of four nickel-titanium rotary instruments in simulated S-shaped canals", J Endod, 35, pp.883~886 (2009)
  5. Carvalho LA, "A comparison of molar root canal preparation using stainless-steel and Nickel-Titanium instruments", J Endod, 25, pp.807~810 (1999)
  6. Sattapan B, "Defects in rotary nickel-titanium files after clinical use", J Endod, 26, pp.161~165 (2000)
  7. Hülsmann M, "Influence of several factors on the success or failure of removal of fractured instruments from the root canal", Endod Dent Traumatol, 15, pp.252~258 (1999)
  8. Thompson SA, "An overview of nickel-titanium alloys used in dentistry", Int Endod J, 33, pp.297~310 (2000)
  9. Plotino G, "Ultrasonics in endodontics: a review of the literature", J Endod, 33, pp.81~95 (2007)
  10. Cujé J, "The outcome of retained instrument removal in a specialist practice", Int Endod J, 43, pp.545~554 (2010)
  11. Alapati SB, "SEM observations of nickel-titanium rotary endodontic instruments that fractured during clinical use", J Endod, 31, pp.40~43 (2005)
  12. Askeland DR, "The science and engineering of materials", pp.2003, Brooks
  13. Spanaki-Voreadi AP, "Failure mechanism of ProTaper Ni-Ti rotary instruments during clinical use: fractographic analysis", Int Endod J, 39, pp.171~178 (2006)
  14. Parashos P, "Factors influencing defects of rotary nickel titanium endodontic instruments after clinical use", J Endod, 30, pp.722~725 (2004)
  15. Souter NJ, "Complications associated with fractured file removal using an ultrasonic technique", J Endod, 31, pp.450~452 (2005)
  16. Torabinejad M, "Procedural accidents, In Principles and practice of endodontics", pp.310~330 (2002), Saunders
  17. Spili P, "The impact of instrument fracture on outcome of endodontic treatment", J Endod, 31, pp.845~850 (2005)
  18. Ruddle CJ, "Nonsurgical retreatment", J Endod, 30, pp.827~845 (2004)

관련 논문AI추천

Comparative evaluation of the effectiveness of ultrasonic tips versus the Terauchi file retrieval kit for the removal of separated endodontic instrumentsNawal Ruchika Roongta, Pruthi Preeti Jain, Talwar Sangeeta, Verma Mahesh대한치과보존학회 · Restorative Dentistry and Endodontics2020 · 제45권 제2호Pattern of endodontic instrument separation and factors affecting its retrieval: a 10-year retrospective observational study in a postgraduate instituteKarthikeyan Paul Kevin Abishek, Narasimhan Srinivasan, Natanasabapathy Velmurugan, Varghese Aswathi대한치과보존학회 · Restorative Dentistry and Endodontics2025 · 제50권 제1호Predictive factors in the retrieval of endodontic instruments: the relationship between the fragment length and locationConsoli Lizzi Eugenia Pilar, Portigliatti Ricardo, Rodríguez Pablo Alejandro대한치과보존학회 · Restorative Dentistry and Endodontics2024 · 제49권 제4호근관 내 file의 파절 시 대처 방안김진우, 김혜정, 박세희, 장훈상, 조경모대한치과보존학회 · Restorative Dentistry and Endodontics2006 · 제31권 제3호니켈티타늄 전동파일 파절의예방 및 처치김현철대한치과의사협회 · 대한치과의사협회지2016 · 제54권 제8호Investigation of fracture prevalence of instruments used in root canal treatments at a faculty of dentistry: a prospective studyEskibağlar Mehmet, Ocak Mevlüt Sinan, Özata Merve Yeniçeri, Öztekin Faruk대한치과보존학회 · Restorative Dentistry and Endodontics2023 · 제48권 제4호