본문 바로가기
KERD
논문KCI 등재

척수 손상에서 새로운 휴대용 요역동학 검사기의 개발 및 유용성

Availability of a Newly Devised Ambulatory Urodynamic System in Spinal Cord Injury

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류재활의학

초록

Objective: To introduce and evaluate the availability of a newly devised ambulatory urodynamic system using the abdominal EMG (electromyography) method instead of the rectal catheter in measuring the abdominal pressure, and to compare with conventional urodynamic system in patients with spinal cord injury. Method: We examined 15 hospitalized subjects with spinal cord injury. Patients were investigated by conventional urodynamics in an examination room, and followed by ambulatory urodynamics in a ward. We used the abdominal EMG and the rectal catheter for the abdominal pressure in the study. We measured urodynamic parameters : volumes of sensations in bladder filling, peak Pdet(detrusor pressure) during filling, maximal Pdet during voiding, bladder capacity, compliance and duration of filling. Results: There was no significant statistic difference in the parameters between the conventional and the ambulatory urodynamics. There were strong correlations between the parameters measured by the rectal catheter and the abdominal EMG in the ambulatory system and between the parameters in the conventional system and in the ambulatory system. Conclusion: There were strong correlation and no significant statistic differences in the parameters between the ambulatory system using abdominal EMG method and the conventional system. Therefore, we suggest that the newly devised ambulatory urodynamic system using abdominal EMG method can be used instead of the conventional non-ambulatory system and the conventional ambulatory system.

Abstract

Objective: To introduce and evaluate the availability of a newly devised ambulatory urodynamic system using the abdominal EMG (electromyography) method instead of the rectal catheter in measuring the abdominal pressure, and to compare with conventional urodynamic system in patients with spinal cord injury. Method: We examined 15 hospitalized subjects with spinal cord injury. Patients were investigated by conventional urodynamics in an examination room, and followed by ambulatory urodynamics in a ward. We used the abdominal EMG and the rectal catheter for the abdominal pressure in the study. We measured urodynamic parameters : volumes of sensations in bladder filling, peak Pdet(detrusor pressure) during filling, maximal Pdet during voiding, bladder capacity, compliance and duration of filling. Results: There was no significant statistic difference in the parameters between the conventional and the ambulatory urodynamics. There were strong correlations between the parameters measured by the rectal catheter and the abdominal EMG in the ambulatory system and between the parameters in the conventional system and in the ambulatory system. Conclusion: There were strong correlation and no significant statistic differences in the parameters between the ambulatory system using abdominal EMG method and the conventional system. Therefore, we suggest that the newly devised ambulatory urodynamic system using abdominal EMG method can be used instead of the conventional non-ambulatory system and the conventional ambulatory system.

참고문헌 (10)

  1. Schafer W, "Good urodynamic practices: uroflowmetry, filling cystometry and pressure-flow studies", Neurourol Urodyn, 21, pp.261~274 (2002)
  2. Malbrain ML, "Different technique to measure intra-abdominal pressure (IAP)-time for a critical re-appraisal", Intensive Care Med, 30, pp.357~371 (2004)
  3. 서정환, "척수손상환자의 요류동태검사에서 직장압 측정에 의한 오류", Annals of Rehabilitation Medicine, 30(6), pp.571~574 (2006), 대한재활의학회
  4. Van Waalwijk van Doorn E, "Standardisation of ambulatory urodynamic monitoring: report of the standardisation sub-committee of the International continence society for ambulatory urodynamic studies", Neurol Urodyn, 19, pp.113~125 (2000)
  5. Amarenco G, "Cough anal reflex: strict relationship between intravesical pressure and pelvic floor muscle electromyographic activity during cough. Urodynamic and electrophysiological study", J Urol, 173, pp.149~152 (2005)
  6. 서정환, "복압의 변화에 따른 복직근의 운동단위활동전위", Annals of Rehabilitation Medicine, 29(5), pp.457~461 (2005), 대한재활의학회
  7. Susset JG, "Cystometry. in: Clinical neurourology. 2nd ed", pp.163~184 (1991), Little Brown & Co
  8. Abrams P, "The standardisation of terminology of lower urinary tract function. The international continence society committee on standardisation of terminology", Scand J Urol Nephrol, 114, pp.5~19 (1988)
  9. Ko HY, "Comparison between conventional cystometry and diuretics stimulated filling cystometry in neurogenic bladder after spinal cord injury", J Korean Acad Rehab Med, 23, pp.948~952 (1999)
  10. Kim YC, "Accuracy of a portable ultrasound scanning in the measurement of post-void residual urine volume in patients with spinal cord injury", J Korean Acad Rehab Med, 21, pp.298~303 (1997)

관련 논문AI추천

이 간행물의 다른 논문 (13건)