This study evaluated analytical methods for the mixture of ethylene glycol (EG) and didecyldimethyl ammonium chloride (DDAC), which are used as the main components of deodorants and biocides. EG and DDAC were nebulized in aerosol into inhalation chambers. For absorbents of EG and DDAC, XAD-7 OVS resin and glass fiber (GF) filters were used, respectively. Instrumental analysis was conducted with gas chromatography/ flame ionization detector (GC/FID) for EG and high performance liquid chromatography with triple-quadrupole mass spectrometer (LC/MS/MS) for DDAC. The method detection limit (MDL) and the limit of quantitation (LOQ) were 24 μg/m3 and 77 μg/m3 for EG, and 13 ng/m3 and 41 ng/m3 for DDAC. The recovery rates ranged from 100.7% to 105.5% for EG, and 97.4% to 98.9% for DDAC. The relative standard deviation (RSD) of EG and DDAC ranged less than 5.72%. Daily monitoring was also conducted at the concentration of 0.2% for 7 days. The results showed that the mean values ranged from 0.728 mg/m3 to 0.739 mg/m3 and the overall accuracy was 11.9% to 16.2%.

Abstract

This study evaluated analytical methods for the mixture of ethylene glycol (EG) and didecyldimethyl ammonium chloride (DDAC), which are used as the main components of deodorants and biocides. EG and DDAC were nebulized in aerosol into inhalation chambers. For absorbents of EG and DDAC, XAD-7 OVS resin and glass fiber (GF) filters were used, respectively. Instrumental analysis was conducted with gas chromatography/ flame ionization detector (GC/FID) for EG and high performance liquid chromatography with triple-quadrupole mass spectrometer (LC/MS/MS) for DDAC. The method detection limit (MDL) and the limit of quantitation (LOQ) were 24 μg/m3 and 77 μg/m3 for EG, and 13 ng/m3 and 41 ng/m3 for DDAC. The recovery rates ranged from 100.7% to 105.5% for EG, and 97.4% to 98.9% for DDAC. The relative standard deviation (RSD) of EG and DDAC ranged less than 5.72%. Daily monitoring was also conducted at the concentration of 0.2% for 7 days. The results showed that the mean values ranged from 0.728 mg/m3 to 0.739 mg/m3 and the overall accuracy was 11.9% to 16.2%.

참고문헌 (17)

  1. 국가표준인증종합정보센터, "자율안전확인 안전기준 부속서 7: 생활화학가정용품", pp.1~43 (2008)
  2. P. Jenkins., "Activity patterns of Californians : use of and proximity to indoor pollutant sources", Atmospheric Environment Part A, 26, pp.2141~2148 (1992)
  3. A. Steinemann, "Fragranced consumer products and undisclosed ingredients", Environmental Impact Assessment Review, 29, pp.32~38 (2009)
  4. S. Hahn., "Consumer exposure to biocieds-identification of relevant sources and evaluation of possible health effects", Environ Health, 72, pp.479~484 (2008)
  5. H. Sütterlin., "Mixtures of quaternary ammonium compounds and anionic organic compounds in the aquatic environment : Elimination and biodegradability in the closed bottle test monitored by LC", MS, Chemosphere, pp.2008, 72[DOI]
  6. M. Owen., "Airborne particle sizes and sources found in indoor air", Atmospheric Environment Part A, 26, pp.2149~2162 (1992)
  7. A. Afshari., "Characterization of indoor sources of fine and ultrafine particles : a study conducted in a full-scale chamber", Indoor air, 15, pp.141~150 (2005)
  8. 환경부, "유해화학물질공정시험기준", pp.7~12 (2012)
  9. W. H. Porter., "Simultaneous determination of ethylene glycol and glycolic acid in serum by gas chromatography-mass spectrometry", Journal of Analytical Toxicology, 23, pp.591~597 (1999)
  10. M. Holcapek., "Trace determination of glycols by HPLC with UV and electrospray ionizaton mass spectrometric detections", Analytical chemistry, 71, pp.2288~2293 (1999)
  11. M. Shibukawa., "Separation and determination of quaternary ammonium compounds by hig-performance liquid chromatography with a hydrophilic polymer column and conductometric detection", J Chromatogr A, 830, pp.321~328 (1999)
  12. 양정선, "흡입챔버 내 didecyldimethylammonium chloride(DDAC) 에어로졸의 분석", 분석과학, 25(5), pp.307~312 (2012), 한국분석과학회
  13. G. Vincent., "Sampling and analysis of quaternary ammonium compounds(QACs)traces in indoor atmosphere", Environ Monit Assess, 133, pp.25~30 (2007)
  14. O. Nunez., "Determination of quaternary ammonium biocides by liquid chromatography-mass spectrometry", J Chromatogr A, 1058, pp.89~95 (2004)
  15. National Institute for Occupational Safety and Health, "NIOSH manual of analytical methods: 5523, Issue 1" (1996)
  16. 살균제 에어로졸의 유해성 및작업장 사용실태 조사연구, "한국산업안전보건연구원", 142~150 (2012)
  17. Criteria for a recommended standard occupational exposure to metalworking fluid, "National Institute for Occupational Safety and Health", 98~102 (1998)