Microgrooves on titanium surface affect peri-implant cell adhesion and soft tissue sealing; an in vitro and in vivo study
2
조회
0
다운로드
2
인용
0
참고문헌
키워드
초록
Purpose: With the significance of stable adhesion of alveolar bone and peri-implant soft tissue on the surface of titanium for successful dental implantation procedure, the purpose of this study was to apply microgrooves on the titanium surface and investigate their effects on peri-implant cells and tissues. Methods: Three types of commercially pure titanium discs were prepared; machined-surface discs (A), sandblasted, large-grit, acid-etched (SLA)-treated discs (B), SLA and microgroove- formed discs (C). After surface topography of the discs was examined by confocal laser scanning electron microscopy, water contact angle and surface energy were measured. Human gingival fibroblasts (hGFs) and murine osteoblastic cells (MC3T3-E1) were seeded onto the titanium discs for immunofluorescence assay of adhesion proteins. Commercially pure titanium implants with microgrooves on the coronal microthreads design were inserted into the edentulous mandible of beagle dogs. After 2 weeks and 6 weeks of implant insertion, the animal subjects were euthanized to confirm peri-implant tissue healing pattern in histologic specimens. Results: Group C presented the lowest water contact angle (62.89±5.66 θ), highest surface energy (45±1.2 mN/m), and highest surface roughness (Ra=22.351±2.766 μm). The expression of adhesion molecules of hGFs and MC3T30E1 cells was prominent in group C. Titanium implants with microgrooves on the coronal portion showed firm adhesion to periimplant soft tissue. Conclusions: Microgrooves on the titanium surface promoted the adhesion of gingival fibroblasts and osteoblastic cells, as well as favorable peri-implant soft tissue sealing.
Abstract
Purpose: With the significance of stable adhesion of alveolar bone and peri-implant soft tissue on the surface of titanium for successful dental implantation procedure, the purpose of this study was to apply microgrooves on the titanium surface and investigate their effects on peri-implant cells and tissues. Methods: Three types of commercially pure titanium discs were prepared; machined-surface discs (A), sandblasted, large-grit, acid-etched (SLA)-treated discs (B), SLA and microgroove- formed discs (C). After surface topography of the discs was examined by confocal laser scanning electron microscopy, water contact angle and surface energy were measured. Human gingival fibroblasts (hGFs) and murine osteoblastic cells (MC3T3-E1) were seeded onto the titanium discs for immunofluorescence assay of adhesion proteins. Commercially pure titanium implants with microgrooves on the coronal microthreads design were inserted into the edentulous mandible of beagle dogs. After 2 weeks and 6 weeks of implant insertion, the animal subjects were euthanized to confirm peri-implant tissue healing pattern in histologic specimens. Results: Group C presented the lowest water contact angle (62.89±5.66 θ), highest surface energy (45±1.2 mN/m), and highest surface roughness (Ra=22.351±2.766 μm). The expression of adhesion molecules of hGFs and MC3T30E1 cells was prominent in group C. Titanium implants with microgrooves on the coronal portion showed firm adhesion to periimplant soft tissue. Conclusions: Microgrooves on the titanium surface promoted the adhesion of gingival fibroblasts and osteoblastic cells, as well as favorable peri-implant soft tissue sealing.
참고문헌 (17)
- Brånemark PI, "Osseointegrated titanium fixtures in the treatment of edentulousness", Biomaterials, 4, pp.25~28 (1983)
- Hermann JS, "Biologic width around titanium implants. A physiologically formed and stable dimension over time", Clin Oral Implants Res, 11, pp.1~11 (2000)
- Geurs NC, "Soft tissue considerations in implant site development", Oral Maxillofac Surg Clin North Am, 22, pp.387~405 (2010)
- Abrahamsson I, "The mucosal attachment to titanium implants with different surface characteristics: an experimental study in dogs", J Clin Periodontol, 29, pp.448~455 (2002)
- Huh JB, "Influence of Implant transmucosal design on early peri-implant tissue responses in beagle dogs", Clin Oral Implants Res, 25, pp.962~968 (2014)
- Owens DK, "Estimation of surface free energy of polymers", J Appl Polym Sci, 13, pp.1741~1747 (1969)
- Bächle M, "A systematic review of the influence of different titanium surfaces on proliferation, differentiation and protein synthesis of osteoblast-like MG63 cells", Clin Oral Implants Res, 15, pp.683~692 (2004)
- Cooper LF, "Fluoride modification effects on osteoblast behavior and bone formation at TiO2 grit-blasted c.p. titanium endosseous implants", Biomaterials, 27, pp.926~936 (2006)
- Kim MJ, "Activation of phospholipase D1 by surface roughness of titanium in MG63 osteoblast-like cell", Biomaterials, 27, pp.5502~5511 (2006)
- Marinucci L, "Effect of titanium surface roughness on human osteoblast proliferation and gene expression in vitro", Int J Oral Maxillofac Implants, 21, pp.719~725 (2006)
- Sader MS, "Effect of three distinct treatments of titanium surface on osteoblast attachment, proliferation, and differentiation", Clin Oral Implants Res, 16, pp.667~675 (2005)
- Cochran DL, "Biologic width around titanium implants. A histometric analysis of the implanto-gingival junction around unloaded and loaded nonsubmerged implants in the canine mandible", J Periodontol, 68, pp.186~198 (1997)
- Ruggeri A, "Supracrestal circular collagen fiber network around osseointegrated nonsubmerged titanium implants", Clin Oral Implants Res, 3, pp.169~175 (1992)
- Broggini N, "Peri-implant inflammation defined by the implant-abutment interface", J Dent Res, 85, pp.473~478 (2006)
- Chehroudi B, "The role of connective tissue in inhibiting epithelial downgrowth on titanium-coated percutaneous implants", J Biomed Mater Res, 26, pp.493~515 (1992)
- Linkevicius T, "Biologic width around implants. An evidence-based review", Stomatologija, 10, pp.27~35 (2008)
- Gargiulo AW, "Dimensions and relations of dentogingival junction in humans", J Periodontol, 32, pp.261~267 (1961)