수술용 로봇 햅틱 시스템 구성을 위한 저관성 마스터 장치 핸들부 개발
Development of Low-inertia Master device for Haptic system configuration of surgical robot
0
조회
0
다운로드
5
인용
0
참고문헌
키워드
초록
When using commercialized robot assisted laparoscopic, surgeon has met some problems to depend only on image of the surgical field. To solve it, there were various researches. The previous study showed that it is possible to estimate the operation force on the commercialized instrument inside patient without sensors. To apply the estimated force to a haptic master console for the laparoscopy surgical robot system, the light haptic master console should be designed. This paper suggests the design of lighter master console handle to reduce a weight of the console whose structure can match with the joint and DOF of an instrument. A cable-conduit mechanism is designed to make light structure to perform a delicate manipulation. The cable-conduit mechanism removes the weight and inertia of link caused by haptic actuator and encoder which is separated from handle link of a manipulator.
Abstract
When using commercialized robot assisted laparoscopic, surgeon has met some problems to depend only on image of the surgical field. To solve it, there were various researches. The previous study showed that it is possible to estimate the operation force on the commercialized instrument inside patient without sensors. To apply the estimated force to a haptic master console for the laparoscopy surgical robot system, the light haptic master console should be designed. This paper suggests the design of lighter master console handle to reduce a weight of the console whose structure can match with the joint and DOF of an instrument. A cable-conduit mechanism is designed to make light structure to perform a delicate manipulation. The cable-conduit mechanism removes the weight and inertia of link caused by haptic actuator and encoder which is separated from handle link of a manipulator.
참고문헌 (19)
- A. Ayav, "Early results of one-year robotic surgery usingthe Da Vinci system to perform advancedlaparocopic procedures", J Gastrointest Surg, 8(6), pp.720~726 (2004)
- M. C. Lee, "Reaction Force Estimation of Surgical Robot Instrument Using Perturbation Observer with SMCSPO Algorithm", Proceedings of AIM 2010, pp.181~186 (2010)
- H. S. Song, "Design of Dexterous Manipulator for MIS", Proceeding of KSME summer symposium, pp.823~828 (2004)
- F. Corcione, "Advantages and limits of robot-assisted laparoscopic surgery", Surgical Endoscopy, 19(1), pp.117~119 (2002)
- P. Giulianotti, "Robotics in General Surgery- Personal Experience in a Large Community Hospital", Archives of Surgery, (238,), pp.777~784 (2003)
- K. Vlachos, "Design and Implementation of a haptic Device for Training in Urological Operations", IEEE Transactions on Robotics and Automation, 19, pp.801~809 (2003)
- S. Payandeh, "Towards new designs of haptic devices for minimally invasive surgery", International Congress Series, pp.775~781 (2003)
- V. Hayward, "Freedom-7: A High Fidelity Seven Axis haptic Deviec with Application to Surgical Training", Experimental Robotics Vi, pp.445~456 (1998)
- A. J. Costello, "Installation of telerobotic surgery and initial experience with telerobotic radical prostatectomy", BJU international, 96(1), pp.34~38 (2005)
- S. M. Yoon, "Evaluation of Possibility of Estimation of Reaction Force of Surgical Robot Instrument using Sliding Perturbation Observer", KROS, 7(1), pp.20~28 (2012)
- Steven E. Butner, "Transformaing a Surgical Robot for Human Telesurgery", IEEE Trans on Robotics and Automation, 19(5), pp.818~824 (2003)
- Gyung Tak Sung, "Robotic Laparoscopic Surgery : A Comparisonof the da Vinci and ZEUS Systems", UROLOGY, 58(6), pp.893~898 (2001)
- J. J. Crisco, "The mechanical Axes of the Wrist Are Oriented Obliquely to the Anatomical Axes", The Journal of Bone & Joint Surgery, 93, pp.169~177 (2011)
- V. Agrawal, "Modeling of Transmission Characteristics Acrossa Cable-Conduit System", IEEE Trans onRobotics, 36(5), pp.914~924 (2010)
- B. Y. Chung, "Ergonics Human Factors" (2009), Minyoungsa
- Varun Agrawal, "Control of Cable Actuated Devices using Smooth Backlash Inverse", IEEE International Conference on Robotics and Automation, pp.1074~1079 (2010)
- Varun Agrawal, "Modeling of Closed Loof Cable-ConduitTransmission System", IEEE InternationalConference on Robotics and Automation (2008)