흉부 전산화단층촬영검사를 위한 최적의 저선량 프로토콜에 관한 팬텀연구

A Phantom Study for the Optimal Low-dose Protocol in Chest Computed Tomography Examination

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류방사선과학

초록

The purpose of this study was to evaluate optimal CT scan parameters to minimize patient dose to the irradiation and maintain satisfactory image quality in low-dose chest computed tomography (CT) scans. In a chest anthropomorphic phantom, chest CT scans were performed at different kVp and mA within reference of 3.4mGy in volume CT Dose Index (CTDIvol). The following quantitative parameters had been statistically evaluated: image noise, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and figure of merit (FOM). Nine radiographers conducted the blind test to select the optimal kVp-mA combination. Results indicated that the kVp-mA combination of 80kVp-90mA, 100kVp-50mA, 120kVp-30mA and 140kVp-30mA were obtained high SNR and CNR. The 120kVp-30mA combination offered good compromise in the FOM, which showed the quality and dose performance. In the blind test, an image of 80kVp-90mA obtained a high score with 4.7 points, and 120kVp-10mA or 140kVp-10mA with a low tube current were observed severe noise and poor image quality, thus resulting in decreased diagnostic accuracy. On the other hand, in the combination of high kVp and high mA(140kVp-90mA), the image quality was improved, but the radiation dose was also increased. the FOM value of 140kVp-90mA was lower than 120kVp-30mA. The application of appropriate scan parameters in low-dose chest CT scans produced satisfactory results in dose and image quality for the accuracy of the clinical diagnosis.

Abstract

The purpose of this study was to evaluate optimal CT scan parameters to minimize patient dose to the irradiation and maintain satisfactory image quality in low-dose chest computed tomography (CT) scans. In a chest anthropomorphic phantom, chest CT scans were performed at different kVp and mA within reference of 3.4mGy in volume CT Dose Index (CTDIvol). The following quantitative parameters had been statistically evaluated: image noise, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and figure of merit (FOM). Nine radiographers conducted the blind test to select the optimal kVp-mA combination. Results indicated that the kVp-mA combination of 80kVp-90mA, 100kVp-50mA, 120kVp-30mA and 140kVp-30mA were obtained high SNR and CNR. The 120kVp-30mA combination offered good compromise in the FOM, which showed the quality and dose performance. In the blind test, an image of 80kVp-90mA obtained a high score with 4.7 points, and 120kVp-10mA or 140kVp-10mA with a low tube current were observed severe noise and poor image quality, thus resulting in decreased diagnostic accuracy. On the other hand, in the combination of high kVp and high mA(140kVp-90mA), the image quality was improved, but the radiation dose was also increased. the FOM value of 140kVp-90mA was lower than 120kVp-30mA. The application of appropriate scan parameters in low-dose chest CT scans produced satisfactory results in dose and image quality for the accuracy of the clinical diagnosis.

참고문헌 (11)

  1. Kubo T, "Standard-dose vs. low-dose CT protocols in the evaluation of localized lung lesions: Capability for lesion characterization-iLEAD study", European Journal of Radiology Open, 3, pp.67~73 (2016)
  2. 이원정, "흉부 CT촬영에서 저선량 프로토콜의 선량과 화질: 표준선량 프로토콜과 비교", 방사선방어학회지, 37(2), pp.84~89 (2012), 테스트학회[DOI]
  3. 김현주, "흉부 CT 검사 시 저 관전압 영상의 화질평가에 관한 연구", 방사선방어학회지, 35(4), pp.135~141 (2010), 테스트학회
  4. 김기남, "폐 전이 결절 진단에서 저선량 다중검출기 CT 영상의 유용성", 대한영상의학회지, 55(3), pp.239~245 (2006), 대한영상의학회
  5. Kraemer A, "Projection based evaluation of CT image quality in dimensional metrology", Digital Industrial Radiology and Computed Tomography, pp.22~25 (2015)
  6. Verdun FR, "Image quality in CT : From physical measurements to model observers", Physica Medica, 31(8), pp.823~843 (2015)
  7. Valentin J, "Managing patient dose in multi-detector computed tomography(MDCT). ICRP Publication 102", Ann ICRP, 37(1), pp.1~79 (2007)
  8. Bhalla AS, "Imaging protocols for CT chest : A recommendation", Indian J Radiol Imaging, 29(3), pp.236~246 (2019)
  9. Zhu X, "Low-dose chest CT : Optimizing radiation protection for patients", American Roentgen Ray Society, 183, pp.809~816 (2004)
  10. Christner JA, "Estimating effective dose for CT using dose–length product compared with using organ doses : Consequences of adopting international commission on radiological protection publication 103 or dual-energy scanning", Medical Physics and Informatics, 194, pp.881~889 (2010)
  11. "ME1, European Communities. European guidelines on quality criteria for diagnostic radiographic images", European Commission, ~ (1997)

관련 논문AI추천

유리선량계를 이용한 관전류자동조절기법과 고정관전류기법에서 저선량 및 고해상 흉부CT의 노이즈 및 선량 비교권대철, 박태석, 이은임, 조규원, 조승연, 한준희(사)한국방사선산업학회 · 방사선산업학회지2017 · 제11권 제3호Image Evaluation and Exposure Dose with the Application of Tube Voltage and Adaptive Statistical Iterative Reconstruction of Low Dose Computed Tomography김기정, 문태준, 이혜남대한방사선과학회 · Journal of Radiological Science and Technology2017 · 제40권 제2호저농도 조영제를 사용한 CT검사에서 저관전압 기법에 따른 유용성 평가정강교, 조평곤대한방사선과학회 · Journal of Radiological Science and Technology2016 · 제39권 제1호CT 장비별 관전압의 변화에 따른 노이즈, SNR 평가 - 자동노출제어방법과 수동노출제어방법 비교 -곽종길, 김미현, 동경래, 문일봉, 박상원, 성호진(사)한국방사선산업학회 · 방사선산업학회지2018 · 제12권 제3호전산화단층촬영에서 관전압과 관전류, 통계적 반복재구성법에 따른 화질과 피폭선량김정훈, 박재윤, 이용기, 차상영, 최재호대한방사선과학회 · Journal of Radiological Science and Technology2017 · 제40권 제3호흉부 전・후방향 검사 시 고관전압 및 산란선 후처리 소프트웨어 적용이 화질과 선량에 미치는 영향김종석, 이승근, 주영철대한방사선과학회 · Journal of Radiological Science and Technology2021 · 제44권 제4호

이 간행물의 다른 논문 (7건)