은(Ag)이온이 함유된 항균필름이 MRI에 미치는 영향
Effect of Antibacterial Film Containing Silver Ions on MRI
0
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
The purpose of these experiments is often to scan infected patients with MRI. Therefore, it is to investigate whether the antibacterial film containing silver ions, which is a non-magnetic substance, affects magnetic resonance imaging. In this experiment, the ACR phantom was used, not the patient. The ACR phantom was wrapped in an antibacterial film and the SNR, CNR, sagittal localization image, and geometrical accuracy were compared before and after. The experiment was performed 10 times and the averaged values were compared. There were no significant differences in the results of all experiments. The FDA recommends removing metal and antibacterial film masks during MRI scans. The reason is that there was one case of injury with facial burns. When I touched the antibacterial film to check the fever during the 2 hour experiment, I did not feel any particular fever. In light of the experimental results, it would be helpful to use an antibacterial film when testing an infected patient. The reason is that there isn't a difference before and after the experiment of SNR, CNR, and sagittal localization images.
Abstract
The purpose of these experiments is often to scan infected patients with MRI. Therefore, it is to investigate whether the antibacterial film containing silver ions, which is a non-magnetic substance, affects magnetic resonance imaging. In this experiment, the ACR phantom was used, not the patient. The ACR phantom was wrapped in an antibacterial film and the SNR, CNR, sagittal localization image, and geometrical accuracy were compared before and after. The experiment was performed 10 times and the averaged values were compared. There were no significant differences in the results of all experiments. The FDA recommends removing metal and antibacterial film masks during MRI scans. The reason is that there was one case of injury with facial burns. When I touched the antibacterial film to check the fever during the 2 hour experiment, I did not feel any particular fever. In light of the experimental results, it would be helpful to use an antibacterial film when testing an infected patient. The reason is that there isn't a difference before and after the experiment of SNR, CNR, and sagittal localization images.
참고문헌 (11)
- Park BJ, "The usefulness of test bolus examination in three dimensional contrast enhanced MR angiography of the carotid artery", J Korean Soc Magn Reson Med, 44, pp.317~323 (2001)
- Brennan SC, "Anxiety and panic during magnetic resonance scans", The Lancet, 332(8609), pp.512~ (1988)
- Tan H, "Estimation of K-space trajectories in spiral MR", Magn Reson Med, 61, pp.396~404 (2009)[DOI]
- Kim YG, "MRI in Practice" (2013), Academia
- Tartaglino LM, "Metallic artifacts on MR images of the postoperative spine : Reduction with fast spin-echo techniques", Radiology, 190(2), pp.565~569 (1994)
- 이정휘, "자기공명영상 장치의 정도관리를 위한 ACR 팬텀의 유용성 및 적정성", 대한영상의학회지, 54(1), pp.47~55 (2006), 대한영상의학회
- "Special Medical Device Quality Control Guide", The Korean Society of Radiology, ~ (2019)
- 신운재, "치아 임플란트와 보철에서 발생하는 자기공명영상의 인공물 감소방안 연구", 한국방사선학회논문지, 13(7), pp.1025~1033 (2019), 한국방사선학회[DOI]
- Lee EK, "Measurement of SNR & CNR in brain MRI lesion", master’s, pp.2016
- Dan MG, "Preparation and characterization of silica nanofibers containing silver nanoparticles", master’s, pp.2006
- "Wear Face Masks with No Metal During MRI", The U.S. Food and Drug Administration (FDA), Exams:FDA Safety Communication Date Issued, ~ (2020)