인터벤션에서 자동노출제어장치와 수동노출 사용 시 두께 변화에 따른 선량감소 방안 연구

A Study on Dose Reduction Method according to Slice Thickness Change using Automatic Exposure Controller and Manual Exposure in Intervention

1

조회

0

다운로드

5

인용

0

참고문헌

키워드

주제분류방사선과학

초록

We aims to perform comparative analysis on the dose area and image qualities varying on the slice thickness when using Automatic Exposure Controller (AEC) and manual exposure; thus, it wants to suggest a measure to reduce exposure dose by setting the optimal examination condition for each slice thickness. The method was to set the thickness as Thin, Normal, and Heavy adult and evaluate the dose area, spatial resolution, low contrast resolution, Signal to Noise Ratio (SNR) and Contrast to Noise Ratio (CNR) according to each slice thickness by using the AEC and the manual exposure controller. The dose area according to each slice thickness all increased both when using the AEC and the manual exposure. However, the manual exposure showed lower dose area product than the AEC. Spatial resolutions and low contrast resolutions were all observed to be higher than the evaluation standard. Also, the SNR and CNR of each thickness all increased when using the AEC. When using the manual exposure, SNR and CNR increased in all cases other than the Heavy Adult. Consequently, the Thin and Normal Adult showed dose reduction about 2 times when using the manual exposure controller, while ensuring the image quality. Heavy adult was able to maintain good image quality by using AEC.

Abstract

We aims to perform comparative analysis on the dose area and image qualities varying on the slice thickness when using Automatic Exposure Controller (AEC) and manual exposure; thus, it wants to suggest a measure to reduce exposure dose by setting the optimal examination condition for each slice thickness. The method was to set the thickness as Thin, Normal, and Heavy adult and evaluate the dose area, spatial resolution, low contrast resolution, Signal to Noise Ratio (SNR) and Contrast to Noise Ratio (CNR) according to each slice thickness by using the AEC and the manual exposure controller. The dose area according to each slice thickness all increased both when using the AEC and the manual exposure. However, the manual exposure showed lower dose area product than the AEC. Spatial resolutions and low contrast resolutions were all observed to be higher than the evaluation standard. Also, the SNR and CNR of each thickness all increased when using the AEC. When using the manual exposure, SNR and CNR increased in all cases other than the Heavy Adult. Consequently, the Thin and Normal Adult showed dose reduction about 2 times when using the manual exposure controller, while ensuring the image quality. Heavy adult was able to maintain good image quality by using AEC.

참고문헌 (24)

  1. Kalra MK, "Whole Spine CT for Evaluation of Scoliosis in Children: Feasibility of Sub-milli Sievert Scanning Protocol", Acta Radiologica, 54(2), pp.226~230 (2013)[DOI]
  2. Dazani GB, "Percutaneous Coronary Intervention for Chronic Total Occlusions", The American Journal of Cardiology, 112(12), pp.1849~1853 (2013)[DOI]
  3. Gorostidi M, "Technique for Precaval and Laterocaval Nodes Excision at Extraperitoneal Paraaortic Lymphadenectomy", Gynecologic Oncology, 148(1), pp.233~234 (2018)[DOI]
  4. Matsunaga Y, "Patient Exposure during Plain Radiography and Mammography in Japen in 1974-2014", Radiation Protection Dosimetry, 176(4), pp.347~353 (2017)[DOI]
  5. International Commission on Radiological Protection (ICRP), "Recommendations of the International Commission on Radiological Protection. ICRP Publication 60" (1990)
  6. International Commission on Radiological Protection (ICRP), "The 2007 Recommendations of the International Commission on Radiological Protection. ICRP Publication 103" (2007)
  7. Crowhurst JA, "Radiation Dose in Coronary Angiography and Intervention: Initial Result from the Establishment of a Multi Center Diagnostic Reference Level in Queensland Public Hospital", Journal of Medical Radiation Sciences, 61(3), pp.135~141 (2014)[DOI]
  8. Morris GM, "Patient Radiation Dose during Fluoroscopically Guided Biventricular Device Implantation", Acta Cardiology, 69(5), pp.491~495 (2014)[DOI]
  9. Coles DR, "Comparison of Radiation Doses From Multislice Computerized Tomography Coronary Angiography and Conventional Diagnostic Angiography", Journal of the American College of Cardiology, 47(9), pp.1840~1845 (2006)[DOI]
  10. International Commission on Radiological Protection (ICRP), "Avoidance of Radiation Injuries from Medical Interventional Procedures. ICRP Publication 85" (2000)
  11. International Commission on Radiological Protection (ICRP), "Diagnostic Reference Levels in Medical Imaging. ICRP Publication 135" (2017)
  12. 황준호, "인터벤션 시 방사선조사 조건에 따른 선량감소 : 면적선량과 영상화질 변화를 중심으로", 방사선기술과학, 40(3), pp.393~400 (2017), 대한방사선과학회[DOI]
  13. Spink C, "Noise Reduction Angiographic Imaging Technology Reduces Radiation Dose during Bronchial Artery Embolization", European Journal of Radiology, 97, pp.115~118 (2017)[DOI]
  14. Doyle P, "Calibrating Automatic Exposure Control Devices for Digital Radiography", Physics in Medicine and Biology, 51(21), pp.5475~5485 (2006)[DOI]
  15. Moore CS, "Automatic Exposure Control Calibration and Optimization for Abdomen, Pelvis and Lumbar spine Imaging with an Agfa Computed Radiography System", Physics in Medicine and Biology, 61(21), pp.N551~N564 (2016)[DOI]
  16. 황준호, "DR(Digital Radiography)에서 자동노출제어장치의 감도변화에 따른 검출기 성능평가 시 흡수선량과 히스토그램을 통한정량적 분석방법에 관한 연구", 한국콘텐츠학회 논문지, 18(1), pp.232~240 (2018), 한국콘텐츠학회[DOI]
  17. 강병삼, "인터벤션 시술 시 면적선량계를 이용한 환자 방사선 선량 평가", 방사선기술과학, 40(1), pp.27~34 (2017), 대한방사선과학회[DOI]
  18. Oliveira da Silva MV, "Comparing Measured and Calculated Doses in Interventional Cardiology Procedures", Radiation Protection Dosimetry, 176(4), pp.439~443 (2017)[DOI]
  19. Shepard SJ, "An Exposure Indicator for Digital Radiography: AAPM Task Group 116 (Executive Summary)", Journal of Medical Physics, 36(7), pp.2898~2914 (2009)[DOI]
  20. Sebert JA, "The Standardized Exposure Index for Digital Radiography: an Opportunity for Optimization of Radiation Dose to the Pediatric Population", Journal of Pediatric Radiology, 41(5), pp.573~581 (2011)[DOI]
  21. 홍선숙, "Digital Radiography System에서 X선 광자 검출 방식에 따른 선량 및 화질 특성에 관한 연구", 전자공학회논문지, 50(12), pp.247~253 (2013), 대한전자공학회
  22. Uffmann M, "Digital radiography:The Balance Between Image and Required Radiation Dose", European Journal Radiology, 72(2), pp.202~208 (2009)[DOI]
  23. Ionita CN, "Modification of the NEMA XR21-2000 Cardiac Phantom for Testing of Imaging Systems used in Endovascular Image Guided Interventions", Physics of Medical Imaging, 7258, pp.72584R~72589R (2009)
  24. 강병삼, "IVR장치의 성능 평가 기준 개발", 한국콘텐츠학회 논문지, 11(1), pp.236~244 (2011), 한국콘텐츠학회

이 간행물의 다른 논문 (10건)