미세병변 진단에서 Optimized Grid을 사용한 영상과 Grid Supperession Software를 사용한 영상의 비교분석
Comparison of Images Using Optimized Grid and Images Using Grid Supperession Software in the Diagnosis of Micro Lesions
0
조회
0
다운로드
1
인용
0
참고문헌
키워드
초록
Quantitative analysis was performed to confirm that moire artifact was removed without loss of image when using grid suppression software in the diagnosis of micro lesions. we showed that grid suppression images can be morphologically different from original images as they are significantly lower than those of the optimized grid in the similarity analysis with reference images in mammographic phantom images. We were confirmed that images of microcalcification with smaller signal than noise were lost because the pixel values of all lesions increased significantly after the grid suppression, The change in contrast using the NORMI 13 X-ray test phantom was reduced to 30% of the reference image, This result was significantly lower than the 90% when using the optimized grid. In conclusion, the use of grid suppression software in clinical images should be carefully considered because of the possibility of misdiagnosis due to micro lesion loss and morphological changes.
Abstract
Quantitative analysis was performed to confirm that moire artifact was removed without loss of image when using grid suppression software in the diagnosis of micro lesions. we showed that grid suppression images can be morphologically different from original images as they are significantly lower than those of the optimized grid in the similarity analysis with reference images in mammographic phantom images. We were confirmed that images of microcalcification with smaller signal than noise were lost because the pixel values of all lesions increased significantly after the grid suppression, The change in contrast using the NORMI 13 X-ray test phantom was reduced to 30% of the reference image, This result was significantly lower than the 90% when using the optimized grid. In conclusion, the use of grid suppression software in clinical images should be carefully considered because of the possibility of misdiagnosis due to micro lesion loss and morphological changes.
참고문헌 (15)
- Michael Sandborg, "The choice of anti-scatter grids in diagnostic radiology: the optimization of image quality and absorbed dose" (1993)
- J. Wang, "Film digitization aliasing artifacts caused by grid line patterns", IEEE Trans. Medical Imaging, 13(2) (1994)
- 김동식, "디지털 방사선 영상에서 그리드 왜곡의 포화 특성에 관한 연구와 적응 필터링에 기초한 제거", 전자공학회논문지 - SP, 48(4), pp.1~11 (2011), 대한전자공학회
- 김동식, "디지털 방사선 영상에서 그리드 각도의 최적화와 변조 모델에 기초한 그리드 왜곡의 제거", 전자공학회논문지 - SP, 48(3), pp.30~41 (2011), 대한전자공학회
- 이후민, "X선 진단장치의 고압정류방식이 대조도 향상에 미치는 영향", 방사선기술과학, 37(3), pp.187~193 (2014), 대한방사선과학회
- 김동식, "고정 그리드를 사용한 디지털 방사선 영상에서 1차원 필터에 기초한 왜곡의 제거", 전자공학회논문지 - SP, 47(5), pp.117~126 (2010), 대한전자공학회
- D.S. Kim, "Grid artifact reduction for direct digital radiography detectors based on rotated stationary grids with homomorphic filtering", Med. Phys, 40(6), pp.061905~ (2013)[DOI]
- Kato M, "Clinical efficacy of image processing of grid detection and suppression (GDS) in computed radiography", Nihon Hoshasen Gijutsu Gakkai Zasshi, 61(8), pp.1158~1169 (2005)[DOI]
- Ryoji Sasada, "Stationary grid pattern removal using 2D technique for moire-free radiographic image display", Proceedings, Medical Imaging 2003: Visualization, Image-Guided Procedures, and Display, 5029 (2003)
- Chih-Yang Lin, "A Study of Grid Artifacts Formation and Elimination in Computed Radiographic Images", J Digit Imaging, 19(4), pp.351~361 (2006)[DOI]
- Ubeda, Carlos, "Influence of the antiscatter grid on dose and image quality in pediatric interventional cardiology X-ay systems", Catheterization and Cardiovascular Interventions, 82(1), pp.51~57 (2013)[DOI]
- 정성훈, "소아 흉부검사 시 격자 사용에 따른 영상 화질 평가", 방사선기술과학, 40(3), pp.371~376 (2017), 대한방사선과학회[DOI]
- Fritz, Shannon, "Guidelines for anti-scatter grid use in pediatric digital radiography", Pediatric radiology, 44(3), pp.313~321 (2014)[DOI]
- Tapiovaara, "A search for improved technique factors in pediatric fluoroscopy", Physics in medicine and biology, 44(2), pp.537~560 (1999)[DOI]
- 이상호, "Sprague Dawley Rat의 미세병변에서 Moire Artifact를 제거하기 위한 Grid suppression software 사용 후 영상분석", 방사선기술과학, 40(4), pp.575~580 (2017), 대한방사선과학회[DOI]