Comparison of HbA1c Analyzers: D-10, Variant II Turbo, Cobas Integra 800, and Afinion AS100

1

조회

0

다운로드

1

인용

0

참고문헌

키워드

주제분류병리학

초록

Background : The purpose of this study was to evaluate the performance and agreement among HbA1c values measured using selected analyzers certified by the National Glycohemoglobin Standardization Program (NGSP) and standardized by the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC). Methods : HbA1c determined using D-10 (Bio-Rad, USA), Variant II Turbo (Turbo; Bio-Rad, USA), Cobas Integra 800 (Integra; Roche, Switzerland) and Afinion AS100 (Afinion; Axis-Shield, Norway) were compared with each other. Precision and method comparisons with Deming regression were evaluated according to CLSI recommendations. We also compared the HbA1c values obtained with each analyzer using either IFCC or NGSP methods by correlation analysis and kappa statistics. Results : The repeatability and method/device precisions of D-10 and Afinion were acceptable. The correlation coefficients of HbA1c were 0.986 for D-10 vs. Afinion, 0.997 for D-10 vs. Turbo, 0.988 for D-10 vs. Integra, and 0.991 for Integra vs. Afinion. The average biases of HbA1c Afinion (IFCC)and HbA1c Integra (IFCC) against HbA1c D-10 (NGSP) were -1.90% and -1.79%, respectively. Kappa agreement statistics for the three diabetic control group HbA1c values of “less than 6.5%,” “6.5%-7.5%,” and “greater than 7.5%” for D-10 vs. Turbo, D-10 vs. Integra, and D-10 vs. Afinion were 0.872, 0.836, and 0.833, respectively. Conclusions : The strong correlations and good clinical agreements of HbA1c between each analyzer expressed in terms of either NGSP or IFCC-derived NGSP indicate that these analyzers can be used interchangeably. (Korean J Lab Med 2010;30:345-50)

Abstract

Background : The purpose of this study was to evaluate the performance and agreement among HbA1c values measured using selected analyzers certified by the National Glycohemoglobin Standardization Program (NGSP) and standardized by the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC). Methods : HbA1c determined using D-10 (Bio-Rad, USA), Variant II Turbo (Turbo; Bio-Rad, USA), Cobas Integra 800 (Integra; Roche, Switzerland) and Afinion AS100 (Afinion; Axis-Shield, Norway) were compared with each other. Precision and method comparisons with Deming regression were evaluated according to CLSI recommendations. We also compared the HbA1c values obtained with each analyzer using either IFCC or NGSP methods by correlation analysis and kappa statistics. Results : The repeatability and method/device precisions of D-10 and Afinion were acceptable. The correlation coefficients of HbA1c were 0.986 for D-10 vs. Afinion, 0.997 for D-10 vs. Turbo, 0.988 for D-10 vs. Integra, and 0.991 for Integra vs. Afinion. The average biases of HbA1c Afinion (IFCC)and HbA1c Integra (IFCC) against HbA1c D-10 (NGSP) were -1.90% and -1.79%, respectively. Kappa agreement statistics for the three diabetic control group HbA1c values of “less than 6.5%,” “6.5%-7.5%,” and “greater than 7.5%” for D-10 vs. Turbo, D-10 vs. Integra, and D-10 vs. Afinion were 0.872, 0.836, and 0.833, respectively. Conclusions : The strong correlations and good clinical agreements of HbA1c between each analyzer expressed in terms of either NGSP or IFCC-derived NGSP indicate that these analyzers can be used interchangeably. (Korean J Lab Med 2010;30:345-50)

참고문헌 (22)

  1. Nathan DM, "The clinical information value of the glycosylated hemoglobin assay", N Engl J Med, 310, pp.341~346 (1984)
  2. Benjamin RJ, "Glycated protein update: implications of recent studies, including the diabetes control and complications trial", Clin Chem, 40, pp.683~687 (1994)
  3. Steffes M, "Hemoglobin A1c measurements over nearly two decades:sustaining comparable values throughout the Diabetes Control and Complications Trial and the Epidemiology of Diabetes Interventions and Complications study", Clin Chem, 51, pp.753~758 (2005)
  4. Little RR, "NGSP Steering Committee. The national glycohemoglobin standardization program: a five-year progress report", Clin Chem, 47, pp.1985~1992 (2001)
  5. Jeppsson JO, "Approved IFCC reference method for the measurement of HbA1c in human blood", Clin Chem Lab Med, 40, pp.78~89 (2002)
  6. Kobold U, "Candidate reference methods for hemoglobin A1c based on peptide mapping", Clin Chem, 43, pp.1944~1951 (1997)
  7. Hoelzel W, "IFCC reference system for measurement of hemoglobin A1c in human blood and the national standardization schemes in the United States, Japan, and Sweden: a method-comparison study", Clin Chem, 50, pp.166~174 (2004)
  8. Sacks DB, "Global harmonization of hemoglobin A1c", Clin Chem, 51, pp.681~683 (2005)
  9. Penttilä IM, "Best use of the recommended IFCC reference method, material and values in HbA1C analyses", Scand J Clin Lab Invest, 65, pp.453~462 (2005)
  10. Mosca A, "Global standardization of glycated hemoglobin measurement:the position of the IFCC Working Group", Clin Chem Lab Med, 45, pp.1077~1080 (2007)
  11. Consensus Committee, "Consensus statement on the worldwide standardization of the hemoglobin A1c measurement: the American Diabetes Association, European Association for the Study of Diabetes, International Federation of Clinical Chemistry and Laboratory Medicine, and the International Diabetes Federation", Diabetes Care, 30, pp.2399~2400 (2007)
  12. Marzullo C, "Evaluation of D10 hemoglobin testing system for hemoglobin A1c assay", Ann Biol Clin, 66, pp.95~99 (2008)
  13. Leon-Justel A, "Influence of reduction in the elution times on HPLC glycohaemoglobin results", Clin Biochem, 42, pp.1582~1584 (2009)
  14. Fleming JK, "Evaluation of HbA1c on the Roche COBAS Integra 800 closed tube system", Clin Biochem, 40, pp.822~827 (2007)
  15. Tholen DW, "Evaluation precision performance of quantitative measurement method: approved guideline. CLSI document EP5-A2. 2nd ed" (2004), NCCLS
  16. Krouwer JS, "Method comparison and bias estimation using patient samples: approved guideline. CLSI document EP9-A2. 2nd ed" (2002), NCCLS
  17. 송정한, "대사질환검사 신빙도조사 결과보고(2008)", 임상검사와 정도관리, 31, pp.143~159 (2009)
  18. Goldstein DE, "Harmonization of gylcohemoglobin measurement. CLSI document C44-A" (2002), NCCLS
  19. Miedema K, "Standardization of HbA1c and optimal range of monitoring", Scand J Clin Lab Invest Suppl, 240, pp.61~72 (2005)
  20. Dhatt GS, "HbA1c: a comparison of NGSP with IFCC transformed values", Clin Chim Acta, 358, pp.81~86 (2005)
  21. Paisooksantivatana K, "NGSP and IFCC-derived NGSP HbA1c can be used interchangeably", Diabetes Res Clin Pract, 85, pp.e22~e25 (2009)
  22. Nordin G, "Recommendation for term and measurement unit for “HbA1c”", Clin Chem Lab Med, 45, pp.1081~1082 (2007)

관련 논문AI추천

이 간행물의 다른 논문 (17건)