Two thickness measurement methods using an electron energy loss spectroscopy (EELS) and 10a convergent beam electron diffraction (CBED) were compared in an Fe-18Mn-0.7C alloy. The thin foil specimen was firstly tilted to satisfy 10a two-beam condition. Low loss spectra of EELS and CBED patterns were acquired in scanning transmission electron microscopy (STEM) and TEM-CBED modes under the two-beam condition. The log-ratio method was used for measuring the thin foil thickness. Kossel-Möllenstedt (K-M) fringe of the 131 diffracted disk of austenite was analyzed to evaluate the thickness. The results prove the good coherency between both methods in the thickness range of 72 ~ 113 nm with a difference of less than 5%.
Abstract
Two thickness measurement methods using an electron energy loss spectroscopy (EELS) and 10a convergent beam electron diffraction (CBED) were compared in an Fe-18Mn-0.7C alloy. The thin foil specimen was firstly tilted to satisfy 10a two-beam condition. Low loss spectra of EELS and CBED patterns were acquired in scanning transmission electron microscopy (STEM) and TEM-CBED modes under the two-beam condition. The log-ratio method was used for measuring the thin foil thickness. Kossel-Möllenstedt (K-M) fringe of the 131 diffracted disk of austenite was analyzed to evaluate the thickness. The results prove the good coherency between both methods in the thickness range of 72 ~ 113 nm with a difference of less than 5%.
참고문헌 (13)
- S. M. Allen, "Foil thickness measurements from convergent-beam diffraction patterns", Phil. Mag. A, 43, pp.325~335 (1981)
- J. E. Bailey, "The dislocation distribution, flow stress, and stored energy in cold-worked polycrystalline silver", Phil. Mag, 5, pp.485~497 (1960)
- F. Delmas, "Quantitative TEM study of the precipitation microstructure in aluminium alloy Al (MgSiCu) 6056 T6", Mater. Sci. Eng. A, 373, pp.80~89 (2004)
- T. Dorin, "Size distribution and volume fraction of T1 phase precipitates from TEM images: Direct measurements and related correction", Micron, 78, pp.19~27 (2015)
- R. F. Egerton, "Electron Energy-Loss Spectroscopy in the Electron Microscope" (1996), Plenum press
- R. F. Egerton, "Measurement of local thickness by electron energy-loss spectroscopy", Ultramicroscopy, 21, pp.231~224 (1987)
- P. Hirsch, "Electron Microscopy of Thin Crystals" (1977), Robert E. Krieger Pub. Comp
- P. M. Kelly, "The determination of foil thickness by scanning transmission electron microscopy", Phys. Status Solidi, 31, pp.771~780 (1975)
- L.E. Murr, "Electron Optical Applications in Materials Science" (1970), McGraw-Hill Book Comp
- Z. Pan, "Measurement of foil thickness in transmission electron microscopy", J. Mater. Sci, 29, pp.1920~1924 (1994)
- E. E. Underwood, "Quantitative stereology" (1970), Addison-Wesley Publishing Comp
- D.B. Willams, "Transmission Electron Microscopy" (2009), Springer
- Z. Yang, "Analytical TEM investigations on concentration gradients surrounding Ni4Ti3 precipitates in Ni-Ti shape memory material", Scripta Mater, 52, pp.1129~1134 (2005)