What is the Adequate Cuff Volume for Tracheostomy Tube? A Pilot Cadaver Study

0

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류재활의학

초록

Objective To determine the patterns of tracheostomy cuff pressure changes with various air inflation amounts in different types of tracheostomy tubes to obtain basic data for appropriately managing long-term tracheostomy. Methods We performed tracheostomy on a 46-year-old male cadaver. Three types of tracheostomy tubes (single-cuffed, double-cuffed, and adjustable flange), divided into 8 different subtypes based on internal tube diameters and cuff diameters, were inserted into the cadaver. Air was inflated into the cuff, and starting with 1 mL air, the cuff pressure was subsequently measured using a manometer. Results For the 7.5 mm/14 mm tracheostomy tube, cuff inflation with 3 mL of air yielded a cuff pressure within the recommended range of 20–30 cmH2O. The 7.5 mm/24 mm tracheostomy tube showed adequate cuff pressure at 5 mL of air inflation. Similar values were observed for the 8.0 mm/16 mm and 8.0 mm/27 mm tubes. Double-cuffed tracheostomy cuff pressures (7.5 mm/20 mm and 8.0 mm/20 mm tubes) at 3 mL air inflation had cuff pressures of 18–20 cmH2O at both the proximal and distal sites. For the adjustable flange tracheostomy tube, cuff pressure at 6 mL of cuff air inflation was within the recommended range. Maximal cuff pressure was achieved at inflation with almost 14 mL of air, unlike other tube types. Conclusion Various types of tracheostomy tubes showed different cuff pressures after inflation. These values might aid in developing guidelines For patients who undergo tracheostomy and are discharged home without cuff pressure manometers, this standard might be helpful to develop guidelines.

Abstract

Objective To determine the patterns of tracheostomy cuff pressure changes with various air inflation amounts in different types of tracheostomy tubes to obtain basic data for appropriately managing long-term tracheostomy. Methods We performed tracheostomy on a 46-year-old male cadaver. Three types of tracheostomy tubes (single-cuffed, double-cuffed, and adjustable flange), divided into 8 different subtypes based on internal tube diameters and cuff diameters, were inserted into the cadaver. Air was inflated into the cuff, and starting with 1 mL air, the cuff pressure was subsequently measured using a manometer. Results For the 7.5 mm/14 mm tracheostomy tube, cuff inflation with 3 mL of air yielded a cuff pressure within the recommended range of 20–30 cmH2O. The 7.5 mm/24 mm tracheostomy tube showed adequate cuff pressure at 5 mL of air inflation. Similar values were observed for the 8.0 mm/16 mm and 8.0 mm/27 mm tubes. Double-cuffed tracheostomy cuff pressures (7.5 mm/20 mm and 8.0 mm/20 mm tubes) at 3 mL air inflation had cuff pressures of 18–20 cmH2O at both the proximal and distal sites. For the adjustable flange tracheostomy tube, cuff pressure at 6 mL of cuff air inflation was within the recommended range. Maximal cuff pressure was achieved at inflation with almost 14 mL of air, unlike other tube types. Conclusion Various types of tracheostomy tubes showed different cuff pressures after inflation. These values might aid in developing guidelines For patients who undergo tracheostomy and are discharged home without cuff pressure manometers, this standard might be helpful to develop guidelines.

참고문헌 (22)

  1. Sengupta P, "Endotracheal tube cuff pressure in three hospitals, and the volume required to produce an appropriate cuff pressure", BMC Anesthesiol, 4, pp.8~ (2004)
  2. Braz JR, "Endotracheal tube cuff pressure : need for precise measurement", Sao Paulo Med J, 117, pp.243~247 (1999)
  3. Stauffer JL, "Complications and consequences of endotracheal intubation and tracheotomy : a prospective study of 150 critically ill adult patients", Am J Med, 70, pp.65~76 (1981)
  4. Leigh JM, "Pressure on the tracheal mucosa from cuffed tubes", Br Med J, 1, pp.1173~1174 (1979)
  5. Nordin U, "The trachea and cuff-induced tracheal injury : an experimental study on causative factors and prevention", Acta Otolaryngol Suppl, 345, pp.1~71 (1977)
  6. Klainer AS, "Surface alterations due to endotracheal intubation", Am J Med, 58, pp.674~683 (1975)
  7. Shelly WM, "Cuffed tubes as a cause of tracheal stenosis", J Thorac Cardiovasc Surg, 57, pp.623~627 (1969)
  8. Bernhard WN, "Intracuff pressures in endotracheal and tracheostomy tubes. Related cuff physical characteristics", Chest, 87, pp.720~725 (1985)
  9. Bernhard WN, "Adjustment of intracuff pressure to prevent aspiration", Anesthesiology, 50, pp.363~366 (1979)
  10. Bryant LR, "Reappraisal of tracheal injury from cuffed tracheostomy tubes : experiments in dogs", JAMA, 215, pp.625~628 (1971)
  11. 박여진, "Care Status of the ALS Patients With Long-Term Use of Tracheostomy Tube", Annals of Rehabilitation Medicine, 39(6), pp.964~970 (2015), 대한재활의학회
  12. 서미리, "Long-Term Outcome of Amyotrophic Lateral Sclerosis in Korean Subjects", Annals of Rehabilitation Medicine, 41(6), pp.1055~1064 (2017), 대한재활의학회
  13. Fernandez R, "Endotracheal tube cuff pressure assessment : pitfalls of finger estimation and need for objective measurement", Crit Care Med, 18, pp.1423~1426 (1990)
  14. Dunn CR, "Determinants of tracheal injury by cuffed tracheostomy tubes", Chest, 65, pp.128~135 (1974)
  15. Lewis FR Jr, "Prevention of complications from prolonged tracheal intubation", Am J Surg, 135, pp.452~457 (1978)
  16. Durbin CG Jr, "Techniques for performing tracheostomy", Respir Care, 50, pp.488~496 (2005)
  17. Hess DR, "Tracheostomy tubes and related appliances", Respir Care, 50, pp.497~510 (2005)
  18. Breatnach E, "Dimensions of the normal human trachea", AJR Am J Roentgenol, 142, pp.903~906 (1984)
  19. St John RE, "Contemporary issues in adult tracheostomy management", Crit Care Nurs Clin North Am, 16, pp.413~430 (2004)
  20. Otoch JP, "Is there a correlation between right bronchus length and diameter with age?", J Thorac Dis, 5, pp.306~309 (2013)
  21. Al-Shaikh B, "Essentials of equipment in anaesthesia, critical care, and peri-operative medicine" (2018), Elsevier
  22. Dullenkopf A, "Fluid leakage past tracheal tube cuffs : evaluation of the new Microcuff endotracheal tube", Intensive Care Med, 29, pp.1849~1853 (2003)

관련 논문AI추천

Guidelines for Tracheostomy From the Korean Bronchoesophagological Society강창현, 김관민, 남인철, 박민우, 박성용, 박일석, 선동일, 송창면, 신유섭, 이영찬, 이종민, 전재현, 정우진, Korean Bronchoesophagological Society대한이비인후과학회 · Clinical and Experimental Otorhinolaryngology2020 · 제13권 제4호Care Status of the ALS Patients With Long-Term Use of Tracheostomy Tube고성화, 김상훈, 박여진, 신명준, 신용범, 이제상, 장재혁대한재활의학회 · Annals of Rehabilitation Medicine2015 · 제39권 제6호The Effect of Positive End-Expiratory Pressure on Air Leakage: Comparison of Cuff Designs심경석, 인준용, 정승현대한중환자의학회 · Acute and Critical Care2014 · 제29권 제1호Association of Oversized Tracheal Tubes and Cuff Overinsufflation With Postintubation Tracheal RupturesAnnekatrin Coordes, Barbara Estel, Rainer O. Seidl, Tobias H. Sudhoff대한이비인후과학회 · Clinical and Experimental Otorhinolaryngology2015 · 제8권 제4호중환자실 환자에서 관혈적 기관 절개술과 경피적 확장 기관 절개술의 비교김경민, 남인철, 신현일, 장경일대한이비인후과학회 · 대한이비인후과학회지 두경부외과학2019 · 제62권 제5호Under- or overpressure: an audit of endotracheal cuff pressure monitoring at the tertiary care centerBiju Viswambharan, Gopala Krishnan, Lakshmi Ramamoorthy, Manjini Jeyaram Kumari대한중환자의학회 · Acute and Critical Care2021 · 제36권 제4호

이 간행물의 다른 논문 (9건)