절대복사온도계의 분광 복사 휘도 감응도를 교정하기 위해 압축된 polytetrafluoroethylene (PTFE)와 반사봉을 사용하여 높은 공간 복사 휘도 분포를 가진 KRISS형 소형 적분구를 제작하였다. 이 적분구의 공간 복사 휘도 균일도는 ±0.009%로 지금까지 국외 표준기관들이 보고한 최고값보다 5배 높았다. 또한, 상용 제품인 소결된 PTFE 적분구의 공간 복사 휘도 균일도를 10배 향상시켰다.
Abstract
A KRISS-type small integrating sphere with a high spatial radiance uniformity was made using pressed polytetrafluoroethylene (PTFE) and a reflective rod to calibrate the spectral radiance responsivity of absolute radiant thermometers. The spatial radiance uniformity of the KRISS-type small integrating sphere was ±0.009%, five times higher than the best value reported by foreign national metrology institutions thus far. In addition,we improved the spatial radiance uniformity of a commercial sintered PTFE integrating sphere by a factor of 10.
참고문헌 (8)
- H. W. Yoon, "Thermodynamic radiation thermometry using radiometers calibrated for radiance responsivity", Int. J. Thermophys, 32, pp.2217~2229 (2011)
- K. Anhalt, "New PTB setup for the absolute calibration of the spectral responsivity of radiation thermometers", Int. J. Thermophys, 30, pp.192~202 (2009)
- T. Keawprasert, "Monochromator-based absolute calibration of radiation thermometers", Int. J. Thermophys, 32, pp.1697~1706 (2011)
- Y. Yamaguchi, "New setup for the absolute spectral responsivity of radiation thermometers with a supercontinuum source", Proc. XX IMEKO World Congress Metrology for Green Growth (2012)
- Y. Yamaguchi, "Supercontinuumsource-based facility for absolute calibration of radiation thermometers", Int. J. Thermophys, 36, pp.1825~1833 (2015)
- P. Y. Barnes, "Diffuse reflectance of sintered and pressed polytetrafluoroethylene(PTFE)", Proc. SPIE, 3426, pp.190~194 (1998)
- D. -J. Shin, "Nonlinearity measurement of UV detectors using light emitting diodes in an integrating sphere", Proc. 9th International Conference on New Developments and Applications in Optical Radiometry, pp.77~78 (2005)
- Y. S. Yoo, "Spectral responsivity calibration of the reference radiation thermometer at KRISS by using a super-continuum laser-based high-accuracy monochromatic source", Metrologia, 53, pp.1354~1364 (2016)