MICS 표준에 기반한 무선 텔레메트리 시스템 개발
Development of a wireless telemetry system based on MICS standard
0
조회
0
다운로드
1
인용
0
참고문헌
키워드
초록
It is said that the desirable bio-signal measurement It and stimulation system should be an implantable type if the several problems such as biocompatibility, electrical safety, and so on are overcome. In addition to the biocompatibility issue, a robust RF communication and a stable electrical power source for the implantable bio-signal measurement and stimulation system are very important matters. In this paper, a wireless telemetry system which adopts the FCC’s approved MICS (medical implant communication service) protocol and a wireless power transmission has been proposed. The proposed system composed of a base station (BS) and an implantable medical device (IMD) has the advantages that the interference with other RF devices can be reduced by the use of the specially assigned MICS frequency band of 402 ᅠMHz to 405 MHz. Also, the proposed system includes various functions of a multi-channel bio-signal acquisition and an electric stimulation. Since the electrical power for the IMD can be provided by the inductive link between PCB patterned coils, the IMD needs no battery so that the IMD can be smaller size and much less dangerous than the active type IMD which includes the internal battery. Finally, the validity as a wireless telemetry system has been demonstrated through the experiments by using the implemented BS and IMD.
Abstract
It is said that the desirable bio-signal measurement It and stimulation system should be an implantable type if the several problems such as biocompatibility, electrical safety, and so on are overcome. In addition to the biocompatibility issue, a robust RF communication and a stable electrical power source for the implantable bio-signal measurement and stimulation system are very important matters. In this paper, a wireless telemetry system which adopts the FCC’s approved MICS (medical implant communication service) protocol and a wireless power transmission has been proposed. The proposed system composed of a base station (BS) and an implantable medical device (IMD) has the advantages that the interference with other RF devices can be reduced by the use of the specially assigned MICS frequency band of 402 ᅠMHz to 405 MHz. Also, the proposed system includes various functions of a multi-channel bio-signal acquisition and an electric stimulation. Since the electrical power for the IMD can be provided by the inductive link between PCB patterned coils, the IMD needs no battery so that the IMD can be smaller size and much less dangerous than the active type IMD which includes the internal battery. Finally, the validity as a wireless telemetry system has been demonstrated through the experiments by using the implemented BS and IMD.
참고문헌 (14)
- 지영준, "무구속 생체신호 측정기술의 현황과 전망", 대한전자공학회지, 32(12), pp.56~68 (2005)
- E. Matei, "A Biomedical implantable FES battery-powered micro-stimulator", IEEE 2008 Custom Integrated Circuits Conference, pp.317~324 (2008)
- 우상효, "전기 자극 강도에 따른 소장 내부에서의 수축력 관계", 센서학회지, 17(1), pp.1~8 (2008), 한국센서학회
- 이승하, "생체신호계측을 위한 체내 이식형 무선송수신 시스템 개발", 센서학회지, 17(1), pp.23~28 (2008), 한국센서학회
- 이태수, "Wireless Healthcare를 위한 무선통신기술의 현황", 제어자동화 시스템공학 회지, 12(2), pp.39~42 (2006)
- 김도현, "WBAN 기술표준화", TTA Journal, (118), pp.102~107 (2008)
- Y. Hu, "A fully integrated low-power BPSK demodulator for implantable medical devices", IEEE Tr. on Circuits and Systems, 52(12), pp.2552~2562 (2005)
- M. Ghovanloo, "A wireless implantable multichannel microstimulating system-on-a-chip with modular architecture", IEEE Tr. on Neural Systems and Rehabilitation Engineering, 12(3), pp.449~457 (2007)
- P. Roberts, "Harvesting the energy of cardiac motion to power a pacemaker", Circulation, 118(18), pp.679~680 (2008), American Heart Association
- H. G. Lim, "Implementation of electromagnetically coupled type charger using h-bridged coil driver for fully implantable middle ear hearing devices", The proceedings of the 21st Int. Technical Conference on Circuit & Systems, ComTableputers and Communications, 3(1), pp.569~572 (2006)
- 임형규, "완전 이식형 인공중이를 위한 베개형 비접촉 충전장치의 설계", 센서학회지, 14(2), pp.78~84 (2005), 한국센서학회