본문 바로가기
KERD
논문KCI 등재

Comparison of Three Blood Collection Tubes for 35 Biochemical Analytes: The Becton Dickinson Barricor Tube, Serum Separating Tube, and Plasma Separating Tube

2

조회

0

다운로드

2

인용

0

참고문헌

키워드

주제분류병리학

초록

The Barricor tube (Becton Dickinson [BD], Sunnyvale, CA, USA) was recently developed to mechanically separate plasma by increasing the centrifugation rate. We compared the Barricor tube with existing serum- and plasma-based tubes based on 35 biochemical analytes and preanalytical turnaround time (TAT). Blood samples were collected from 30 healthy volunteers in a Barricor tube, serum separating tube (SST, Vacutainer SST II Tube 8.5 mL, #368972; BD), or plasma separating tube (PST, Vacutainer PST Tube 8.0 mL, #367964; BD) in random order. Next, 27 chemistry analytes, six immunochemistry analytes, and two cardiac markers were compared using Passing-Bablok regression and the Bland-Altman method. Preanalytical TAT was measured for each tube. The Barricor tube exhibited bias exceeding the desirable limit for nine and four analytes compared with the SST and PST, respectively. The Barricor tube lactate dehydrogenase value showed a bias of -10.29% and -9.86% compared with that of the SST and PST, respectively. The preanalytical TAT of Barricor tube was 8.8 minutes, which was the shortest among the three tubes. The clinical performance of the Barricor tube was equivalent to that of the SST and PST for most analytes, with an apparent advantage in preanalytical TAT. When using the Barricor tube, the reference range needs to be changed for some analytes that exceed the desirable bias limit.

Abstract

The Barricor tube (Becton Dickinson [BD], Sunnyvale, CA, USA) was recently developed to mechanically separate plasma by increasing the centrifugation rate. We compared the Barricor tube with existing serum- and plasma-based tubes based on 35 biochemical analytes and preanalytical turnaround time (TAT). Blood samples were collected from 30 healthy volunteers in a Barricor tube, serum separating tube (SST, Vacutainer SST II Tube 8.5 mL, #368972; BD), or plasma separating tube (PST, Vacutainer PST Tube 8.0 mL, #367964; BD) in random order. Next, 27 chemistry analytes, six immunochemistry analytes, and two cardiac markers were compared using Passing-Bablok regression and the Bland-Altman method. Preanalytical TAT was measured for each tube. The Barricor tube exhibited bias exceeding the desirable limit for nine and four analytes compared with the SST and PST, respectively. The Barricor tube lactate dehydrogenase value showed a bias of -10.29% and -9.86% compared with that of the SST and PST, respectively. The preanalytical TAT of Barricor tube was 8.8 minutes, which was the shortest among the three tubes. The clinical performance of the Barricor tube was equivalent to that of the SST and PST for most analytes, with an apparent advantage in preanalytical TAT. When using the Barricor tube, the reference range needs to be changed for some analytes that exceed the desirable bias limit.

참고문헌 (19)

  1. Fournier JE, "Evaluation of BD Vacutainer Barricor blood collection tubes for routine chemistry testing on a Roche Cobas 8000 Platform", Clin Biochem, 58, pp.94~99 (2018)
  2. Füzéry AK, "The BD Barricor blood collection tube is an acceptable and robust alternative to the PST for use with the Beckman AccuTnI+3 assay", Clin Biochem, 50, pp.851~857 (2017)
  3. Dupuy AM, "Comparison of Barricor vs. lithium heparin tubes for selected routine biochemical analytes and evaluation of post centrifugation stability", Biochem Med (Zagreb), 28, pp.020902~ (2018)
  4. Cadamuro J, "Influence of centrifugation conditions on the results of 77 routine clinical chemistry analytes using standard vacuum blood collection tubes and the new BD-Barricor tubes", Biochem Med (Zagreb), 28, pp.010704~ (2018)
  5. Arslan FD, "The local clinical validation of a new lithium heparin tube with a barrier:BD Vacutainer Barricor LH Plasma tube", Biochem Med (Zagreb), 27, pp.030706~ (2017)
  6. Padoan A, "Quality of plasma samples and BD Vacutainer Barricor tubes: effects of centrifugation", Clin Chim Acta, 483, pp.271~274 (2018)
  7. Ramakers C, "BD Vacutainer Barricor tube in the emergency department:reduced hemolysis rates using partial draw tubes with reduced vacuum", Clin Chem Lab Med, 56, pp.e31~e32 (2018)
  8. "Desirable specifications for total Error, imprecision, and bias, derived from intra- and inter-individual biologic variation"[URL]
  9. "Clinical laboratory improvement amendments of 1988: final rule", Department of Health and Human Services, Health Care Financing Administration; Federal Register 57, no. 7146, ~ (1992)
  10. Erdal EP, "The economic impact of poor sample quality in clinical chemistry laboratories: results from a global survey", Ann Clin Biochem, 54, pp.230~239 (2017)
  11. Boyanton BL Jr, "Stability studies of twenty-four analytes in human plasma and serum", Clin Chem, 48, pp.2242~2247 (2002)
  12. Schrapp A, "The right blood collection tube for therapeutic drug monitoring and toxicology screening procedures: standard tubes, gel or mechanical separator?", Clin Chim Acta, 488, pp.196~201 (2019)
  13. Lippi G, "Influence of hemolysis on routine clinical chemistry testing", Clin Chem Lab Med, 44, pp.311~316 (2006)
  14. Zhang DJ, "Effect of serum-clot contact time on clinical chemistry laboratory results", Clin Chem, 44, pp.1325~1333 (1998)
  15. Monneret D, "Stability of routine biochemical analytes in whole blood and plasma from lithium heparin gel tubes during 6-hr storage", J Clin Lab Anal, 30, pp.602~609 (2016)
  16. Chan FK, "Detection of necrosis by release of lactate dehydrogenase activity", Methods Mol Biol, 979, pp.65~70 (2013)
  17. Bowen RA, "Impact of blood collection devices on clinical chemistry assays", Clin Biochem, 43, pp.4~25 (2010)
  18. Dimeski G, "What is the most suitable blood collection tube for glucose estimation?", Ann Clin Biochem, 52, pp.270~275 (2015)
  19. Lippi G, "Impact of blood cell counts and volumes on glucose concentration in uncentrifuged serum and lithium-heparin blood tubes", Clin Chem Lab Med, 56, pp.2125~2131 (2018)

관련 논문AI추천

Evaluation of the Barricor Tube in 28 Routine Chemical Tests and Its Impact on Turnaround Time in an Outpatient Clinic김인숙, Lee Han Sol, Lee Sun Min, Moon Soo Young, Park Min Soon대한진단검사의학회 · Annals of Laboratory Medicine2021 · 제41권 제3호Comparison of Red Blood Cell Hemolysis Using Plasma and Serum Separation Tubes for Outpatient Specimens고대현, 민원기, 원다해, 이우창, 정태동, Sail Chun대한진단검사의학회 · Annals of Laboratory Medicine2015 · 제35권 제2호14종의 생화학적 항목에 대한 진공 채혈관 신속 응고 혈청 분리관 및 혈장분리관 간의 비교Choi Ju Hee, Ha Changhee, Han Hye Young, Kim Hyun-Joong, Kim Sung Hea, Lee Yeon Jae, Seo Jong Do, Son Eun Hwa, Yun Yeo-Min대한진단검사의학회 · Laboratory Medicine Online2025 · 제15권 제3호검사용기 재질과 저장시간에 따른 일반화학 검사의 비교김각현, 신옥현, 이은주대한임상검사과학회 · 대한임상검사과학회지2001 · 제33권 제2호49종의 검사에 대한 새로운 두 종류의 플라스틱 진공채혈관(Sekisui INSEPACK및 녹십자 Green Vac-Tube)과 BD Vacutainer 진공채혈관의 비교기창석, 오수현대한진단검사의학회 · Annals of Laboratory Medicine2007 · 제27권 제1호새로운 국산 V-Tube 진공채혈관과 BD Vacutainer의 비교평가조덕, 기승정, 김수현, 서순팔, 신명근, 신종희, 양동욱, 원용관, 원은정, 장민중대한진단검사의학회 · Laboratory Medicine Online2013 · 제3권 제3호

이 간행물의 다른 논문 (16건)