진단용엑스선촬영장치의 간접 선량 계산법 개발
Development of Indirect Dosimetry by Calculation Method in the Diagnostic X-ray Equipment
0
조회
0
다운로드
2
인용
0
참고문헌
키워드
초록
The purpose of this study was to develop the indirect dosimetry by calculation (IDC) method for diagnostic X-ray equipment. The experiments were performed with inverter type X-ray tubes: Toshiba (E7252X, Japan) and Varian (RAD-14, USA). For the development method, we first applied the standard quality of X-ray beam shown in the TRS457 document, and second, to produce the constants of trendline for the IDC, the total filtration on X-ray beam was subdivided. Third, in order to increase the precision, the energy region was divided into the high energy region and the low energy region and developed by the IDC. In order to verify the IDC, mean dose (mR) values were measured for three Toshiba X-ray tubes and three Varian X-ray tubes at clinical medical institutions and then compared with the IDC on the 2013. As a result, compared with the previous study, the accuracy of the IDC of this study were improved by 2.71% and 9.91% in Toshiba and Varian X-ray tubes, respectively
Abstract
The purpose of this study was to develop the indirect dosimetry by calculation (IDC) method for diagnostic X-ray equipment. The experiments were performed with inverter type X-ray tubes: Toshiba (E7252X, Japan) and Varian (RAD-14, USA). For the development method, we first applied the standard quality of X-ray beam shown in the TRS457 document, and second, to produce the constants of trendline for the IDC, the total filtration on X-ray beam was subdivided. Third, in order to increase the precision, the energy region was divided into the high energy region and the low energy region and developed by the IDC. In order to verify the IDC, mean dose (mR) values were measured for three Toshiba X-ray tubes and three Varian X-ray tubes at clinical medical institutions and then compared with the IDC on the 2013. As a result, compared with the previous study, the accuracy of the IDC of this study were improved by 2.71% and 9.91% in Toshiba and Varian X-ray tubes, respectively
참고문헌 (23)
- Kimme-Smith C, "Effects of reduced exposure on computed radiography: comparison of nodule detection accuracy with conventional and asymmetric screen-film radiographs of a chest phantom", American Journal of Roentgenology, 165(2), pp.269~273 (1995)
- Al Khalifah K, "Comparison between conventional radiography and digital radiography for various kVp and mAs settings using a pelvic phantom", Radiography, 10(2), pp.119~125 (2004)
- Van Metter R, "Medical imaging 2006: physics of medical imaging", pp.6142~ (2006), SPIE
- Seibert JA, "Computed radiography x-ray exposure trends", Acad Radiol, 4, pp.313~318 (1996)
- "IEC 60601-2-54:2009", AMD1:2015, Part 2-54: Particularrequirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy, 2015
- Hanan FA, "Assessment of Entrance skin dose and effective dose of some routine Examinations using calculation technique" (2007), King Saud University
- Suchart K, "An Estimation of X-Radiation Output using Mathematic Model", American Journal of Applied Sciences, 8(9), pp.923~926 (2011)
- National Based Research of Patient Radiation National Based Research of Patient Radiation, "Sung DW", Korea Food & Drug Administration, ~ (2013)
- "IEC 61267:2005, Medical diagnostic X-ray equipment – Radiations conditions for use in the determination of characteristics", ~ (2005)
- "IEC 61674:2012, Medical electrical equipment -Dosimeters with ionization chambers and/or semiconductor detectors as used in X-ray diagnostic imaging" (2012)[URL]
- Seoung YH, "Evaluation of Accuracy and Reproducibility of Clinical General X-ray System by Safety Management Standard of Diagnostic Radiation Generators", J. Health. Med. Sci., Cheongju Univ, 6(2), pp.41~48 (2018)
- Establishing the standard X-ray beam qualities for calibration of dosimeters used in diagnostic radiology following IAEA-TRS457, "Duong VT", ~, pp.No. VINATOM-AR—13 (2014)
- Suchart K, "An Estimation of X-Radiation Output using Mathematic Model", American Journal of Applied Sciences, 8(9), pp.923~926 (2011)
- 노영훈, "촬영 및 투시용 X선 장치의 기본안전과 필수 성능에 관한 개별 기준규격의 동향", 방사선기술과학, 38(4), pp.331~336 (2015), 대한방사선과학회[DOI]
- Technical White Paper: Monitoring and Tracking of Fluoroscopic Dose, , ~, pp.CRCPD Publication #E-10-7 (2010)
- Mori T, "The Calculation and Measurement of an Entrances Surface Dose in Diagnostic X-ray", Committee of Radiation Protection (NDD group) in the Ibaragi branch of Japanese Society of Radiological Technology (1984)
- Edmonds IR, "Calculation of patient skin dose from diagnostic X-ray procedures", The British journal of radiology, 57(680), pp.733~734 (1984)
- Tung CJ, "Evaluations of gonad and fetal doses for diagnostic radiology", Proc Nat Sci Counc Repub China B, 23(3), pp.107~113 (1999)
- Suzuki M, "The Expression for an Entrances Surface Dose of Patients in Diagnostic X-ray–NDD Method", Committee of Radiation Protection in the Ibaragi branch of Japanese Society of Radiological Technology (1996)
- 김성철, "선질계수에 의한 피부입사선량 계산법", 한국콘텐츠학회 논문지, 10(2), pp.258~267 (2010), 한국콘텐츠학회
- 성호진, "피폭선량 산출을 통한 피부입사선량 계산: 머리 및 손목을 중심으로", 방사선기술과학, 39(3), pp.305~312 (2016), 대한방사선과학회[DOI]
- Brosi P, "Copper filtration in pediatric digital X-ray imaging : its impact on image quality and dose", Radiol Phys Technol, 4, pp.148~155 (2011)
- 한재복, "입사 표면 선량 계산에 따른 진단용 X-선 촬영시 피폭선량 비교 연구", 한국콘텐츠학회 논문지, 11(12), pp.357~363 (2011), 한국콘텐츠학회