마그네트론 스퍼터링 시스템을 이용한 정형외과용 PEEK의 타이타늄/하이드록시아파타이트 이중 코팅층의 표면 특성 분석
Surface Characteristics of Titanium/Hydroxyapatite Double Layered Coating on Orthopedic PEEK by Magnetron Sputtering System
0
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
In this study, we have fabricated pure titanium (Ti)/hydroxyapatite (HA) double layer coating on medical grade PEEK from magnetron sputtering system, an investigation was performed whether the surface can be had more improve bio-active for orthopedi/dental applications than that of non-coated one. Pure Ti and HA coating layer were obtained by a radio-frequency and direct current power magnetron sputtering system. The microstructures surface, mechanical properties and wettability of the pure Ti/HA double layer deposited on the PEEK were analyzed by scanning electron microscopy (SEM), energy dispersive X-ray spectroscopy (EDS), X-ray diffraction (XRD), nano-indentation, and contact angle test. According to the EDS and XRD results, the composition and crystal structure of pure Ti and HA coated surface were verified. The elastic modulus and hardness value were increased by pure Ti and HA coating, and the pure Ti/HA double layer coating surface has the highest value. The contact angle showed higher value for pure Ti/HA double layered coating specimens than that of non-coated (PEEK) surface.
Abstract
In this study, we have fabricated pure titanium (Ti)/hydroxyapatite (HA) double layer coating on medical grade PEEK from magnetron sputtering system, an investigation was performed whether the surface can be had more improve bio-active for orthopedi/dental applications than that of non-coated one. Pure Ti and HA coating layer were obtained by a radio-frequency and direct current power magnetron sputtering system. The microstructures surface, mechanical properties and wettability of the pure Ti/HA double layer deposited on the PEEK were analyzed by scanning electron microscopy (SEM), energy dispersive X-ray spectroscopy (EDS), X-ray diffraction (XRD), nano-indentation, and contact angle test. According to the EDS and XRD results, the composition and crystal structure of pure Ti and HA coated surface were verified. The elastic modulus and hardness value were increased by pure Ti and HA coating, and the pure Ti/HA double layer coating surface has the highest value. The contact angle showed higher value for pure Ti/HA double layered coating specimens than that of non-coated (PEEK) surface.
참고문헌 (17)
- Ma R, "Current Strategies to Improve the Bioactivity of PEEK", Int. J. Mol. Sci., 15, pp.5426~5445 (2014)[DOI]
- A. Godara, "The influence of sterilization processes on the micromechanical properties of carbon fiber-reinforced PEEK composites for bone implant applications", Acta Biomater., 3, pp.209~220 (2007)[DOI]
- A. Katzer, "Polyetheretherketone-Cytotoxicity and mutagenicity in vitro", Biomaterials, 23, pp.1749~1759 (2002)[DOI]
- Y. Zheng, "Bone-like apatite coating on functionalized poly(etheretherketone) surface via tailored silanization layers technique", Mater Sci Eng C Mater Biol Appl., 55, pp.512~523 (2015)[DOI]
- X. Wu, "Nano-TiO2", PEEK bioactive composite as a bone substitute material: in vitro and in vivo studies, 2012, pp.7, 1215~1225
- Baştan, F. E., "Electrophoretic co-deposition of PEEKhydroxyapatite composite coatings for biomedical applications", Colloids and Surfaces B: Biointerfaces, 169, pp.176~182 (2018)[DOI]
- 정용훈, "HA", Ti 복합층 코팅한 Ti-30Ta-(3~15)Nb 합금의 전기화학적 특성, 2008, pp.41(2), 57~62[DOI]
- "Standard Specification for Polyetheretherketone (PEEK) Polymers for Surgical Implant Applications, ASTM F2026-17", ~
- "Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process, ISO 10993-1:2009", ~
- J. U. Kim, "Measurement of oxide thin film dissolution rate on the HA?coated Ti alloy by scanning electron microscopy and impedance spectroscopy", Surface and Interface Analysis, 44, pp.1468~1472 (2012)[DOI]
- V. Kyriazis, "Skeletal calcium", phosphorus ratio measuring techniques and results. I. Microscopy and microtomography, 2014, pp.4, 1027~1034
- "Joint Committee on Powder Diffraction Standard (JCPDS) and International Center for Diffraction Data (ICDD, 44-1294)", ~
- H. J. Kim, "Hydroxyapatite formation on biomedical Ti-Ta-Zr alloys by magnetron sputtering and electrochemical deposition", Thin Solid Films, 572, pp.119~125 (2014)[DOI]
- J. G. C. Wolke, "In vivo dissolution behavior of various RF magnetron sputtered Ca?P coatings", J. Biomed. Mater. Res., 39, pp.524~530 (1998)[DOI]
- H. Isaksson, "Precision of nanoindentation protocols for measurement of viscoelasticity in cortical and trabecular bone", Journal of biomechanics, 43, pp.2410~2417 (2010)[DOI]
- K. Das, "TiO2 nanotubes on Ti: Influence of nanoscale morphology on bone cell-materials interaction", J. Biomed. Mater. Res. A, 90, pp.225~237 (2009)
- B. G. Zhang, "Bioactive coatings for orthopaedic implants-recent trends in development of implant coatings. International journal of molecular sciences", Int. J. Mol. Sci., 15, pp.11878~11921 (2014)[DOI]