본문 바로가기
KERD
논문KCI 등재

Contamination Rates in Duodenoscopes Reprocessed Using Enhanced Surveillance and Reprocessing Techniques: A Systematic Review and Meta-Analysis

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류내과학

초록

Background/Aims: Multiple outbreaks of multidrug-resistant organisms have been reported worldwide due to contaminatedduodenoscopes. In 2015, the United States Food and Drug Administration recommended the following supplemental enhancedsurveillance and reprocessing techniques (ESRT) to improve duodenoscope disinfection: (1) microbiological culture, (2) ethyleneoxide sterilization, (3) liquid chemical sterilant processing system, and (4) double high-level disinfection. A systematic review andmeta-analysis was performed to assess the impact of ESRT on the contamination rates. Methods: A thorough and systematic search was performed across several databases and conference proceedings from inceptionuntil January 2021, and all studies reporting the effectiveness of various ESRTs were identified. The pooled contamination rates ofpost-ESRT duodenoscopes were estimated using the random effects model. Results: A total of seven studies using various ESRTs were incorporated in the analysis, which included a total of 9,084 post-ESRTduodenoscope cultures. The pooled contamination rate of the post-ESRT duodenoscope was 5% (95% confidence interval [CI]:2.3%–10.8%, inconsistency index [I2]=97.97%). Pooled contamination rates for high-risk organisms were 0.8% (95% CI: 0.2%–2.7%,I2=94.96). Conclusions: While ESRT may improve the disinfection process, a post-ESRT contamination rate of 5% is not negligible. Ongoingefforts to mitigate the rate of contamination by improving disinfection techniques and innovations in duodenoscope design toimprove safety are warranted.

Abstract

Background/Aims: Multiple outbreaks of multidrug-resistant organisms have been reported worldwide due to contaminatedduodenoscopes. In 2015, the United States Food and Drug Administration recommended the following supplemental enhancedsurveillance and reprocessing techniques (ESRT) to improve duodenoscope disinfection: (1) microbiological culture, (2) ethyleneoxide sterilization, (3) liquid chemical sterilant processing system, and (4) double high-level disinfection. A systematic review andmeta-analysis was performed to assess the impact of ESRT on the contamination rates. Methods: A thorough and systematic search was performed across several databases and conference proceedings from inceptionuntil January 2021, and all studies reporting the effectiveness of various ESRTs were identified. The pooled contamination rates ofpost-ESRT duodenoscopes were estimated using the random effects model. Results: A total of seven studies using various ESRTs were incorporated in the analysis, which included a total of 9,084 post-ESRTduodenoscope cultures. The pooled contamination rate of the post-ESRT duodenoscope was 5% (95% confidence interval [CI]:2.3%–10.8%, inconsistency index [I2]=97.97%). Pooled contamination rates for high-risk organisms were 0.8% (95% CI: 0.2%–2.7%,I2=94.96). Conclusions: While ESRT may improve the disinfection process, a post-ESRT contamination rate of 5% is not negligible. Ongoingefforts to mitigate the rate of contamination by improving disinfection techniques and innovations in duodenoscope design toimprove safety are warranted.

참고문헌 (27)

  1. "Reducing the risk of infection from reprocessed duodenoscopes" (2019)[URL]
  2. "Health, Education, Labor, Pensions Committee. Preventable tragedies: superbugs and how ineffective monitoring of medical device safety fails patients" (2016)[URL]
  3. "Design of endoscopic retrograde cholangiopancreatography (ERCP) duodenoscopes may impede effective cleaning: FDA safety communication" (2015)[URL]
  4. "Supplemental measures to enhance duodenoscope reprocessing:FDA safety communication" (2015)[URL]
  5. "522 Postmarket Surveillance Studies Database" (2016)[URL]
  6. "522 Postmarket Surveillance Studies Database" (2016)[URL]
  7. Rauwers AW, "High prevalence rate of digestive tract bacteria in duodenoscopes : a nationwide study", Gut, 67, pp.1637~1645 (2018)
  8. Ross AS, "A quarantine process for the resolution of duodenoscope-associated transmission of multidrug-resistant escherichia coli", Gastrointest Endosc, 82, pp.477~483 (2015)
  9. Mark JA, "Results of duodenoscope culture and quarantine after manufacturer-recommended cleaning process", Gastrointest Endosc, 91, pp.1328~1333 (2020)
  10. Bartles RL, "A randomized trial of single versus double high-level disinfection of duodenoscopes and linear echoendoscopes using standard automated reprocessing", Gastrointest Endosc, 88, pp.306~313 (2018)
  11. Gromski MA, "Double high-level disinfection versus liquid chemical sterilization for reprocessing of duodenoscopes used for ERCP : a prospective randomized study", Gastrointest Endosc, 93, pp.927~931 (2021)
  12. Higa JT, "Optimizing duodenoscope reprocessing : rigorous assessment of a culture and quarantine protocol", Gastrointest Endosc, 88, pp.223~229 (2018)
  13. Naryzhny I, "Impact of ethylene oxide gas sterilization of duodenoscopes after a carbapenem-resistant Enterobacteriaceae outbreak", Gastrointest Endosc, 84, pp.259~262 (2016)
  14. Rex DK, "A double-reprocessing high-level disinfection protocol does not eliminate positive cultures from the elevators of duodenoscopes", Endoscopy, 50, pp.588~596 (2018)
  15. Snyder GM, "Randomized comparison of 3high-level disinfection and sterilization procedures for duodenoscopes", Gastroenterology, 153, pp.1018~1025 (2017)
  16. Higgins JPT, "Cochrane handbook for systematic reviews of interventions. Version 5.1.0" (2011), Cochrane
  17. Moher D, "Preferred reporting items for systematic reviews and meta-analyses : the PRISMA statement", Ann Intern Med, 151, pp.264~269 (2009)
  18. Stang A, "Critical evaluation of the Newcastle-Ottawa scale for the assessment of the quality of nonrandomized studies in meta-analyses", Eur J Epidemiol, 25, pp.603~605 (2010)
  19. "Duodenoscope surveillance sampling and culturing protocols" (2018)[URL]
  20. DerSimonian R, "Meta-analysis in clinical trials", Controlled Clinical Trials, 7, pp.177~188 (1986)
  21. Guyatt GH, "GRADE guidelines : 7. Rating the quality of evidence--inconsistency", J Clin Epidemiol, 64, pp.1294~1302 (2011)
  22. "The FDA continues to remind facilities of the importance of following duodenoscope reprocessing instructions: FDA safety communication" (2019)[URL]
  23. Larsen S, "Rate and impact of duodenoscope contamination : a systematic review and meta-analysis", EClinicalMedicine, 25, pp.100451~ (2020)
  24. "The FDA is recommending transition to duodenoscopes with innovative designs to enhance safety: FDA Safety Communication" (2019)[URL]
  25. Pasricha PJ, "Novel and effective disposable device that provides 2-way protection to the duodenoscope from microbial contamination", Gastrointest Endosc, 92, pp.199~208 (2020)
  26. Ross AS, "Novel single-use duodenoscope compared with 3 models of reusable duodenoscopes for ERCP : a randomized bench-model comparison", Gastrointest Endosc, 91, pp.396~403 (2020)
  27. Bang JY, "Equivalent performance of single-use and reusable duodenoscopes in a randomised trial", Gut, 70, pp.838~844 (2021)

관련 논문AI추천