하모닉 직교 방식의 초음파 고조파 영상화 기법

Ultrasound Harmonic Imaging Method based on Harmonic Quadrature Demodulation

0

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

A harmonic quadrature demodulation method to extract the second harmonic component from focused ultrasound signals after a single transmit-receive event is proposed. In the proposed method, the focused ultrasound signal is converted into baseband inphase and quadrature components by multiplying with sine and cosine signals both having twice the center frequency of the transmitted signal and filtering the two modulated signals. The quadrature component is then passed through a Hilbert filter to be added to the inphase component, which leaves only the envelope of the second harmonic component. A novel phase estimation technique is employed in the proposed method to avoid the phase mismatch between the focused signal and the two modulating signals. The proposed method is verified through both theoretical analysis and computer simulations. It is shown that compared to the pulse inversion scheme the proposed method provides almost the same results for stationary targets and significantly improved harmonic to fundamental ratio for moving targets.

Abstract

A harmonic quadrature demodulation method to extract the second harmonic component from focused ultrasound signals after a single transmit-receive event is proposed. In the proposed method, the focused ultrasound signal is converted into baseband inphase and quadrature components by multiplying with sine and cosine signals both having twice the center frequency of the transmitted signal and filtering the two modulated signals. The quadrature component is then passed through a Hilbert filter to be added to the inphase component, which leaves only the envelope of the second harmonic component. A novel phase estimation technique is employed in the proposed method to avoid the phase mismatch between the focused signal and the two modulating signals. The proposed method is verified through both theoretical analysis and computer simulations. It is shown that compared to the pulse inversion scheme the proposed method provides almost the same results for stationary targets and significantly improved harmonic to fundamental ratio for moving targets.

참고문헌 (21)

  1. C.L. Reid, "Accuracy of evaluation of the presence and severity of aortic and mitral regurgitation by contrast 2-dimensional echocardiography", J Am Cardiol, 52, pp.519~524 (1983)
  2. P.J.A. Frinking, "Ultrasound contrast imaging: current and new potential methods", Ultrasound Med. & Biol, 26(6), pp.965~975 (2000)
  3. P.J.A. Frinking, "A new ultrasound contrast imaging approach based on the combination of multiple imaging pulses and a separate release burst", IEEE Trans. UFFC, 48(3), pp.643~651 (2001)
  4. N. de Jong, "Higher harmonics of vibrating gas-filled microspheres. Part one: simulations", Ultrasonics, 32(6), pp.447~453 (1994)
  5. B. Schrope, "Second harmonic ultrasonic blood perfusion measurement", Ultrasound Med. & Biol, 19(7), pp.567~579 (1993)
  6. B. Schrope, "Simulated capillary blood flow measurement using a nonlinear ultrasonic contrast agent", Ultrasonic imaging, 14(2), pp.134~158 (1992)
  7. M. Kornbluth, "Native Tissue Harmonic Imaging Improves Endocardial Border Definition and Visualization of Cardiac Structures", J Am Soc Echocardiogr, 11, pp.693~701 (1998)
  8. D. H. Spencer, "Use of harmonic imaging without echocardiographic contrast to improve two-dimensional image quality", J Am Cardiol, 82, pp.794~799 (1998)
  9. M. Averkiou, "A new imaging technique based on the nonlinear properties of tissues", IEEE Ultrasonics Symposium, 통권 : 2, pp.1561~1566 (1997)
  10. M. Averkiou, "Tissue harmonic imaging", IEEE Ultrasonics Symposium, 통권 : 2, pp.1563~1572 (2000)
  11. F. Tranquart, "Clinical Use of Ultrasound Tissue Harmonic Imaging", Ultrasound in Med.& Biol, 25(6), pp.889~894 (1999)
  12. J.J. Hwang, "Two Pulse Technique For Ultrasound Harmonic Imaging", US patent,5,951,478, ~ (1999)
  13. P.N. Burns, "Nonlinear Imaging", Ultrasound in Med.& Biol, 26(1), pp.S19~S22 (2000)
  14. C.C. Shen, "Motion artifacts of pulse inversion based tissue harmonic imaging", IEEE Trans. UFFC, 49(9), pp.1203~1211 (2002)
  15. H. Yao, "Post-beamforming second-order Volterra filter for nonlinear pulse-echo imaging", ICASSP, 2, pp.1133~1136 (2002)
  16. P. Phukpattaranont, "Post-beamforming Volterra filter for contrast agent imaging", Proc. IEEE EMBS, 2, pp.1172~1175 (2003)
  17. A. Trucco, "Harmonic beamforming: a new approach to removing the linear contribution from harmonic imaging", IEEE Ultrasonics Symposium, 통권 : 1, pp.457~460 (2003)
  18. A. Trucco, "Harmonic Beamforming: Performance Analysis and Imaging Results", IEEE Trans. IM, 55(6), pp.1965~1974 (2006)
  19. S.J. Roome, "Analysis of quadrature detectors using complex envelope notation", Proc. IEEE RSP, 136(2), pp.95~100 (1989)
  20. J.G. Proakis, "Digital communications", pp.338~339 (2001), McGrawHill
  21. L. Hoff, "Acoustic characterization of contrast agents for medical ultrasound imaging", pp.54~126 (2001), Kluwer Academic Publishers

관련 논문AI추천

이 간행물의 다른 논문 (11건)