본문 바로가기
KERD
논문KCI 미등재

Comparative study on the specimen thickness measurement using EELS and CBED methods

1

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류자연과학일반

초록

Two thickness measurement methods using an electron energy loss spectroscopy (EELS) and 10a convergent beam electron diffraction (CBED) were compared in an Fe-18Mn-0.7C alloy. The thin foil specimen was firstly tilted to satisfy 10a two-beam condition. Low loss spectra of EELS and CBED patterns were acquired in scanning transmission electron microscopy (STEM) and TEM-CBED modes under the two-beam condition. The log-ratio method was used for measuring the thin foil thickness. Kossel-Möllenstedt (K-M) fringe of the 131 diffracted disk of austenite was analyzed to evaluate the thickness. The results prove the good coherency between both methods in the thickness range of 72 ~ 113 nm with a difference of less than 5%.

Abstract

Two thickness measurement methods using an electron energy loss spectroscopy (EELS) and 10a convergent beam electron diffraction (CBED) were compared in an Fe-18Mn-0.7C alloy. The thin foil specimen was firstly tilted to satisfy 10a two-beam condition. Low loss spectra of EELS and CBED patterns were acquired in scanning transmission electron microscopy (STEM) and TEM-CBED modes under the two-beam condition. The log-ratio method was used for measuring the thin foil thickness. Kossel-Möllenstedt (K-M) fringe of the 131 diffracted disk of austenite was analyzed to evaluate the thickness. The results prove the good coherency between both methods in the thickness range of 72 ~ 113 nm with a difference of less than 5%.

참고문헌 (13)

  1. S. M. Allen, "Foil thickness measurements from convergent-beam diffraction patterns", Phil. Mag. A, 43, pp.325~335 (1981)
  2. J. E. Bailey, "The dislocation distribution, flow stress, and stored energy in cold-worked polycrystalline silver", Phil. Mag, 5, pp.485~497 (1960)
  3. F. Delmas, "Quantitative TEM study of the precipitation microstructure in aluminium alloy Al (MgSiCu) 6056 T6", Mater. Sci. Eng. A, 373, pp.80~89 (2004)
  4. T. Dorin, "Size distribution and volume fraction of T1 phase precipitates from TEM images: Direct measurements and related correction", Micron, 78, pp.19~27 (2015)
  5. R. F. Egerton, "Electron Energy-Loss Spectroscopy in the Electron Microscope" (1996), Plenum press
  6. R. F. Egerton, "Measurement of local thickness by electron energy-loss spectroscopy", Ultramicroscopy, 21, pp.231~224 (1987)
  7. P. Hirsch, "Electron Microscopy of Thin Crystals" (1977), Robert E. Krieger Pub. Comp
  8. P. M. Kelly, "The determination of foil thickness by scanning transmission electron microscopy", Phys. Status Solidi, 31, pp.771~780 (1975)
  9. L.E. Murr, "Electron Optical Applications in Materials Science" (1970), McGraw-Hill Book Comp
  10. Z. Pan, "Measurement of foil thickness in transmission electron microscopy", J. Mater. Sci, 29, pp.1920~1924 (1994)
  11. E. E. Underwood, "Quantitative stereology" (1970), Addison-Wesley Publishing Comp
  12. D.B. Willams, "Transmission Electron Microscopy" (2009), Springer
  13. Z. Yang, "Analytical TEM investigations on concentration gradients surrounding Ni4Ti3 precipitates in Ni-Ti shape memory material", Scripta Mater, 52, pp.1129~1134 (2005)