본문 바로가기
KERD
논문KCI 등재

Performance Accuracy of Wrist-Worn Oximetry and Its Automated Output Parameters for Screening Obstructive Sleep Apnea in Children

1

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

Objectives: Obstructive sleep apnea (OSA) increases the risk of perioperative adverse events in children. While polysomnography (PSG) remains the reference standard for OSA diagnosis, oximetry is a valuable screening tool. The traditional practice is the manual analysis of desaturation clusters derived from a tabletop device using the McGill oximetry score. However, automated analysis of wearable oximetry data can be an alternative. This study investigated the accuracy of wrist-worn oximetry with automated analysis as a preoperative OSA screening tool. Methods: Healthy children scheduled for adenotonsillectomy underwent concurrent overnight PSG and wrist-worn oximetry. PSG determined the obstructive apnea-hypopnea index (OAHI). Oximetry data were auto-analyzed to determine 3% oxygen desaturation index (ODI3) and visually scored as per McGill criteria. The logistic regression model assessed the predictive performance of ODI3 for detecting the presence and severity of OSA after adjusting for covariates. Results: Seventy-six children (34 females), aged (mean±standard deviation) 5.7±1.6 years were classified, based on PSG-derived OAHI, as no OSA (n=31), mild (n=31), and moderate-severe OSA (n=14). Oximetric ODI3 was identified as the sole predictor of moderate-severe OSA (OAHI≥5 events/h) (odds ratio 1.38, 95% confidence interval 1.15, 1.65, <i>p</i>=0.001). The best diagnostic performance was at ODI3=5 events/h (78.6% sensitivity, 75.8% specificity [receiver operating characteristic-area under the curve {ROC-AUC}=0.857]). ODI3 was also more sensitive than the McGill oximetry score in diagnosing moderate-severe OSA (78.6% by ODI3 vs. 33.0% by McGill). The performance was suboptimal for any level of OSA (OAHI≥1 event/h) (75.6% sensitivity, 61.3% specificity [ROC-AUC=0.709]). Conclusions: Wrist-worn oximetry-derived automated ODI3 can reliably identify moderate-severe OSA in children undergoing adenotonsillectomy, making it a potentially useful preoperative OSA screening tool.

Abstract

Objectives: Obstructive sleep apnea (OSA) increases the risk of perioperative adverse events in children. While polysomnography (PSG) remains the reference standard for OSA diagnosis, oximetry is a valuable screening tool. The traditional practice is the manual analysis of desaturation clusters derived from a tabletop device using the McGill oximetry score. However, automated analysis of wearable oximetry data can be an alternative. This study investigated the accuracy of wrist-worn oximetry with automated analysis as a preoperative OSA screening tool. Methods: Healthy children scheduled for adenotonsillectomy underwent concurrent overnight PSG and wrist-worn oximetry. PSG determined the obstructive apnea-hypopnea index (OAHI). Oximetry data were auto-analyzed to determine 3% oxygen desaturation index (ODI3) and visually scored as per McGill criteria. The logistic regression model assessed the predictive performance of ODI3 for detecting the presence and severity of OSA after adjusting for covariates. Results: Seventy-six children (34 females), aged (mean±standard deviation) 5.7±1.6 years were classified, based on PSG-derived OAHI, as no OSA (n=31), mild (n=31), and moderate-severe OSA (n=14). Oximetric ODI3 was identified as the sole predictor of moderate-severe OSA (OAHI≥5 events/h) (odds ratio 1.38, 95% confidence interval 1.15, 1.65, <i>p</i>=0.001). The best diagnostic performance was at ODI3=5 events/h (78.6% sensitivity, 75.8% specificity [receiver operating characteristic-area under the curve {ROC-AUC}=0.857]). ODI3 was also more sensitive than the McGill oximetry score in diagnosing moderate-severe OSA (78.6% by ODI3 vs. 33.0% by McGill). The performance was suboptimal for any level of OSA (OAHI≥1 event/h) (75.6% sensitivity, 61.3% specificity [ROC-AUC=0.709]). Conclusions: Wrist-worn oximetry-derived automated ODI3 can reliably identify moderate-severe OSA in children undergoing adenotonsillectomy, making it a potentially useful preoperative OSA screening tool.

참고문헌 (30)

  1. Waters KA, "Sleep disorders in children", Med J Aust, 199, pp.S31~S35 (2013)[DOI]
  2. Patino M, "Obstructive sleep apnoea in children : perioperative considerations", Br J Anaesth, 111(Suppl 1), pp.I83~I95 (2013)[DOI]
  3. von Ungern-Sternberg BS, "Effect of albuterol premedication vs placebo on the occurrence of respiratory adverse events in children undergoing tonsillectomies : the REACT randomized clinical trial", JAMA Pediatr, 173, pp.527~533 (2019)[DOI]
  4. Ramgolam A, "Deep or awake removal of laryngeal mask airway in children at risk of respiratory adverse events undergoing tonsillectomy-a randomised controlled trial", Br J Anaesth, 120, pp.571~580 (2018)[DOI]
  5. Ohn M, "Preoperative identification of children at high risk of obstructive sleep apnea", Paediatr Anaesth, 30, pp.221~231 (2020)[DOI]
  6. Kang M, "Trends in diagnosing obstructive sleep apnea in pediatrics", Children(Basel), 9, pp.306~ (2022)[DOI]
  7. Twiss J, "Overnight oximetry for evaluating paediatric obstructive sleep apnoea : technical specifications and interpretation guidelines", J Paediatr Child Health, 55, pp.1279~ (2019)[DOI]
  8. Chung F, "Oxygen desaturation index from nocturnal oximetry : a sensitive and specific tool to detect sleep-disordered breathing in surgical patients", Anesth Analg, 114, pp.993~1000 (2012)[DOI]
  9. Álvarez D, "Automated screening of children with obstructive sleep apnea using nocturnal oximetry : an alternative to respiratory polygraphy in unattended settings", J Clin Sleep Med, 13, pp.693~702 (2017)[DOI]
  10. Hornero R, "Nocturnal oximetry–based evaluation of habitually snoring children", Am J Respir Crit Care Med, 196, pp.1591~1598 (2017)[DOI]
  11. Fabius TM, "Validation of the oxygen desaturation index in the diagnostic workup of obstructive sleep apnea", Sleep Breath, 23, pp.57~63 (2019)[DOI]
  12. Del Campo F, "Oximetry use in obstructive sleep apnea", Expert Rev Respir Med, 12, pp.665~681 (2018)[DOI]
  13. Gao X, "Diagnostic accuracy of level IV portable sleep monitors versus polysomnography for pediatric obstructive sleep apnea : a systematic review and meta-analysis", Sleep Med, 87, pp.127~137 (2021)[DOI]
  14. Ma JR, "Value of pulse oximetry watch for diagnosing pediatric obstructive sleep apnea", hypopnea syndrome, 2018, pp.138, 175~179
  15. Thavagnanam S, "WRISTOX2 is a reliable tool to diagnose obstructive sleep apnoea syndrome", Int J Pediatr Otorhinolaryngol, 151, pp.110930~ (2021)[DOI]
  16. Jorgensen G, "An Australasian commentary on the AASM manual for the scoring of sleep and associated events", Sleep Biol Rhythms, 18, pp.163~185 (2020)[DOI]
  17. Berry RB, "AASM scoring manual updates for 2017(version 2. 4)", J Clin Sleep Med, 13, pp.665~666 (2017)[DOI]
  18. Berry RB, "Rules for scoring respiratory events in sleep : update of the 2007 AASM manual for the scoring of sleep and associated events. Deliberations of the sleep apnea definitions task force of the American Academy of Sleep Medicine", J Clin Sleep Med, 8, pp.597~619 (2012)[DOI]
  19. Nixon GM, "Planning adenotonsillectomy in children with obstructive sleep apnea : the role of overnight oximetry", Pediatrics, 113(1 Pt 1), pp.e19~e25 (2004)[DOI]
  20. Chervin RD, "Pediatric sleep questionnaire(PSQ) : validity and reliability of scales for sleep-disordered breathing, snoring, sleepiness, and behavioral problems", Sleep Med, 1, pp.21~32 (2000)[DOI]
  21. Bursac Z, "Purposeful selection of variables in logistic regression", Source Code Biol Med, 3, pp.17~ (2008)[DOI]
  22. Fan J, "Understanding receiver operating characteristic(ROC)curves", CJEM, 8, pp.19~20 (2006)[DOI]
  23. Fluss R, "Estimation of the Youden index and its associated cutoff point", Biom J, 47, pp.458~472 (2005)[DOI]
  24. Martínez-Camblor P, "The Youden index in the generalized receiver operating characteristic curve context", Int J Biostat, 15, pp.20180060~ (2019)[DOI]
  25. Brockmann PE, "Gender dimorphism in pediatric OSA : is it for real?", Respir Physiol Neurobiol, 245, pp.83~88 (2017)[DOI]
  26. Horne RSC, "Are there gender differences in the severity and consequences of sleep disordered in children?", Sleep Med, 67, pp.147~155 (2020)[DOI]
  27. Brouillette RT, "Nocturnal pulse oximetry as an abbreviated testing modality for pediatric obstructive sleep apnea", Pediatrics, 105, pp.405~412 (2000)[DOI]
  28. McNamara F, "Arousal pattern following central and obstructive breathing abnormalities in infants and children", J Appl Physiol (1985), 81, pp.2651~2657 (1996)[DOI]
  29. Mograss MA, "Movement", arousals. Description, classification, and relationship to sleep apnea in children, 1994, pp.150(6 Pt 1), 1690~1696
  30. Kirk VG, "Comparison of home oximetry monitoring with laboratory polysomnography in children", Chest, 124, pp.1702~1708 (2003)[DOI]