CMOS 기반의 DSLR 카메라 시스템과 DR 시스템의 변조전달함수 비교

Comparison of MTF of CMOS-based Digital Camera System and DR System

1

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

The purpose of this study is to apply the RQA5 conditions recommended by IEC 61267 and comparethe spatial resolution to MTF using a CMOS-based DSLR camera system and a DR system to improve imagequality. The study was conducted to use as basic data to confirm. The edge phantom used to acquire images for MTFmeasurement was the TX5 tungsten edge test device of IEC 62220-1 standard, and the wire quality and edge methodsuggested by IEC were used for the measurement. Images were evaluated using a GOS (gadolinium oxysulfide)-basedscintillator for the DR-based X-ray system and a CsI:Tl scintillator for the DSLR camera system. In order to evaluateimages using a DR system and a DSLR camera system, the resolution evaluation using the RQA-5 quality suggestedby IEC applied MTF. In order to evaluate images using a DR system and a DSLR camera system, the resolutionevaluation using the quality of RQA-5 presented by IEC was performed by applying MTF. In the DR system, whenMTF was 2% and the spatial frequency was 0.98924 and 10%, The spatial frequency was 0.83627 LP/mm. WhenMTF was 50%, the spatial frequency was 0.4745 LP/mm. In the DSLR camera system, the MTF was 0.99251 at 2%,and the spatial frequency at 10% was 0.90956 LP/mm. When MTF was 50%, the spatial frequency was 0.3771 LP/mm. In conclusion, in digital X-ray inspection, image quality can be confirmed by comparing the MTF accordingto the resolution of the GOS-based DR system and the CMOS-based DSLR camera. Images acquired using the DRsystem can be used as basic data for diagnosis and treatment.

Abstract

The purpose of this study is to apply the RQA5 conditions recommended by IEC 61267 and comparethe spatial resolution to MTF using a CMOS-based DSLR camera system and a DR system to improve imagequality. The study was conducted to use as basic data to confirm. The edge phantom used to acquire images for MTFmeasurement was the TX5 tungsten edge test device of IEC 62220-1 standard, and the wire quality and edge methodsuggested by IEC were used for the measurement. Images were evaluated using a GOS (gadolinium oxysulfide)-basedscintillator for the DR-based X-ray system and a CsI:Tl scintillator for the DSLR camera system. In order to evaluateimages using a DR system and a DSLR camera system, the resolution evaluation using the RQA-5 quality suggestedby IEC applied MTF. In order to evaluate images using a DR system and a DSLR camera system, the resolutionevaluation using the quality of RQA-5 presented by IEC was performed by applying MTF. In the DR system, whenMTF was 2% and the spatial frequency was 0.98924 and 10%, The spatial frequency was 0.83627 LP/mm. WhenMTF was 50%, the spatial frequency was 0.4745 LP/mm. In the DSLR camera system, the MTF was 0.99251 at 2%,and the spatial frequency at 10% was 0.90956 LP/mm. When MTF was 50%, the spatial frequency was 0.3771 LP/mm. In conclusion, in digital X-ray inspection, image quality can be confirmed by comparing the MTF accordingto the resolution of the GOS-based DR system and the CMOS-based DSLR camera. Images acquired using the DRsystem can be used as basic data for diagnosis and treatment.

참고문헌 (22)

  1. Jung HW, "Performance characteristic of a CsI(Tl)flat panel detector radiography system", Journal of Radiological Science and Technology, 35(2), pp.109~117 (2012)
  2. Lee SH, "An analysis and survey of patient care curriculum of radiological science undergraduate students", Journal of Radiation Industry, 14(4), pp.335~341 (2020)[DOI]
  3. Kim JA, "Telehealth in the developing world", Healthcare Informatics Research, 16(2), pp.140~141 (2010)[DOI]
  4. Kong C, "Image quality evaluation of CsI : Tl and Gd2O2S detectors in the indirect-conversion DR system", Journal of the Korean Society of Radiology, 11(1), pp.27~35 (2017)[DOI]
  5. Bugby SL, "A comparison of CsI : Tl and GOS in a scintillator-CCD detector for nuclear medicine imaging", Journal of Instrumentation, 11(9), pp.09009~ (2016)[DOI]
  6. Seong YH, "Evaluation of the spatial resolution for exposure class in computed radiography by using the modulation transfer function", The Journal of Digital Policy & Management, 11(8), pp.273~279 (2013)[DOI]
  7. Lee JW, "Evaluation of the response of the modulation transfer function to a computed radiography photostimulable phosphor imaging plate for various numbers of X-ray exposures", Journal of the Korean Physical Society, 58(2), pp.326~333 (2011)
  8. Dobbins JT, "DQE(f)of four generations of computed radiography acquisition devices", Medical Physics, 22(10), pp.1581~1593 (1995)[DOI]
  9. Granfors PR, "Performance of a 41×41 cm2amorphous silicon flat panel X-ray detector for radiographic imaging applications", Medical Physics, 27(6), pp.1324~1333 (2000)[DOI]
  10. Panna A, "Performance of low-cost X-ray area detectors with consumer digital cameras", Journal of Instrumentation (2015)[DOI]
  11. Hejazi S, "System considerations in CCD-based X-ray imaging for digital chest radiography and digital mammography", Medical Physics, 24(2), pp.287~297 (1997)[DOI]
  12. Fan H, "DR with a DSLR: digital radiography with a digital single-lens reflex camera", Proceedings of SPIEThe International Society for Optical Engineering, 7622, pp.76225E~ (2010)
  13. Nguyen VG, "Digital radiography with a consumer camera : Imaging denoising and deblurring", IEIE Transactions on Smart Processing and Computing, 10(5), pp.398~406 (2021)[DOI]
  14. Chavarria M, "X-ray imaging detector for radiological applications adapted to the context and requirements of low- and middle-income countries", Review of Scientific Instruments, 93, pp.034102~ (2022)[DOI]
  15. Marks IH, "Medical equipment donation in low-resource settings: a review of the literature and guidelines for surgery and anaesthesia in low-income and middle-income countries", BMJ Global Health, 4, pp.e001785~ (2019)[DOI]
  16. Vassallo DJ, "Experience with a low-cost telemedicine system in three developing countries", Journal of Telemedicine and Telecare, 7(1), pp.56~58 (2001)[DOI]
  17. Kuhls-Gilcrist A, "Accurate MTF measurement in digital radiography using noise response", Medical Physics, 37(2), pp.724~735 (2010)[DOI]
  18. You BG, "MTF evaluation and clinical application according to the characteristic kernels in the computed tomography", Korean Journal of Medical Physics, 18(2), pp.55~64 (2007)
  19. Lee JY, "X-ray-acquired imaging and detection radiography system using digital radiography with a DSLR digital camera: preliminary results of a pilot study", Radiation Effects and Defects in Solids, 178(5-6), pp.654~667 (2023)[DOI]
  20. Cho HM, "Construction and performance evaluation of digital radiographic system", Korean Journal of Medical Physics, 18(3), pp.144~148 (2007)
  21. Kocaman S, "Radiometric and geometric accuracy analysis of RATSAT pan imagery", The International Archives of the Photogrammetry Remote Sensing and Spatial Information Sciences. XLI-B, 1, pp.349~353 (2016)[DOI]
  22. Kong C, "Image Quality Evaluation of CsI : Tl and Gd2O2S Detectors in the Indirect-Conversion DR System", Journal of the Korean Society of Radiology, 11(1), pp.27~35 (2017)[DOI]

관련 논문AI추천

이 간행물의 다른 논문 (28건)