약물검사 키트의 분석 민감도에 대한 연구

A Study of Analytical Sensitivity on TDM Test Kit in Clinical Chemistry

0

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류기타자연과학

초록

.

Abstract

Analytical sensitivity on TDM test is the lowest concentration that can be distinguished from background noise. The aim of study was to evaluate analytical sensitivity that is also referred to as the  lower limit of detection(LLD) about difference between zero calibrator and isotonic saline sample. We tested for 10 days with zero calibrators and 0.85% saline samples while running trilevel control samples under control. Raw data divided by two groups calculated mean and standard deviation from two sample populations and analytical sensitivity by X+2SD. In comparison with isotonic saline samples and zero calibrators, there were significant differences in phenytoin, phenobarbital and vancomycin, etc. Especially analytical sensitivity on phenytoin is at the same level as the upper limit of analytical measurement range with 40 μg/mL. We think the cause of this is matrix interference. In conclusion, we were sure that standard protocol for analytical sensitivity as lower limit of analytical measurement range on TDM test must be measured with zero standard rather than an isotonic saline sample and type 1 reagent DW for reducing matrix effects within interactions between different materials in a mixture.