Performance of an Automated Fluorescence Antinuclear Antibody Image Analyzer

2

조회

0

다운로드

1

인용

0

참고문헌

키워드

주제분류병리학

초록

Background: The gold standard for antinuclear antibody (ANA) screening is the indirect immunofluorescence (IIF) assay with human epithelial cells (HEp-2). However, a number of substantial disadvantages of manual IIF assays have highlighted the need for the automation and standardization of fluorescent ANA (FANA) testing. We evaluated the performance of EUROPattern Suite (Euroimmun AG, Germany), an automated FANA image analyzer, with regard to ANA detection and pattern recognition compared with conventional manual interpretation using the fluorescence microscopic IIF assay. Methods: A total of 104 samples including 70 ANA-positive sera and 34 ANA-negative sera collected from September to October 2015 were included. The sensitivity, specificity, and pattern recognition function were evaluated to determine the performance of EUROPattern Suite compared with the manual IIF assay results. Results: The sensitivity and specificity of EUROPattern Suite for ANA detection were 94.3% and 94.1%, respectively. The concordance rate between the two methods was 94.2%. For pattern recognition, 45.7% of the samples were assigned identical ANA patterns including simple and mixed. When major pattern matching was considered, 83.7% (41/49) and 95.2% (20/21) of the samples with simple and mixed patterns, respectively, showed concordant results between the two methods. Conclusions: EUROPattern Suite, an automated FANA image analyzer, provides a viable option for distinguishing between positive and negative results, although the ability to assign specific patterns is insufficient to replace manual microscopic interpretation. This automated system may increase efficiency in laboratories, in which a large number of samples need to be processed.

Abstract

Background: The gold standard for antinuclear antibody (ANA) screening is the indirect immunofluorescence (IIF) assay with human epithelial cells (HEp-2). However, a number of substantial disadvantages of manual IIF assays have highlighted the need for the automation and standardization of fluorescent ANA (FANA) testing. We evaluated the performance of EUROPattern Suite (Euroimmun AG, Germany), an automated FANA image analyzer, with regard to ANA detection and pattern recognition compared with conventional manual interpretation using the fluorescence microscopic IIF assay. Methods: A total of 104 samples including 70 ANA-positive sera and 34 ANA-negative sera collected from September to October 2015 were included. The sensitivity, specificity, and pattern recognition function were evaluated to determine the performance of EUROPattern Suite compared with the manual IIF assay results. Results: The sensitivity and specificity of EUROPattern Suite for ANA detection were 94.3% and 94.1%, respectively. The concordance rate between the two methods was 94.2%. For pattern recognition, 45.7% of the samples were assigned identical ANA patterns including simple and mixed. When major pattern matching was considered, 83.7% (41/49) and 95.2% (20/21) of the samples with simple and mixed patterns, respectively, showed concordant results between the two methods. Conclusions: EUROPattern Suite, an automated FANA image analyzer, provides a viable option for distinguishing between positive and negative results, although the ability to assign specific patterns is insufficient to replace manual microscopic interpretation. This automated system may increase efficiency in laboratories, in which a large number of samples need to be processed.

참고문헌 (18)

  1. Voigt J, "Automated indirect immunofluorescence evaluation of antinuclear autoantibodies on HEp-2 cells", Clin Dev Immunol, 2012, pp.651058~ (2012)
  2. Meroni PL, "ANA screening: an old test with new recommendations", Ann Rheum Dis, 69, pp.1420~1422 (2010)
  3. Roggenbuck D, "Standardization of automated interpretation of immunofluorescence tests", Clin Chim Acta, 421, pp.168~169 (2013)
  4. Bizzaro N, "Automated antinuclear immunofluorescence antibody screening: a comparative study of six computer-aided diagnostic systems", Autoimmun Rev, 13, pp.292~298 (2014)
  5. Schouwers S, "Value-added reporting of antinuclear antibody testing by automated indirect immunofluorescence analysis", Clin Chem Lab Med, 52, pp.547~551 (2014)
  6. Krause C, "EUROPattern Suite technology for computer-aided immunofluorescence microscopy in autoantibody diagnostics", Lupus, 24, pp.516~529 (2015)
  7. Gerlach S, "Automated Evaluation of Crithidia luciliae Based Indirect Immunofluorescence Tests: A Novel Application of the EUROPattern-Suite Technology", J Immunol Res, 2015, pp.742402~ (2015)
  8. Szalat R, "Anti-NuMA1 and anti-NuMA2 (anti-HsEg5) antibodies: Clinical and immunological features: A propos of 40 new cases and review of the literature", Autoimmun Rev, 9, pp.652~656 (2010)
  9. 김지명, "전신 자가면역질환에서 Line Immunoassay를 이용한 항 ENA항체 및 항 dsDNA항체의 검출", Annals of Laboratory Medicine, 28(5), pp.353~361 (2008), 대한진단검사의학회
  10. Rothberger H, "Disease distributions in patients with multiple patterns of nuclear staining detected by FANA (immunofluorescent antinuclear antibody) tests", Clin Exp Rheumatol, 4, pp.201~207 (1986)
  11. Meroni PL, "Automated tests of ANA immunofluorescence as throughput autoantibody detection technology: strengths and limitations", BMC Med, 12, pp.38~ (2014)
  12. Ganapathy V, "Autoimmunity to the nuclear autoantigen DFS70 (LEDGF): what exactly are the autoantibodies trying to tell us?", Arthritis Rheum, 50, pp.684~688 (2004)
  13. Mahler M, "Importance of the dense fine speckled pattern on HEp-2 cells and anti-DFS70 antibodies for the diagnosis of systemic autoimmune diseases", Autoimmun Rev, 11, pp.642~645 (2012)
  14. Mahler M, "The clinical significance of the dense fine speckled immunofluorescence pattern on HEp-2 cells for the diagnosis of systemic autoimmune diseases", Clin Dev Immunol, 2012, pp.494356~ (2012)
  15. Chan EK, "Report of the First International Consensus on Standardized Nomenclature of Antinuclear Antibody HEp-2 Cell Patterns 2014-2015", Front Immunol, 6, pp.412~ (2015)
  16. Damoiseaux J, "International consensus on ANA patterns (ICAP): the bumpy road towards a consensus on reporting ANA results", Auto Immun Highlights, 7, pp.1~ (2016)
  17. Mulliez SM, "Impact of the routine implementation of automated indirect immunofluorescence antinuclear antibody analysis: 1 year of experience", Clin Chem Lab Med, 54, pp.183~186 (2016)
  18. Wiliem A, "A bag of cells approach for antinuclear antibodies HEp-2 image classification", Cytometry A, 87, pp.549~557 (2015)

관련 논문AI추천

이 간행물의 다른 논문 (17건)