초록
PURPOSE. The aim of this study was to evaluate the clinical performance and reliability of plasma sprayed nanostructured zirconia (NSZ) coating. MATERIALS AND METHODS. This study consisted of three areas of analysis: (1) Mechanical property: surface roughness of NSZ coating and bond strength between NSZ coating and titanium specimens were measured, and the microstructure of bonding interface was also observed by scanning election microscope (SEM). (2) Biocompatibility: hemolysis tests, cell proliferation tests, and rat subcutaneous implant test were conducted to evaluate the biocompatibility of NSZ coating. (3) Mechanical compatibility: fracture and artificial aging tests were performed to measure the mechanical compatibility of NSZcoated titanium abutments. RESULTS. In the mechanical study, 400 μm thick NSZ coatings had the highest bond strength (71.22 ± 1.02 MPa), and a compact transition layer could be observed. In addition, NSZ coating showed excellent biocompatibility in both hemolysis tests and cell proliferation tests. In subcutaneous implant test, NSZcoated plates showed similar inflammation elimination and fibrous tissue formation processes with that of titanium specimens. Regarding fatigue tests, all NSZ-coated abutments survived in the five-year fatigue test andshowed sufficient fracture strength (407.65-663.7 N) for incisor teeth. CONCLUSION. In this study, the plasmasprayed NSZ-coated titanium abutments presented sufficient fracture strength and biocompatibility, and it was demonstrated that plasma spray was a reliable method to prepare high-quality zirconia coating.
Abstract
PURPOSE. The aim of this study was to evaluate the clinical performance and reliability of plasma sprayed nanostructured zirconia (NSZ) coating. MATERIALS AND METHODS. This study consisted of three areas of analysis: (1) Mechanical property: surface roughness of NSZ coating and bond strength between NSZ coating and titanium specimens were measured, and the microstructure of bonding interface was also observed by scanning election microscope (SEM). (2) Biocompatibility: hemolysis tests, cell proliferation tests, and rat subcutaneous implant test were conducted to evaluate the biocompatibility of NSZ coating. (3) Mechanical compatibility: fracture and artificial aging tests were performed to measure the mechanical compatibility of NSZcoated titanium abutments. RESULTS. In the mechanical study, 400 μm thick NSZ coatings had the highest bond strength (71.22 ± 1.02 MPa), and a compact transition layer could be observed. In addition, NSZ coating showed excellent biocompatibility in both hemolysis tests and cell proliferation tests. In subcutaneous implant test, NSZcoated plates showed similar inflammation elimination and fibrous tissue formation processes with that of titanium specimens. Regarding fatigue tests, all NSZ-coated abutments survived in the five-year fatigue test andshowed sufficient fracture strength (407.65-663.7 N) for incisor teeth. CONCLUSION. In this study, the plasmasprayed NSZ-coated titanium abutments presented sufficient fracture strength and biocompatibility, and it was demonstrated that plasma spray was a reliable method to prepare high-quality zirconia coating.
참고문헌 (27)
- Long M, "Titanium alloys in total joint replacement-a materials science perspective", Biomaterials, 19, pp.1621~1639 (1998)
- Wadhwani CP, "Laboratory technique for coloring titanium abutments to improve esthetics", J Prosthet Dent, 115, pp.409~411 (2016)
- 이정진, "Evaluation of effect of galvanic corrosion between nickel-chromium metal and titanium on ion release and cell toxicity", The Journal of Advanced of Prosthodontics, 7(2), pp.172~177 (2015), 대한치과보철학회[DOI]
- Guo H, "Adenovirus-mediated small interfering RNA targeting tumor necrosis factor-α inhibits titanium particle-induced osteoclastogenesis and bone resorption", Int J Mol Med, 32, pp.296~306 (2013)
- Zhu W, "Nondestructive inspection of phase transformation in zirconia-containing hip joints by confocal Raman spectroscopy", J Biomed Opt, 18, pp.127002~ (2013)
- Céline Brunot-Gohin, "Biocompatibility study of lithium disilicate and zirconium oxide ceramics for esthetic dental abutments", Journal of Periodontal & Implant Science, 46(6), pp.362~371 (2016), 대한치주과학회
- Kuo C, "Phase transformation behavior of 3mol% yttria partially-stabilized ZrO2(3Y-PSZ)precursor powder by an isothermal method", Ceramics International, 40, pp.3243~3251 (2014)
- Mühlemann S, "Bending moments of zirconia and titanium implant abutments supporting all-ceramic crowns after aging", Clin Oral Implants Res, 25, pp.74~81 (2014)
- D’Ercole S, "Microleakage of bacteria in different implant-abutment assemblies : an in vitro study", J Appl Biomater Funct Mater, 13, pp.e174~e180 (2015)
- Marcelli E, "Effect of intermediate ZrO2-CaO coatings deposited by cold thermal spraying on the titaniumporcelain bond in dental restorations", J Prosthet Dent, 112, pp.1201~1211 (2014)
- Vardelle A, "A perspective on plasma spray technology", Plasma Chem Plasma Process, 35, pp.491~509 (2015)
- Prasauskas T, "Filter media properties of mineral fibres produced by plasma spray", Environ Technol, 37, pp.1315~1324 (2016)
- Guo M, "Kgp, rgpA, and rgpB DNA vaccines induce antibody responses in experimental peri-implantitis", J Periodontol, 85, pp.1575~1581 (2014)
- Reill MI, "Influence of core material on fracture resistance and marginal adaptation of restored root filled teeth", Int Endod J, 41, pp.424~430 (2008)
- Kohal RJ, "Stability of prototype twopiece zirconia and titanium implants after artificial aging : an in vitro pilot study", Clin Implant Dent Relat Res, 11, pp.323~329 (2009)
- Amarnath LP, "In vitro hemocompatibility testing of UV-modified hyaluronan hydrogels", Biomaterials, 27, pp.1416~1424 (2006)
- Khor KA, "Microstructure and mechanical properties of plasma sprayed HA", YSZ, Biomaterials, pp.2004, 25[DOI]
- Zhang Y, "Mesoporous bioactive glass nanolayer-modified zirconia coatings on Ti-6Al-4V with improved in vitro bioactivity", Int J Appl Glass Sci, 7, pp.216~228 (2016)
- Jemata A, "Effects of TiO2 on microstructural, mechanical properties and in-vitro bioactivity of plasma sprayed yttria stabilised zirconia coatings for dental application", Ceram Int, 44, pp.4271~4281 (2018)
- Wongkamhaeng K, "Effect of surface modification on in-depth transformations and flexural strength of zirconia ceramics", J Prosthodont, 28, pp.e364~e375 (2019)
- Guan SH, "Energy landscape of zirconia phase transitions", J Am Chem Soc, 137, pp.8010~8013 (2015)
- Pittayachawan P, "Flexural strength, fatigue life, and stress-induced phase transformation study of Y-TZP dental ceramic", J Biomed Mater Res B Appl Biomater, 88, pp.366~377 (2009)
- Tortopidis D, "The variability of bite force measurement between sessions, in different positions within the dental arch", J Oral Rehabil, 25, pp.681~686 (1998)
- Gehrke P, "In vitro fatigue and fracture resistance of one-and two-piece CAD", CAM zirconia implant abutments, 2015, pp.30, 546~554
- Alsahhaf A, "Fracture resistance of zirconia-based implant abutments after artificial long-term aging", J Mech Behav Biomed Mater, 66, pp.224~232 (2017)
- Xing R, "The influence of surface nanoroughness, texture and chemistry of TiZr implant abutment on oral biofilm accumulation", Clin Oral Implants Res, 26, pp.649~656 (2015)
- Herbst D, "Evaluation of surface characteristics of titanium and cobalt chromium implant abutment materials", SADJ, 68, pp.350~ (2013)