Trace analysis of Cd and Pb at surface modified thick film graphite electrode with Bi nanopowder has been carried out using square-wave anodic stripping voltammetry (SWASV) technique. Bi nanopowder synthesized by gas condensation (GC) method showed the size of 50~100 nm with BET surface area, ABET = 6.8 m² g-1. For a strong adhesion of the Bi nanopowder onto the screen printed carbon paste electrode, nafion solution was added into Bi-containing suspension. From the SWASV, it was found that the Bi nanopowder electrode exhibited a well-defined responses relating to the oxidations of Cd and Pb. The current peak intensity increased with increasing concentration of Cd and Pb. From the linear relationship between Cd/Pb concentrations and peak current, the sensitivity of the Bi nanopowder electrode was quantitatively estimated. The detection limit of the electrode was estimated to be 0.15 mg/l and 0.07 mg/l for Cd and Pb, respectively, on the basis of the signal-to-noise characteristics (S/N=3) of the response for the 1.0 μg/l solution under a 10 min accumulation.

Abstract

Trace analysis of Cd and Pb at surface modified thick film graphite electrode with Bi nanopowder has been carried out using square-wave anodic stripping voltammetry (SWASV) technique. Bi nanopowder synthesized by gas condensation (GC) method showed the size of 50~100 nm with BET surface area, ABET = 6.8 m² g-1. For a strong adhesion of the Bi nanopowder onto the screen printed carbon paste electrode, nafion solution was added into Bi-containing suspension. From the SWASV, it was found that the Bi nanopowder electrode exhibited a well-defined responses relating to the oxidations of Cd and Pb. The current peak intensity increased with increasing concentration of Cd and Pb. From the linear relationship between Cd/Pb concentrations and peak current, the sensitivity of the Bi nanopowder electrode was quantitatively estimated. The detection limit of the electrode was estimated to be 0.15 mg/l and 0.07 mg/l for Cd and Pb, respectively, on the basis of the signal-to-noise characteristics (S/N=3) of the response for the 1.0 μg/l solution under a 10 min accumulation.

참고문헌 (16)

  1. J. P. Riley, , Anal. Chim. Acta, 181, pp.117~ (1986)
  2. G. Jialong, , Talanta, 32, pp.1072~ (1985)
  3. C. M. Elson, , Anal. Chim. Acta, 134, pp.393~ (1982)
  4. M. A. Floyd, , Anal. Chem, 52, pp.431~ (1980)
  5. P. Schramel, , J. Angerer Int. Arch. Occup. Environ. Health, 69, pp.219~ (1997)
  6. G. Kefala, , Talanta, 61, pp.603~ (2003)
  7. L. Lin, , Talanta, 65, pp.144~ (2005)
  8. L. Lin, , Anal. Chim. Acta, 535, pp.9~ (2005)
  9. S. B. Hocevar, , Electrochim. Acta, 51, pp.706~ (2005)
  10. I. Svancara, , Electroanalysis, 18
  11. J. Wang, , Talanta, 69, pp.914~ (2006)
  12. A. Y. Yermakov, , Mater. Sci. Forum, 386, pp.455~ (2002)
  13. Y. R. Uhm, , Phys. Stat. Sol. A, 201, pp.1934~ (2004)
  14. B. S. Han, , IEEE Transactions on magnetics, 42, pp.3779~ (2006)
  15. Y. R. Uhm, , Mater. Sci. and Eng. A, 449, pp.813~ (2007)
  16. S. B. Hocevar, , Electroanalysis, 14, pp.112~ (2002)

관련 논문AI추천

이 간행물의 다른 논문 (9건)