본문 바로가기
KERD
논문KCI 등재

Early Results of Total Knee Arthroplasty Using a Built-in 3-Degree External Rotation Prosthesis

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

Purpose: We compared the outcomes of total knee arthroplasty (TKA) using the Genesis II prosthesis in which 3o of external rotation was incorporated into the femoral component and Nexgen LPS prosthesis in which the degree of external rotation was determined by the extent of intraoperative bone cutting. Materials and Methods: Of the patients who underwent TKA for osteoarthritis between November 2006 and December 2009, 326 patients (572 knees) were included in this study. The prosthesis of choice was Genesis II in 199 knees and Nexgen LPS in 373 knees. The mean follow-up was 27 months (range, 24 to 51 months). The American Knee Society score (KSS) and function score (FS) were used for clinical assessments. The maximum flexion angle and patellar tilt were compared between the groups. Results: At the final follow-up, no significant intergroup difference was observed regarding the KSS (p=0.217), FS (p=0.238), maximum flexion angle (p=0.124), and patellar tilt (p=0.156). Conclusions: There was no significant difference in the clinical outcomes and patellar tilt between the prosthesis in which 3o of external rotation was incorporated into the femoral component and the prosthesis in which external rotation was determined based on bone cutting.

Abstract

Purpose: We compared the outcomes of total knee arthroplasty (TKA) using the Genesis II prosthesis in which 3o of external rotation was incorporated into the femoral component and Nexgen LPS prosthesis in which the degree of external rotation was determined by the extent of intraoperative bone cutting. Materials and Methods: Of the patients who underwent TKA for osteoarthritis between November 2006 and December 2009, 326 patients (572 knees) were included in this study. The prosthesis of choice was Genesis II in 199 knees and Nexgen LPS in 373 knees. The mean follow-up was 27 months (range, 24 to 51 months). The American Knee Society score (KSS) and function score (FS) were used for clinical assessments. The maximum flexion angle and patellar tilt were compared between the groups. Results: At the final follow-up, no significant intergroup difference was observed regarding the KSS (p=0.217), FS (p=0.238), maximum flexion angle (p=0.124), and patellar tilt (p=0.156). Conclusions: There was no significant difference in the clinical outcomes and patellar tilt between the prosthesis in which 3o of external rotation was incorporated into the femoral component and the prosthesis in which external rotation was determined based on bone cutting.

참고문헌 (12)

  1. Anouchi YS, "The effects of axial rotational alignment of the femoral component on knee stability and patellar tracking in total knee arthroplasty demonstrated on autopsy specimens", Clin Orthop Relat Res, 287, pp.170~177 (1993)
  2. Poilvache PL, "Rotational landmarks and sizing of the distal femur in total knee arthroplasty", Clin Orthop Relat Res, 331, pp.35~46 (1996)
  3. Kaper BP, "The effect of builtin external femoral rotation on patellofemoral tracking in the genesis II total knee arthroplasty", J Arthroplasty, 15, pp.964~969 (2000)
  4. Bindelglass DF, "Patellar tilt and subluxation in total knee arthroplasty: relationship to pain, fixation, and design", Clin Orthop Relat Res, 286, pp.103~109 (1993)
  5. Shrout PE, "Intraclass correlations : uses in assessing rater reliability", Psychol Bull, 86, pp.420~428 (1979)
  6. Nicoll D, "Internal rotational error of the tibial component is a major cause of pain after total knee replacement", J Bone Joint Surg Br, 92, pp.1238~1244 (2010)
  7. Thompson JA, "Biomechanical effects of total knee arthroplasty component malrotation : a computational simulation", J Orthop Res, 29, pp.969~975 (2011)
  8. Ries MD, "Can rotational congruity be achieved in both flexion and extension when the femoral component is externally rotated in total knee arthroplasty?", Knee, 5, pp.37~41 (1998)
  9. Liau JJ, "The influence of contact alignment of the tibiofemoral joint of the prostheses in in vitro biomechanical testing", Clin Biomech(Bristol, Avon), 14, pp.717~721 (1999)
  10. Bourne RB, "Ten-year results of the first 100 Genesis II total knee replacement procedures", Orthopedics, 30(8 Suppl), pp.83~85 (2007)
  11. Bhandari M, "The Genesis II in primary total knee replacement : a systematic literature review of clinical outcomes", Knee, 19, pp.8~13 (2012)
  12. McCalden RW, "A randomized controlled trial comparing “high-flex” vs “standard” posterior cruciate substituting polyethylene tibial inserts in total knee arthroplasty", J Arthroplasty, 24(6 Suppl), pp.33~38 (2009)

관련 논문AI추천

High-Flexion Total Knee Arthroplasty Using NexGen LPS-Flex System: Minimum 5-year Follow-up Results서정탁, 이승준, 홍성민대한슬관절학회 · Knee Surgery and Related Research2015 · 제27권 제3호The Correlation between Posterior Tibial Slope and Maximal Angle of Flexion after Total Knee Arthroplasty김종민, 빈성일, Keong-Hwan Kim대한슬관절학회 · Knee Surgery and Related Research2012 · 제24권 제3호Comparison of Mobile-Bearing and Fixed-Bearing Designs in High Flexion Total Knee Arthroplasty: Using a Navigation System서정탁, Shi Hwan Park, Taewan Kim대한슬관절학회 · Knee Surgery and Related Research2012 · 제24권 제1호Early Clinical Outcomes of a New Posteriorly Stabilized Total Knee Arthroplasty Prosthesis: Comparisons with Two Established Prostheses김태균, 이성엽, Nimesh P. Jain, Suri Chong, Vivek M. Morey, Yeon Gwi Kang대한슬관절학회 · Knee Surgery and Related Research2017 · 제29권 제3호Results of Total Knee Arthroplasty with NexGen LPS-flex Implant Using Navigation System (Brain Lab): Results with a 5-year Follow-up강철형, 배기철, 이경재, 이시욱, 조철현, Hong Kwan Shin, Ji Suk Bae, Young Kook Lee대한슬관절학회 · Knee Surgery and Related Research2012 · 제24권 제4호Clinical Outcome of IM-Guided Total Knee Arthroplasty with Inappropriate Femoral Resection in Coronal Plane문영완, 서재곤, 이근호, 장성원, 한정훈대한슬관절학회 · Knee Surgery and Related Research2013 · 제25권 제1호

이 간행물의 다른 논문 (9건)