본문 바로가기
KERD
논문KCI 등재

임플란트 고정체의 나사산 형태와 하중조건에 따른 응력분석

Finite Element Stress Analysis of the Implant Fixture According to the Thread Configuration and the Loading Condition

2

조회

0

다운로드

0

인용

0

참고문헌

주제분류치의학

초록

Abstract

The purpose of this study was to compare the v-shape thread with the square shape thread of fixture in the view of stress distribution pattern using finite element stress analysis. The finite element model was designed with the parallel placement of two standard fixtures(4.0 mm diameter × 11.5 mm length) on the region of mandibular 1st and 2nd molars. Three dimensional finite element model was created with the components of the implant and surrounding bone. This study simulated loads of 200 N at the central fossa in a axial direction (load A), 200 N at the buccal offset load that is 2 mm apart from central fossa in a axial direction (load B), 200 N at the buccal offset load that was 4 mm apart from central fossa in a axial direction (load C). These forces of load A',B',C' were applied to a 15° inward oblique direction at that same site with 200 N. Von Mises stress values were recorded and compared in the supporting bone, fixture, and abutment screw. The following results have been made based on this study : 1. The highest stress concentration occurred at the cervical region of the implant fixture. 2. Von Mises stress value of off-site region was higher than that of central fossa region. 3. Square shape thread type showed more even stress distribution in the vertical and oblique force than V-shape thread type. 4. Stress distribution was the most effective in the case of buccal offset load (2, 4 mm distance from central fossa) in the square shape thread type. 5. V-shape thread type revealed higher von Mises stress value than square shape thread type in all environmental condition. The results from numerical analyses concluded that square shape thread type had the lower destructive stress and more stress distribution between the fixture and bone interface than V-shape thread type. Therefore, square shape thread type was regarded as optimal thread configuration in biomechanical concepts.

참고문헌 (11)

  1. "The biomechanics of force distribution in implant-supported prostheses" (1993)
  2. "Force transfer in implant dentistry" (1992), basic concepts and principles
  3. "In vivo bone response to biomechanical loading at the bone", dental-implant interface, 1999, ~[DOI]
  4. "Finite element analysis of thread-form configurations in a stepped screw implant Journal of Oral Rehabilitation 31", 31, pp.233~239 (2004)
  5. "thread-form parameter optimization for implant body design", 19(4) (1998)
  6. "Applications of finite element analysis in implant dentistry a review of literatures" (2001)
  7. "Finite element analysis of a stepped screw osseointegrated dental implant" (2004), In press
  8. "Bone stress distribution for three endosseous implants" (1989)
  9. "Evaluation of design parameters of osseointegrated dental implants using finite element analysis Journal of Oral Rehabilitation 29", pp.565~574 (2002)
  10. "Finite element stress and strain analysis of the bone surrounding a dental implant effect of variations in bone modulus Department of Prosthodontics College of Dentistry", pp.233 501 82~41 759 (1992), Chosun University 421
  11. "Finite element stress and strain analysis of the bone surrounding a dental implant effect of variations in bone modulus Department of Prosthodontics College of Dentistry", pp.233 501 82~41 759 (1992), Chosun University 421

관련 논문AI추천

이 간행물의 다른 논문 (8건)