The effect of sample handling on the rheological measurement of regenerated silk fibroin formic acid solution using parallel plate geometry

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

The effect of sample handling condition on the rheological measurement of regenerated silk fibroin formic acid solution using parallel plate geometry was investigated. In case of loading method, the loading by pouring sample solution resulted in the best reproducibility of rheological measurement. Loading with spoon showed a high variance of viscosity value at low shear rate region (0.01 ~ 1 sec^(-1)) while loading with syringe exhibited a low reproducibility of viscosity at high shear region (1 ~ 100 sec^(-1)) with a disappearance of shear thinning phenomenon. It was revealed that the sample loading with small extra amount lead to the most reproducible result. The sample loading with the exact amount for the measuring plate resulted in a lack of reproducibility of high shear viscosity, while the loading with large extra volume produced a limited consistency of low shear viscosity. It was turned out that 3 min. of waiting time before measurement was the optimum condition for reliable result. When the waiting time was less than 1 min., the low shear viscosity was obtained with a lack of consistency. On the other hand, the sample solution started drying when the waiting time increased up to 5 min.

Abstract

The effect of sample handling condition on the rheological measurement of regenerated silk fibroin formic acid solution using parallel plate geometry was investigated. In case of loading method, the loading by pouring sample solution resulted in the best reproducibility of rheological measurement. Loading with spoon showed a high variance of viscosity value at low shear rate region (0.01 ~ 1 sec^(-1)) while loading with syringe exhibited a low reproducibility of viscosity at high shear region (1 ~ 100 sec^(-1)) with a disappearance of shear thinning phenomenon. It was revealed that the sample loading with small extra amount lead to the most reproducible result. The sample loading with the exact amount for the measuring plate resulted in a lack of reproducibility of high shear viscosity, while the loading with large extra volume produced a limited consistency of low shear viscosity. It was turned out that 3 min. of waiting time before measurement was the optimum condition for reliable result. When the waiting time was less than 1 min., the low shear viscosity was obtained with a lack of consistency. On the other hand, the sample solution started drying when the waiting time increased up to 5 min.

참고문헌 (18)

  1. Baek HS, "Enhanced chondrogenic responses of articular chondrocytes onto porous silk fibroin scaffolds treated with microwave-induced argon plasma", Surface & Coatings Technology, 202(22-23), pp.5794~5797 (2008), ELSEVIER SCIENCE SA
  2. Hee Jung Cho, "The Effect of Dissolution Condition on the Yield, Molecular Weight, and Wet- and Electro-spinnability of Regenerated Silk Fibroins Prepared by LiBr Aqueous Solution", International Journal of Industrial Entomology, 20(2), pp.99~105 (2010), 한국잠사학회
  3. Deitzel JM, "The effect of processing variables on the morphology of electrospun nanofibers and textiles", Polymer, 42, pp.261~272 (2001)
  4. Fang Q, "In vitro and in vivo research on using Antheraea pernyi silk fibroin as tissue engineering tendon scaffolds", Mater Sci Eng, 29, pp.1527~1534
  5. Gupta P, "Electrospinning of linear homopolymers of poly(methyl methacrylate): exploring relationships between fiber formation, viscosity, molecular weight and concentration in a good solvent", Polymer, 46, pp.4799~4810 (2005)
  6. Jeong L, "Fabrication and applications of silk Fibroin nanofibers", Polym Sci Technol, 16, pp.556~567
  7. Kim J, "Comparison of methods for the repair of acute tympanic membrane perforations: Silk patch vs. paper patch", WOUND REPAIR AND REGENERATION, 18, pp.132~138 (2010), BLACKWELL PUBLISHING
  8. Ko JS, "Miscibility, structural characteristics, and thermal behavior of wet spun regenerated silk fibroin", Nylon 6 blend filaments, 2010, pp.11, 14~20
  9. Mckee MG, "Correlations of solution rheology with electrospun fiber formation of linear and branched polyesters", Macromolecules, 37, pp.1760~1767 (2004)
  10. Min B, "Electrospinning of silk fibroin nanofibers and its effect on the adhesion and spreading of normal human keratinocytes and fibroblasts in vitro", BIOMATERIALS, 25, pp.1289~1297 (2004), ELSEVIER SCI LTD
  11. Minoura N, "Attachment and growth of cultured fibroblast cells on silk protein matrices", J. Biomed Mater Res, 29, pp.1215~1221 (1995)
  12. Sachiko S, "Regeneration of Bombyx mori silk by electrospinning—part 1: processing parameters and geometric properties", Polymer, 44, pp.5721~5727 (2003)
  13. Sakabe H, "In vivo blood compatibility of regenerated silk fibroin", Sen-i Gakkaish, 45, pp.487~490 (1989)
  14. Santin M, "In vitro evaluation of the inflammatory potential of the silk fibroin", J Biomed Mater Res, 46, pp.382~389 (1999)
  15. Schneider A, "Biofunctionalized electrospun silk mats as a topical bioactive dressing for accelerated wound healing", Acta Biomaterialia, 5, pp.2570~2578 (2009)
  16. Sill TJ, "Electrospinning: applications in drug delivery and tissue engineering", Biomaterials, 29, pp.2989~2006 (2008)
  17. Um IC, "Wet spinning of silk polymer: I. Effect of coagulation conditions on the morphological feature of filament", Int J Biol Macromol, 34, pp.89~105 (2004)
  18. Wang Y, "Cartilage tissue engineering with silk scaffolds and human articular chondrocytes", Biomaterials, 27, pp.4434~4442 (2006)

관련 논문AI추천