인공와우 이식술 후 발생한 자석전위 1예
Magnet Displacement Following Cochlear Implantation
1
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
Abstract
Background and Objectives:To describe a rare complication of cochlear implantation, such as magnet displacement which can be diagnosed easily with simple skull radiographs and treated with minimal surgery. Materials and Method:The record of male who experienced CI magnet dislodgement was reviewed. Results:The patient was seen with a displaced magnet under the skin 15 days after a head trauma (1 month after implantation). The patient was implanted with the Nucleus 24 device at the age of 3 years. Plain skull radiographs showed the dislocated magnet. Surgery was limited to the revision of the receiver/stimulator site and reinsertion of the magnet to its pocket. Intraoperative neural response telemetry was normal in the patient. Postoperative auditory Conclusion:The displaced magnet can be successfully ma-naged by a simple revision procedure. The posibility of magnet migration should be considered in cases of device malfunction. (Korean J Otorhinolaryngol-Head Neck Surg 2007 ;50 :1051-3)
참고문헌 (4)
- "Cochlear implant magnet displa-cement with minor head trauma", 133(2), pp.275~277 (2005)
- "Magnet displacement folowing cochlear implantation", 26(4), pp.646~648 (2005)
- "Eisenmenger BA. Surgical and nonsurgical complications associated with cochlear prosthesis impl-antation. Am J Otol 1990", (lph mb), pp.415~20