Background and Objectives A lack of investigators for polysomnography has risen due toincreased demand since health insurance started to cover the cost of the test. We examined thereliability of the automated scoring of polysomnography, which has been deployed to improvethis imbalance. Subjects and Method We analyzed the data of 20 patients who underwent level 1 polysomnographyfrom April 1 to July 27, 2019. The software from Noxturnal (Nox Medical) was usedfor the scoring of the Polysomnography data. Each of the polysomnographic data was scoredboth by the automated scoring system and by a skilled technician. Results Twenty patients were analyzed. There was no significant difference between automatedscoring and manual scoring in sleep latency, apnea index, and rapid eye movementsleep stage ratio. However, the concordance rate of the sleep stage by epoch was 83.32%, andthere was a significant difference with regards to apnea-hypoapnea index (AHI) and respiratorydisturbance index (RDI). Two obvious errors were noted in the automated scoring that couldbe easily fixed; the failure to recognize wakefulness during sleep and the fragmentation of respiratoryevents. When two errors were corrected, many differences in polysomnography parameters,including AHI and RDI, were eliminated. Conclusion It showed 80% coincidence of epoch in the sleep stage between the automatedscoring and manual scoring. However, there was no difference in AHI and RDI when the fragmentedrespiratory events of the automated scoring were adjusted. Therefore, automated scoringis considered to be useful if only a little modification could be made.
Abstract
Background and Objectives A lack of investigators for polysomnography has risen due toincreased demand since health insurance started to cover the cost of the test. We examined thereliability of the automated scoring of polysomnography, which has been deployed to improvethis imbalance. Subjects and Method We analyzed the data of 20 patients who underwent level 1 polysomnographyfrom April 1 to July 27, 2019. The software from Noxturnal (Nox Medical) was usedfor the scoring of the Polysomnography data. Each of the polysomnographic data was scoredboth by the automated scoring system and by a skilled technician. Results Twenty patients were analyzed. There was no significant difference between automatedscoring and manual scoring in sleep latency, apnea index, and rapid eye movementsleep stage ratio. However, the concordance rate of the sleep stage by epoch was 83.32%, andthere was a significant difference with regards to apnea-hypoapnea index (AHI) and respiratorydisturbance index (RDI). Two obvious errors were noted in the automated scoring that couldbe easily fixed; the failure to recognize wakefulness during sleep and the fragmentation of respiratoryevents. When two errors were corrected, many differences in polysomnography parameters,including AHI and RDI, were eliminated. Conclusion It showed 80% coincidence of epoch in the sleep stage between the automatedscoring and manual scoring. However, there was no difference in AHI and RDI when the fragmentedrespiratory events of the automated scoring were adjusted. Therefore, automated scoringis considered to be useful if only a little modification could be made.
참고문헌 (21)
- 이영찬, "Prevalence of Snoring and High Risk of Obstructive Sleep Apnea Syndrome in Young Male Soldiers in Korea", Journal of Korean Medical Science, 28(9), pp.1373~1377 (2013), 대한의학회
- Peppard PE, "Prospective study of the association between sleep-disordered breathing and hypertension", N Engl J Med, 342(19), pp.1378~1384 (2000)
- Canessa N, "Obstructive sleep apnea : Brain structural changes and neurocognitive function before and after treatment", Am J Respir Crit Care Med, 183(10), pp.1419~1426 (2011)
- Yang HC, "The efficacy of drug induced sleep endoscopy using multimodality monitoring system", PLoS One, 13(12), pp.e0209775~ (2018)
- Sung CM, "The clinical characteristics of patients with an isolate epiglottic collapse", Auris Nasus Larynx, 47(3), pp.450~457 (2020)
- Yoon DW, "Evaluation of the feasibility and preference of Nox-A1 type 2 ambulatory device for unattended home sleep test : A randomized crossover study", Sleep Biol Rhythms, 17(3), pp.297~304 (2019)
- Xu L, "Validation of the Nox-T3 portable monitor for diagnosis of obstructive sleep apnea in Chinese adults", J Clin Sleep Med, 13(5), pp.675~683 (2017)
- Cairns A, "A pilot validation study for the NOX T3(TM)portable monitor for the detection of OSA", Sleep Breath, 18(3), pp.609~614 (2014)
- Epstein LJ, "Clinical guideline for the evaluation, management and long-term care of obstructive sleep apnea in adults", J Clin Sleep Med, 5(3), pp.263~276 (2009)
- Berry RB, "AASM scoring manual updates for 2017 (version 2.4)", J Clin Sleep Med, 13(5), pp.665~666 (2017)
- Kushida CA, "Practice parameters for the indications for polysomnography and related procedures : An update for 2005", Sleep, 28(4), pp.499~521 (2005)
- Collop NA, "Scoring variability between polysomnography technologists in different sleep laboratories", Sleep Med, 3(1), pp.43~47 (2002)
- Schaltenbrand N, "Sleep stage scoring using the neural network model : Comparison between visual and automatic analysis in normal subjects and patients", Sleep, 19(1), pp.26~35 (1996)
- Malhotra A, "Performance of an automated polysomnography scoring system versus computer-assisted manual scoring", Sleep, 36(4), pp.573~582 (2013)
- Redline S, "The scoring of respiratory events in sleep : Reliability and validity", J Clin Sleep Med, 3(2), pp.169~200 (2007)
- Oztürk O, "The concordance of manuel (visual) scoring and automatic analysis in sleep staging", Tuberk Toraks, 57(3), pp.306~313 (2009)
- Barreiro B, "Comparison between automatic and manual analysis in the diagnosis of obstructive sleep apnea-hypopnea syndrome", Arch Bronconeumol, 39(12), pp.544~548 (2003)
- Sforza E, "Automatic Oxford-Medilog 9200 sleep staging scoring : Comparison with visual analysis", J Clin Neurophysiol, 13(3), pp.227~233 (1996)
- Carrasco O, "Visual and different automatic scoring profiles of respiratory variables in the diagnosis of sleep apnoea-hypopnoea syndrome", Eur Respir J, 9(1), pp.125~130 (1996)
- Cachada N, "Comparison of manual and automatic scoring of limited channel sleep studies: Noxturnal software correlates well with manual scoring in severe OSA", Eur Respir J, 50, pp.PA2301~ (2017)
- Younes M, "Staging sleep in polysomnograms : Analysis of inter-scorer variability", J Clin Sleep Med, 12(6), pp.885~894 (2016)