We aim to evaluate safety of radiation by measuring leakage dose and patient(phantom) incident dose of ZEN-PX II dental portable equipment developed by G company. Measurement for leakage dose of equipment is conducted on the top, at the bottom, on the left, on the right and at the back. Dose measurement incident on the subject with the area dosimeter when using the phantom and measurement the leakage dose of equipment when using the phantom are evaluated. Comparing the right with the highest leakage dose as a 0 cm, 25 cm, 50 cm, 75 cm and 100 cm dose measurement at the measurement height of 100 cm, 64.2 uR was reduced to 47.3 uR in the senser mode 0.32sec. Even in film mode it was measured at 414.4 uR and about 27% lower at 162.6 uR. As the result of this study, when the irradiation time is 2 sec the right side dose is 290.5 uR and sensor mode is 0.32 sec the right side dose is 64.2 uR
Abstract
We aim to evaluate safety of radiation by measuring leakage dose and patient(phantom) incident dose of ZEN-PX II dental portable equipment developed by G company. Measurement for leakage dose of equipment is conducted on the top, at the bottom, on the left, on the right and at the back. Dose measurement incident on the subject with the area dosimeter when using the phantom and measurement the leakage dose of equipment when using the phantom are evaluated. Comparing the right with the highest leakage dose as a 0 cm, 25 cm, 50 cm, 75 cm and 100 cm dose measurement at the measurement height of 100 cm, 64.2 uR was reduced to 47.3 uR in the senser mode 0.32sec. Even in film mode it was measured at 414.4 uR and about 27% lower at 162.6 uR. As the result of this study, when the irradiation time is 2 sec the right side dose is 290.5 uR and sensor mode is 0.32 sec the right side dose is 64.2 uR
참고문헌 (10)
- 신귀순, "국내 치과방사선의 현황 및 이용 실태", 방사선기술과학, 33(2), pp.109~120 (2010), 대한방사선과학회
- Kenneth PH, "Radiation safty for the NOMAD™ portable X-ray system in a temporary morgue setting", J Forensic Sci, 53, pp.917~921 (2008)
- 김은경, "이동용 치과 X선 발생장치의 누설 및 산란 선량에 관한 연구", Imaging Science in Dentistry, 37(2), pp.65~68 (2007), 대한영상치의학회
- Radiation safety of portable dental X-ray systems, "Kim EK", KFDA, ~ (2011)
- Radiation Safety of Portable Dental X-ray System, "KFDA", Radiation safety management, ~, pp.Radiation safety management series No.37 (2014)
- Schueler BA, "Operator shielding : how and why", Tech Vasc Interv Radiol, 13, pp.167~171 (2010)
- Le Heron J, "Radiation protection of medical staff", Eur J Radiol, 76, pp.20~23 (2010)
- Medical Radiation Protection in Japan, "Chapter 4of the Medical Law Enforcement Rules", ~ (2011)
- 고종경, "치과 파노라마 촬영에서 공간선량률 분석", 방사선기술과학, 39(4), pp.509~516 (2016), 대한방사선과학회[DOI]