고압동결고정을 이용한 애기장대 줄기 의 cryo-SEM 분석법
Cryo-SEM Methodology of Arabidopsis thaliana Stem Using High-Pressure Freezing
1
조회
0
다운로드
1
인용
0
참고문헌
키워드
초록
The scanning electron microscopy is an ideal technique for examining plant surface at high resolution. Most hydrate samples, however, must be fix and dehydrate for observation in the scanning electron microscope. Because the microscopes operate under high vacuum, most specimens, especially biological samples,cannot withstand water removal by the vacuum system without morphological distortion. Cryo-techniques can observe in their original morphology and structure without various artifacts from conventional sample preparation. Rapid cooling is the method of choice for preparing plant samples for scanning electron microscopy in a defined physiological state. As one of cryo-technique, highpressure freezing allows for fixation of native non-pretreated samples up to 200 μm thick and 2 mm wide with minimal or no ice crystal damage for the freezing procedure. In this study, we could design to optimize structural preservation and imaging by comparing cryo-SEM and convention SEM preparation, and observe a fine,well preserved Arabidopsis stem’s inner ultrastructure using HPF and cryo-SEM. These results would suggest a useful method of cryo-preparation and cryo-SEM for plant tissues, especially intratubule and vacuole rich structure.
Abstract
The scanning electron microscopy is an ideal technique for examining plant surface at high resolution. Most hydrate samples, however, must be fix and dehydrate for observation in the scanning electron microscope. Because the microscopes operate under high vacuum, most specimens, especially biological samples,cannot withstand water removal by the vacuum system without morphological distortion. Cryo-techniques can observe in their original morphology and structure without various artifacts from conventional sample preparation. Rapid cooling is the method of choice for preparing plant samples for scanning electron microscopy in a defined physiological state. As one of cryo-technique, highpressure freezing allows for fixation of native non-pretreated samples up to 200 μm thick and 2 mm wide with minimal or no ice crystal damage for the freezing procedure. In this study, we could design to optimize structural preservation and imaging by comparing cryo-SEM and convention SEM preparation, and observe a fine,well preserved Arabidopsis stem’s inner ultrastructure using HPF and cryo-SEM. These results would suggest a useful method of cryo-preparation and cryo-SEM for plant tissues, especially intratubule and vacuole rich structure.
참고문헌 (6)
- Beckett ND, "Low-temperature scanning electron microscopy. Ultrastructural Techniques for Microorganisms", pp.45~86 (1986), Plenum Publishing Corporation
- Boyde A, "Biological specimen preparation for the scanning electron microscope: an overview", Proceedings of the Fifth Annual Electron Microscopy Symposium (1972)
- Echlin P, "The preparation and X-ray microanalysis of bulk frozen hydrated vacuolated plant tissue", J Microsc, 141, pp.329~348 (1986)
- 이은주, "Cryo-Methods와 Cryo-SEM을 이용한 Cyanobacteria, Synechocystis sp. PCC 6803 미세구조 관찰", 한국현미경학회지, 39(1), pp.65~72 (2009), 한국현미경학회
- Sargent JA, "Low temperature scanning electron microscopy: advantages and applications", Scanning Microsc, 2, pp.835~849 (1988)
- Shimoni E, "On optimizing high-pressure freezing: from heat transfer theory to a new microbiopsy device", J Microsc, 192, pp.236~247 (1998)