문 개폐 여부와 차폐체 설치 유무에 따른 공간산란선량 측정 : X선 촬영 시 피폭선량 감소방안에 대한 연구
Measurement of the Spatial Scattering Dose by Opening, Closing Door and Installing Shielding : A Study on the Reduction of Exposure Dose in Radiography
1
조회
0
다운로드
3
인용
0
참고문헌
키워드
초록
Recently, due to the increased use of medical radiation, the radiation exposure of radiation workers should be considered as well as medical exposure of patients. And it is recommended to close the door during radiography. however, In this study, when the door was inevitably opened for radiography, the proposed method was to install the shield as a method of reducing the exposure dose. And its efficiency was analyzed. In simple chest radiography, the measurement point was changed according to the measurement location. Dose rate were measured 10 times for each condition using a dosimeter. And the average value was derived. Using this, the change of dose according to the opening and closing of the door and the installation of the shield was analyzed. Using this, we compared and analyzed the dose change according to the door opening and closing and the installation of the shield, and significance was verified through the SPSS ver. 24. Depending on whether the door was opened or closed, 11,215.35%, 159.0%, 101.9% increased in front of the door in the consol room, behind the wall and behind the lead glass. Depending on the installing of the shield, the 49.2%, 29.6%, 19.9%, 30.6% decrease in front of the door in the examination and consol room, behind the wall and lead glass. In addition, statistical analysis was showed that there were significant differences in both the results according to whether the door was opened or closed and shielding(p<.05). Close the door during radiography. However, when the door should be opened, it was confirmed that the dose rate were reduced by installing the shield. Therefore, to optimize radiation protection, it is recommended to install shields when opening the door.
Abstract
Recently, due to the increased use of medical radiation, the radiation exposure of radiation workers should be considered as well as medical exposure of patients. And it is recommended to close the door during radiography. however, In this study, when the door was inevitably opened for radiography, the proposed method was to install the shield as a method of reducing the exposure dose. And its efficiency was analyzed. In simple chest radiography, the measurement point was changed according to the measurement location. Dose rate were measured 10 times for each condition using a dosimeter. And the average value was derived. Using this, the change of dose according to the opening and closing of the door and the installation of the shield was analyzed. Using this, we compared and analyzed the dose change according to the door opening and closing and the installation of the shield, and significance was verified through the SPSS ver. 24. Depending on whether the door was opened or closed, 11,215.35%, 159.0%, 101.9% increased in front of the door in the consol room, behind the wall and behind the lead glass. Depending on the installing of the shield, the 49.2%, 29.6%, 19.9%, 30.6% decrease in front of the door in the examination and consol room, behind the wall and lead glass. In addition, statistical analysis was showed that there were significant differences in both the results according to whether the door was opened or closed and shielding(p<.05). Close the door during radiography. However, when the door should be opened, it was confirmed that the dose rate were reduced by installing the shield. Therefore, to optimize radiation protection, it is recommended to install shields when opening the door.
참고문헌 (14)
- Jeong TS, "The analysis of radiation exposure of hospital radiation workers", Journal of The Korean Society for Radiation Oncology, 18(2), pp.157~166 (2000)
- Choo SS, "Current status of radiation safety management by medical use of radiation", Journal of The Korean Association for Radiation Protection, 15(1), pp.39~46 (1990)
- Lee GG, "A study on the job satisfaction of radiologist", Master's, pp.2001
- 김유현, "진단방사선영역에서 방사선장치의 이용실태 및 환자피폭선량에 관한 조사연구", 의학물리, 16(1), pp.10~15 (2005), 한국의학물리학회
- 임인철, "영상의학과 엑스선 발생장치의 누설 및 산란선량 측정", 한국방사선학회논문지, 5(3), pp.155~159 (2011), 한국방사선학회
- Lee HL, "Radiation doses from occupational exposure of the radiological workers", The Report of National Institute of Health, 29(2), pp.460~466 (1992)
- 이주아, "의료기관 방사선작업종사자와 임상실습 학생의 피폭선량 분석", 한국산학기술학회논문지, 17(8), pp.442~448 (2016), 한국산학기술학회[DOI]
- Ryu JK, "Occupational radiation exposure to hospital workers : On the effect of hematological change", The Korean Journal of Nuclear Medicine Technology, 12(3), pp.157~170 (2008)
- "ICRP 103, Recommendations of the international commision on radiological protection", ICRP Publication, ~ (2007)
- "IEC 60601-1-3 Medical electrical equipment, Part 1. General requirements for radiation protection in diagnostic X-ray equipment dissertation", 13(60), pp.331~338 (2013)
- 최원근, "방사선촬영 중 제어실 문의 열린 방향에 따른 차폐효과", 한국산학기술학회논문지, 11(9), pp.3347~3352 (2010), 한국산학기술학회
- 최성관, "X선 촬영에서 조사야 크기에 따른 산란선량의 변화", 한국콘텐츠학회 논문지, 13(3), pp.198~203 (2013), 한국콘텐츠학회[DOI]
- Shin GS, "Radiation physics 1", pp.247~249 (2014), Shinkwang Publising Company
- Lee SH, "Evaluation on the radiation exposure of radiation workers in proton therapy", The Journal of Korean society for Radiation Therapy, 24(2), pp.107~104 (2012)