Performance Evaluation of the Beckman Coulter DxN VERIS Hepatitis B Virus (HBV) Assay in Comparison With the Abbott RealTime HBV Assay

0

조회

0

다운로드

1

인용

0

참고문헌

키워드

주제분류병리학

초록

The detection and quantification of hepatitis B virus (HBV) DNA plays an important role in diagnosing and monitoring HBV infection as well as in assessing the therapeutic response. We compared the analytical performance of a random access, fully automated HBV assay—DxN VERIS Molecular Diagnostics System (Beckman Coulter, Brea, CA, USA)—with that of Abbott RealTime HBV assay (Abbott Laboratories, Des Plaines, IL, USA). The between-day precision of the VERIS assay ranged from 0.92% (mean 4.68 log IU/mL) to 4.15% (mean 2.09 log IU/mL) for pooled sera from HBV patients. HBV DNA levels measured by the VERIS HBV assay correlated with the calculated HBV DNA levels (r2=0.9994; P<0.0001). The lower limit of quantification was estimated as 8.76 IU/mL (Probit analysis, 95% confidence interval: 7.32–12.00 IU/mL). Passing-Bablok regression analysis showed good concordance between the VERIS and RealTime assays for 187 chronic HBV samples (y=-0.2397+0.9712x; r=0.981), as well as for 20 drug-resistant HBV genotype C positive samples (y=-0.5415+0.9954x; r=0.961). The VERIS assay demonstrated performance similar to the RealTime assay and is suitable for high-throughput HBV DNA monitoring in large hospital laboratories.

Abstract

The detection and quantification of hepatitis B virus (HBV) DNA plays an important role in diagnosing and monitoring HBV infection as well as in assessing the therapeutic response. We compared the analytical performance of a random access, fully automated HBV assay—DxN VERIS Molecular Diagnostics System (Beckman Coulter, Brea, CA, USA)—with that of Abbott RealTime HBV assay (Abbott Laboratories, Des Plaines, IL, USA). The between-day precision of the VERIS assay ranged from 0.92% (mean 4.68 log IU/mL) to 4.15% (mean 2.09 log IU/mL) for pooled sera from HBV patients. HBV DNA levels measured by the VERIS HBV assay correlated with the calculated HBV DNA levels (r2=0.9994; P<0.0001). The lower limit of quantification was estimated as 8.76 IU/mL (Probit analysis, 95% confidence interval: 7.32–12.00 IU/mL). Passing-Bablok regression analysis showed good concordance between the VERIS and RealTime assays for 187 chronic HBV samples (y=-0.2397+0.9712x; r=0.981), as well as for 20 drug-resistant HBV genotype C positive samples (y=-0.5415+0.9954x; r=0.961). The VERIS assay demonstrated performance similar to the RealTime assay and is suitable for high-throughput HBV DNA monitoring in large hospital laboratories.

참고문헌 (16)

  1. Schweitzer A, "Estimations of worldwide prevalence of chronic hepatitis B virus infection: a systematic review of data published between 1965 and 2013", Lancet, 386, pp.1546~1555 (2015)
  2. Zhang Q, "Incidence of natural resistance mutations in naïve chronic hepatitis B patients : a systematic review and meta-analysis", J Gastroenterol Hepatol, 30, pp.252~261 (2015)
  3. European Association for the Study of the Liver, "EASL 2017 Clinical Practice Guidelines on the management of hepatitis B virus infection", J Hepatol, 67, pp.370~398 (2017)
  4. 한미순, "Comparison of the QIAGEN artus HBV QS-RGQ Assay With the Roche COBAS AmpliPrep", COBAS TaqMan HBV Assay for Quantifying Viral DNA in Sera of Chronic Hepatitis B Patients, 2017, pp.37(3), 248~253[DOI]
  5. Terrault NA, "AASLD guidelines for treatment of chronic hepatitis B", Hepatology, 63, pp.261~283 (2016)
  6. Pawlotsky JM, "Virologic monitoring of hepatitis B virus therapy in clinical trials and practice : recommendations for a standardized approach", Gastroenterology, 134, pp.405~415 (2008)
  7. Chevaliez S, "Performance of version 2.0 of the Cobas AmpliPrep", Cobas TaqMan Real-time PCR assay for hepatitis B virus DNA quantification, 2010, pp.48, 3641~3647
  8. Chevaliez S, "New virologic tools for management of chronic hepatitis B and C", Gastroenterology, 142, pp.1303~1313 (2012)[DOI]
  9. Yeh ML, "Comparison of the Abbott RealTime HBV assay with the Roche Cobas AmpliPrep", Cobas TaqMan HBV assay for HBV DNA detection and quantification, 2014, pp.60, 206~214
  10. Fourati S, "Evaluation of a new random-access HBV DNA molecular assay : The VERIS HBV assay", J Clin Virol, 92, pp.69~74 (2017)
  11. Braun P, "A European multicentre study on the comparison of HBV viral loads between VERIS HBV assay and Roche COBAS® TAQMAN® HBV test, Abbott RealTime HBV assay, Siemens VERSANT HBV assay, and Qiagen artus HBV RG kit", J Clin Virol, 95, pp.76~83 (2017)
  12. The Korean Association for the Study of the Liver(KASL), "KASL clinical practice guidelines : management of chronic hepatitis B", Clin Mol Hepatol, 22, pp.18~75 (2016)
  13. 허희재, "Performance of the Real-Q EBV Quantification Kit for Epstein-Barr Virus DNA Quantification in Whole Blood", Annals of Laboratory Medicine, 37(2), pp.147~150 (2017), 대한진단검사의학회
  14. 김한아, "Automated Nucleic Acid Extraction Systems for Detecting Cytomegalovirus and Epstein-Barr Virus Using Real-Time PCR: A Comparison Study Between the QIAsymphony RGQ and QIAcube Systems", Annals of Laboratory Medicine, 37(2), pp.129~136 (2017), 대한진단검사의학회
  15. 박종은, "Performance Evaluation of the Real-Q Cytomegalovirus (CMV) Quantification Kit Using Two Real-Time PCR Systems for Quantifying CMV DNA in Whole Blood", Annals of Laboratory Medicine, 36(6), pp.603~606 (2016), 대한진단검사의학회
  16. 조지현, "B형간염바이러스 유전자형의 국내 분포에 대한 연구", Clinical and Molecular Hepatology, 15(2), pp.140~147 (2009), 대한간학회

관련 논문AI추천

Comparison of the QIAGEN artus HBV QS-RGQ Assay With the Roche COBAS AmpliPrep/COBAS TaqMan HBV Assay for Quantifying Viral DNA in Sera of Chronic Hepatitis B Patients김현숙, 나현진, 박용정, 한미순대한진단검사의학회 · Annals of Laboratory Medicine2017 · 제37권 제3호Performance Evaluation of the Roche Cobas 5800 HBV and HCV Tests: Comparison of the 200 and 500 μL ProtocolsCho Eun-Jung, Jeong Seri, Kim Han-Sung, Kim Hyun Soo, Kim Jae-Seok, Lee Su-Kyung, Lee Young Kyung, Song Wonkeun대한진단검사의학회 · Annals of Laboratory Medicine2024 · 제44권 제3호Elecsys Hepatitis B Surface Antigen Quantitative Assay: Performance Evaluation and Correlation with Hepatitis B Virus DNA during 96 Weeks of Follow-up in Chronic Hepatitis B Patients허정, 김신영, 김형회, 손한철, 이선민, 이은엽, 이현지, 장철훈대한진단검사의학회 · Annals of Laboratory Medicine2012 · 제32권 제6호Performance evaluation of the HepB Typer-Entecavir kit for detection of entecavir resistance mutations in chronic hepatitis B김수옥, 박준용, 신수경, 안상훈, 유왕돈, 천지용, 한광협, 홍선표대한간학회 · Clinical and Molecular Hepatology2013 · 제19권 제4호Performance Evaluation of Aptima HBV and HCV Quant Assays in the Panther System김보연, 김윤정, 이경아, 최유정대한진단검사의학회 · Laboratory Medicine Online2021 · 제11권 제3호실시간 정량적 중합효소연쇄반응을 이용한 Abbott RealTime HBVQuantification Kit의 성능평가김명희, 안동희, 오흥범, 차충환, 최성은대한진단검사의학회 · Annals of Laboratory Medicine2008 · 제28권 제2호