The tele-monitoring system (TMS) refers to a system that monitors the emission state of pollutants on a 24-hour basis by online-connecting a remote water quality control center with automatic measurement devicesattached to the discharging outlets of wastewater processing facilities and wastewater-producing businesses. The core of the TMS is the reliability of data from the continuous water monitoring equipment (CWME). Thedata reliability depends upon the accuracy and uniformity of the monitoring equipment as well as the representativenessof samples. The water sampling system is a major factor affecting the measuring resultsbecause it consists of a series of continuous sample collection processes from the effluent outlet to the CWME. Experiments were performed with a specially designed sampling system. The mixing characteristics of eachsample in the designed system were investigated using three-dimensional computational analyses of fluid andtracer tests. In addition, the residence time distribution (RTD) and the dispersion number were analyzed. Theresults showed that the sampling system is applicable as a standard of performance test for sampling baths.

Abstract

The tele-monitoring system (TMS) refers to a system that monitors the emission state of pollutants on a 24-hour basis by online-connecting a remote water quality control center with automatic measurement devicesattached to the discharging outlets of wastewater processing facilities and wastewater-producing businesses. The core of the TMS is the reliability of data from the continuous water monitoring equipment (CWME). Thedata reliability depends upon the accuracy and uniformity of the monitoring equipment as well as the representativenessof samples. The water sampling system is a major factor affecting the measuring resultsbecause it consists of a series of continuous sample collection processes from the effluent outlet to the CWME. Experiments were performed with a specially designed sampling system. The mixing characteristics of eachsample in the designed system were investigated using three-dimensional computational analyses of fluid andtracer tests. In addition, the residence time distribution (RTD) and the dispersion number were analyzed. Theresults showed that the sampling system is applicable as a standard of performance test for sampling baths.

참고문헌 (14)

  1. 국립환경과학원, "수질오염공정시험기준", pp.1036~1037 (2011)
  2. 국립환경과학원, "환경측정기기 형식승인·정도검사 등에관한 고시, 제2013-35호", ~ (2013)
  3. 김영일, "침전지 추적자 실험의 해석을 위한 Dispersion Number의 적용", 상하수도학회지, 21(2), pp.187~193 (2007), 대한상하수도학회
  4. 수질자동측정망에 대한측정불확도 및 숙련도 시험 (T-N장비를 중심으로)”, "박종호", 한국물환경학회·대한상하수도학회, 539~540 (2010)
  5. 방기웅, "추적자 주입기의 형태 및 주입지점에 따른 침전지의 수리특성평가", 한국도시환경학회지, 11(2), pp.169~176 (2011), 한국도시환경학회
  6. 신영섭, "비정렬 격자계에서 S.I.P. 최적화 방법을 이용한 점성유동 수치해석", 대한조선학회 논문집, 49(2), pp.196~202 (2012), 대한조선학회[DOI]
  7. 이충열, "수질자동측정시스템에서 온라인 TOC 자동측정장치의 불확도 산출", 생태와 환경, 40(2), pp.193~200 (2007), 한국하천호수학회
  8. 환경부, "수질 TMS 설치·운영 업무편람", pp.54~66 (2012)
  9. 환경부, "환경측정기기의 정확도", 정밀도 개선연구(IV), 2012
  10. B. Morrill, "Sedimentation Basin Research and Design", Journal of the American Water Works Association, 24(9), pp.1442~1463 (1932)
  11. D. D. W. Tsai, "A Method of Short-Circuiting Comparison", Water Resource Manage, 26, pp.2689~2702 (2012)
  12. F. L. Hart, "Hydraulic Analysis of Model Treatment Units", Journal of the Environmental Engineering Division, pp.785~798 (1978)
  13. O. Levenspiel, "Chemical Reaction Engineering", pp.296~303 (1999), John Wiley&sons Publishing
  14. Y. Wang, "Diagnosis of Membrane Bioreactor Performance Through Residence Time Distribution Measurement", Desalination, 236(1), pp.120~216 (2009)