HTO 형태 시료 조제 조건에 따른 삼중수소 계측에 미치는 영향 평가

Evaluation of Effects on Tritium Measurement According to HTO Type Sample Preparation Conditions

1

조회

0

다운로드

2

인용

0

참고문헌

키워드

주제분류방사선과학

초록

In this study, for the measurement of 3H(tritium) radioactivity concentration, a study was conducted on whether the type of cocktail and the material of the vial had an effect on the measurement before liquid scintillation counter measurement on HTO-type samples that had undergone physical and chemical pretreatment. As a result of the study, the efficiency according to the type of cocktail was higher in Ultima Gold LLT than Ultima Flo-AP cocktail with polyethylene (1.49%), glass (5.10%), and teflon (6.58%), respectively. Regarding the effect according to the type of vial, the efficiency and SQP(E) of both Ultima Gold LLT and Ultima Flo-AP showed the highest values in the order of teflon, polyethylene, and glass.

Abstract

In this study, for the measurement of 3H(tritium) radioactivity concentration, a study was conducted on whether the type of cocktail and the material of the vial had an effect on the measurement before liquid scintillation counter measurement on HTO-type samples that had undergone physical and chemical pretreatment. As a result of the study, the efficiency according to the type of cocktail was higher in Ultima Gold LLT than Ultima Flo-AP cocktail with polyethylene (1.49%), glass (5.10%), and teflon (6.58%), respectively. Regarding the effect according to the type of vial, the efficiency and SQP(E) of both Ultima Gold LLT and Ultima Flo-AP showed the highest values in the order of teflon, polyethylene, and glass.

참고문헌 (17)

  1. Yeom YS, "An evaluation on uncertainty in LSC calibration and verification" (2011), University of Chosun
  2. 한상준, "국내 원전 삼중수소 방사능 배출 및 환경 거동에 대한분석 및 고찰", 방사선방어학회지, 40(4), pp.267~276 (2015), 테스트학회
  3. KRISS, "Liquid scintillation method", ~ (1997)
  4. "HASL-300 3H-04-RC. Tritium in water-Liquid scintillation counting", ~ (1997)
  5. Wallac Company, "1220 quantulus liquid scintillation counter, instrument manual", ~ (1995)
  6. Packard Instrument Company., "Laboratory manual for liquid scintillation counting", ~ (1988)
  7. PerkinElmer, "Quantulus 1220 liquid scintillation counter instrument manual", ~ (2004)
  8. L'Annunziata MF, "Handbook of radioactivity analysis" (2003), Academic Press
  9. Development on the measurement procedures of the C-14 activity of the nuclear power plants worker's urine, "Lee GJ", Chosun University, ~ (2008)
  10. 한상준, "LSC 장비를 이용한 교정시 Activity 및 Geometry 차이에 의한 영향 평가", 방사선방어학회지, 33(1), pp.21~26 (2008), 테스트학회
  11. KINS, "Manual of standard procedures for environmental sampling and radioactivity analysis", Korea Institute of Nuclear Safety, KINS
  12. KINS, "Development of procedure manual for the radionuclides analysis", ~ (2013)
  13. A Canberra Company, "Liquid scintillation analysis", Science and Technology, ~ (1998)
  14. PerkinElmer, "Cocktails for liquid scintillation counting", ~ (1997)
  15. 조정원, "먹는 샘물의 방사성물질 측정 및 유해성 평가", 방사선기술과학, 44(3), pp.247~252 (2021), 대한방사선과학회
  16. 장은성, "몬테카를로 시뮬레이션에서의 다양한 환경 샘플에 대한 Marinelli 비이커 측정 및 자기 흡수 보정과 적용", 방사선기술과학, 40(4), pp.605~611 (2017), 대한방사선과학회[DOI]
  17. A research on development and correction of standard source of liquid scintillation counter for analyzing environment sample, "KETEP", ~, pp.R-2005-0-005 (2007)

관련 논문AI추천

이 간행물의 다른 논문 (13건)