초록

Gamma spectroscopy analysis is widely used for radioactivity analysis, and various factors are required for radioactivity calculations. Among the factors, K3 for each sample significantly influences the results. The previous methods of correcting the self-absorption effect include a computational simulation method and a method that requires making a CRM(certified reference material) identical to the sample medium. However, the above methods have limitations when used in small institutions because they require specialized program utilization skills or high manufacturing costs and large facilities. The aim of this study is to develop a method that can be easily and rapidly applied to radioactivity analysis. After filling the beaker with water, we placed the radiation source in a uniform position and used the measured value as the benchmark. Next, a correction factor was derived based on the difference in the radiation source count of the benchmark and the identically measured sample. For the radiation source, Eu-152, which emits a broad range of energy within the measurement range of gamma rays, and Cs-134 and Cs-137, which are indicator nuclides in environmental radiation analysis, were used. The sample was selected within the density range of 0.26–2.11 g/cm3, and the correction factor was derived by calculating the count difference of each sample compared to the reference value of water. This study presents a faster and more convenient method than the existing research methods for determining the self-absorption effect correction, which has become increasingly necessary.

Abstract

Gamma spectroscopy analysis is widely used for radioactivity analysis, and various factors are required for radioactivity calculations. Among the factors, K3 for each sample significantly influences the results. The previous methods of correcting the self-absorption effect include a computational simulation method and a method that requires making a CRM(certified reference material) identical to the sample medium. However, the above methods have limitations when used in small institutions because they require specialized program utilization skills or high manufacturing costs and large facilities. The aim of this study is to develop a method that can be easily and rapidly applied to radioactivity analysis. After filling the beaker with water, we placed the radiation source in a uniform position and used the measured value as the benchmark. Next, a correction factor was derived based on the difference in the radiation source count of the benchmark and the identically measured sample. For the radiation source, Eu-152, which emits a broad range of energy within the measurement range of gamma rays, and Cs-134 and Cs-137, which are indicator nuclides in environmental radiation analysis, were used. The sample was selected within the density range of 0.26–2.11 g/cm3, and the correction factor was derived by calculating the count difference of each sample compared to the reference value of water. This study presents a faster and more convenient method than the existing research methods for determining the self-absorption effect correction, which has become increasingly necessary.

참고문헌 (18)

  1. 김창종, "국내 시중 유통 수산물에 대한 방사능 농도 조사", 한국식품과학회지, 47(6), pp.789~792 (2015), 한국식품과학회[DOI]
  2. 이완로, "실용적이고 간단한 환경시료의 감마핵종자체흡수보정 방법", 방사선방어학회지, 31(1), pp.1~6 (2006), 테스트학회
  3. 지영용, "방사성폐기물드럼 핵종분석에서 감마선 감쇠보정 방법들의 비교평가", Journal of Nuclear Fuel Cycle and Waste Technology, 4(3), pp.275~284 (2006), 한국방사성폐기물학회
  4. Kim CJ, "A Study on the Determination Method of Line Dampening Coefficients for Self-absorption Correction in Gamma-ray Analysis", Journal of Radiation Protection and Research, 11, pp.122~123 (2010)
  5. Seo BK, "Corrections of Self-Absorption Effect Using the Monte Carlo Method in the Radioactivity Analysis of Environmental Samples", Journal of Radiation Protection and Research, 26(2), pp.51~58 (2001)
  6. 장은성, "몬테카를로 시뮬레이션에서의 다양한 환경 샘플에 대한 Marinelli 비이커 측정 및 자기 흡수 보정과 적용", 방사선기술과학, 40(4), pp.605~611 (2017), 대한방사선과학회[DOI]
  7. 이규영, "몬테카를로 전산해석을 이용한 콘크리트 코어시료의 직경과 밀도에 따른 보정인자 계산", 한국방사선학회논문지, 14(5), pp.503~510 (2020), 한국방사선학회[DOI]
  8. "KASTO 16-80212-062. Standard Calibration Procedure of Semiconductor Detector for Gamma-ray Measurements", Korea Association of Standard & Testing Organizations, ~ (2016)
  9. Jo MH, "Uncertainty Evaluation of Energy Calibration Function of Gamma Nuclide Analysis Device", Journal of Radiation Protection and Research, 11, pp.114~115 (2011)
  10. 송병철, "방사선 측정기의 품질관리 기법에 대한 고찰", 분석과학, 24(6), pp.414~420 (2011), 한국분석과학회
  11. Lee WG, "Gamma Measurement of Soil and Food in Gyeonggi Province", The Journal of Industrial Liaison Research Institute, 13, pp.35~43 (2007)
  12. Yoo JH, "Investigation for some of the consumer goods which contain radioactive materials on the viewpoint of mineralogical and chemical characteristics", Mineralogical Society of Korea and Petrological Society of Korea, (5), pp.109~111 (2009)
  13. 윤욱, "옥천층군 일대의 지표방사능과 감마선량 평가", 지질공학, 30(4), pp.577~588 (2020), 대한지질공학회
  14. "Standards and Specifications of Food-Part 8 General Test Method: 9.9.2. Testing of radionuclides by High Purity Germanium detector", MFDS Notice No. 2021-79, ~ (2021)
  15. ISO, "Measurement of radioactivity in the environment–Soil–Part.2: Guidance for the selection of the sampling strategy, Sampling and pre-treatment of samples", ~ (2015)
  16. 조정원, "먹는 샘물의 방사성물질 측정 및 유해성 평가", 방사선기술과학, 44(3), pp.247~252 (2021), 대한방사선과학회
  17. 이세령, "담배연기와 담뱃잎 내 함유된 방사능 농도분석 및 위해도 평가", 방사선기술과학, 44(5), pp.489~494 (2021), 대한방사선과학회
  18. 이유상, "Radon Hazard Review of Spilled Groundwater and Tap Water in Incheon Metropolitan City Subway Station", 방사선기술과학, 44(6), pp.671~677 (2021), 대한방사선과학회

관련 논문AI추천