Development of the rapid preconcentration method for determination of actinides in large volume seawater sample using Actinide resin
0
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
A simple and rapid preconcentration method of actinide from seawater using Actinide resin was developed and tested with the seawater spiked with a known U and Th. The developed method of Actinide resin based on column chromatography is less time-consuming and requires less labor compared with a typical co-precipitation technique for preconcentration of actinides. U and Th, which are relatively weak-bonded with Actinide resin among actinides, were used to determine the optimum flow rate of seawater sample and evaluate the capacity of Actinide resin to concentrate actinides from seawater. A flow rate of 50 mL min−1 was available with Actinide resin 2 mL (BV, bed volume). When 5 or 10 L of seawater containing U were loaded on Actinide resin (2 mL, BV) at 50 mL min−1, the recovery of U was 93 % and 86 %, respectively. For extraction of actinides bound with Actinide resin, we compared three methods: solvent extraction, ashing-acid digestion, and ashingmicrowave digestion. Ashing-microwave digestion method shows the best performance of which is the recovery of 100 % for U and 81 % for Th. For the preconcentration of actinides in 200 L of seawater, a typical coprecipitation method requires 2-3 days, but the developed method in this study is achieved the high recovery of actinides within 12 h.
Abstract
A simple and rapid preconcentration method of actinide from seawater using Actinide resin was developed and tested with the seawater spiked with a known U and Th. The developed method of Actinide resin based on column chromatography is less time-consuming and requires less labor compared with a typical co-precipitation technique for preconcentration of actinides. U and Th, which are relatively weak-bonded with Actinide resin among actinides, were used to determine the optimum flow rate of seawater sample and evaluate the capacity of Actinide resin to concentrate actinides from seawater. A flow rate of 50 mL min−1 was available with Actinide resin 2 mL (BV, bed volume). When 5 or 10 L of seawater containing U were loaded on Actinide resin (2 mL, BV) at 50 mL min−1, the recovery of U was 93 % and 86 %, respectively. For extraction of actinides bound with Actinide resin, we compared three methods: solvent extraction, ashing-acid digestion, and ashingmicrowave digestion. Ashing-microwave digestion method shows the best performance of which is the recovery of 100 % for U and 81 % for Th. For the preconcentration of actinides in 200 L of seawater, a typical coprecipitation method requires 2-3 days, but the developed method in this study is achieved the high recovery of actinides within 12 h.
참고문헌 (28)
- Sources and effects of ionizing radiation, "United Nations Scientific Committee on the Effects of Atomic Radiation", UNSCEAR, ~ (2010)
- K. Buesseler, , Annu. Rev. Mar. Sci, 9, pp.173~203 (2017)
- "Report on Treated Water Disposal"[URL]
- Marine Environmental Radioactivity Survey, "C.S. Kim", Korea Institute of Nuclear Safety, ~, pp.KINS (2018), ER-092
- L. León Vintró, , Sci. Total. Environ, 237-238, pp.77~91 (1999)
- S. H. Lee, , J. Radioanal. Nucl. Chem, 248(3), pp.757~764 (2001)
- Q. Cehn, , J. Radioanal. Nucl. Chem, 253(3), pp.451~458 (2002)
- W. Men, , Sci. Rep, 8, pp.1892~ (2018)
- S. L. Maxwell, , J. Radioanal. Nucl. Chem, 300, pp.1175~1189 (2014)
- E. P. Horwitz, , React. Funct. Polym, 33, pp.25~36 (1997)
- W. C. Burnett, , J. Radioanal. Nucl. Chem, 226, pp.121~127 (1997)
- I. W. Croudace, , Anal. Chim. Acta, 577, pp.111~118 (2006)
- N. Navarro, , Appl. Radiat. Isot, 61, pp.287~291 (2004)
- R. H. Sankhe, , J. Radioanal. Nucl. Chem, 302, pp.617~622 (2014)
- S. M. Pike, , J. Radioanal. Nucl, Chem, 296, pp.369~374 (2013)
- H. Kim, , Talanta, 217 (2020)[DOI]
- H. Tazoe, , Talanta, 152, pp.219~227 (2016)
- S. L. Maxwell, , J. Radioanal. Nucl. Chem, 303, pp.709~717 (2015)
- 박지영, "원자력 시설 주변 환경 감시를 위한 토양 중 우라늄 동위원소 신속 분석법 확립", 분석과학, 31(3), pp.134~142 (2018), 한국분석과학회[DOI]
- L. Cao, , Talanta, 151, pp.30~41 (2016)
- S. A. Yim, , J. Radat. Prot, 35(3), pp.117~123 (2010)
- J. Zheng, , Appl. Radiat. Isot, 70, pp.1944~1948 (2012)
- E. Braysher, , J. Radioanal. Nul. Chem, 321, pp.183~196 (2019)
- S. Maischak, "Solid-phase extraction for the separation of actinides from radioactive waste", WM’01 Conference (2001)
- D. F. Lupton, , X-Ray Spectrom, 26, pp.132~140 (1997)
- J. Lehto, "Chemistry and Analysis of Radionuclides", pp.255~ (2012), WILEY-VCH
- 임종명, "실내 건축자재 중 천연방사성핵종의 정량분석법 연구", 분석과학, 30(5), pp.252~261 (2017), 한국분석과학회[DOI]
- H. Kim, , J. Radioanal. Nucl. Chem, 304, pp.321~327 (2015)