In this study, a postural change detection sensor module (PCDSM) was developed to detect postural changes in activities of daily living (ADL) and falls. The PCDSM consists of eight mercury sensors that measure angle variations in 360° rotation and 90° tilting. From the preliminary study, the output characteristics of the PCDSM were confirmed with the angle variations of rotational motion and a tilting table. Three experiments were conducted to test rotational motion, postural changes, and falling and lying. The results confirmed that the PCDSM could effectively detect postural changes, movement patterns, and falls or non-falls.
Abstract
In this study, a postural change detection sensor module (PCDSM) was developed to detect postural changes in activities of daily living (ADL) and falls. The PCDSM consists of eight mercury sensors that measure angle variations in 360° rotation and 90° tilting. From the preliminary study, the output characteristics of the PCDSM were confirmed with the angle variations of rotational motion and a tilting table. Three experiments were conducted to test rotational motion, postural changes, and falling and lying. The results confirmed that the PCDSM could effectively detect postural changes, movement patterns, and falls or non-falls.
참고문헌 (15)
- P. Kannus, "Prevention of falls and consequent injuries in elderly people", Lancet, 366, pp.1885~1893 (2005)
- C. P. Dykes, "Fall prevention in acute care hospitals", JAMA, 304(17), pp.1912~1918 (2010)
- K. Aminian, "Spatial-temporal parameters of gait measured by ambulatory system using miniature gyroscopes", J. Biomech, 35, pp.689~699 (2002)
- C. C. Yang, "A review of accelerometrybased wearable motion detectors for physical activity monitoring", Sensors, 10, pp.7772~7788 (2010)
- R. Williamson, "Detecting absolute human knee angle and angular velocity using accelerometers and rate gyroscopes", Med. Biol. Eng. Comput, 39, pp.1~9 (2001)
- J. Rueterboriesa, "Methods for gait event detection and analysis in ambulatory systems", Med. Eng. Phys, 32, pp.545~552 (2010)
- R. S. Goldberg, "The importance of swing-phase initial conditions in stiff-knee gait", J. Biomech, 36(8), pp.1111~1116 (2003)
- A. Mansfield, "The use of accelerometry to detect heel contact events for use as a sensor in FES assisted walking", Med. Eng. Phys, 25, pp.879~885 (2003)
- N. A. Alba, "Novel hydrogel-based reparation-free EEG electrode", IEEE Trans. Neural Syst. Rehabil. Eng., 18(4), pp.415~423 (2010)
- M. J. Burke, "A micropower dry-electrode ECG preamplifier", IEEE Trans. Biomed. Eng., 47(2), pp.155~162 (2000)