요실금 치료용 자기 자극기의 자속밀도 및 유도전류 측정
Measurement of Magnetic Flux and Induced Current in Magnetic Stimulation for Urinary Incontinence Treatment
1
조회
0
다운로드
0
인용
0
참고문헌
키워드
초록
A simple method for measuring magnetic flux and induced current in magnetic nerve stimulation for urinary incontinence treatment is proposed. Unlike electric nerve stimulation, direct measurement of the induced current in magnetic nerve stimulation is impossible. Since induced currents stimulate nerves or muscles in magnetic nerve stimulation, measuring induced current is very important in validating stimulation efficacy and securing safety. The magnetic flux measuring system is composed of 6 layers with pick-up coils of 7 by 7 in each layer, and the induced current measuring system is composed of 6 layers with 7 concentric circular coils in each layer. The proposed method can be used in the design or performance test of a magnetic nerve stimulator for many clinical applications such as urinary incontinence treatment, activation of peripheral nerves, and transcranial magnetic stimulation.
Abstract
A simple method for measuring magnetic flux and induced current in magnetic nerve stimulation for urinary incontinence treatment is proposed. Unlike electric nerve stimulation, direct measurement of the induced current in magnetic nerve stimulation is impossible. Since induced currents stimulate nerves or muscles in magnetic nerve stimulation, measuring induced current is very important in validating stimulation efficacy and securing safety. The magnetic flux measuring system is composed of 6 layers with pick-up coils of 7 by 7 in each layer, and the induced current measuring system is composed of 6 layers with 7 concentric circular coils in each layer. The proposed method can be used in the design or performance test of a magnetic nerve stimulator for many clinical applications such as urinary incontinence treatment, activation of peripheral nerves, and transcranial magnetic stimulation.
참고문헌 (34)
- S.-J. Oh, "Prevalence of urinary incontinence and incontinence-related quality of life in Korean women: a population-based study", J. Korean Continence Society, 7, pp.73~80 (2003)
- H. J. Song, "Epidemiologic study of urinary incontinence for Korean women over 30 years old", J. Korean Continence Society, 5, pp.24~38 (2001)
- J. W. Yarnell, "The prevalence and severity of urinary incontinence in women", J. Epidemiol. Community Health, 35, pp.71~74 (1981)
- J. T. Anger, "The prevalence of urinary incontinence among community dwelling adult women: results from the National Health and Nutrition Examination Survey", J. Urol, 175, pp.601~604 (2006)
- 김노수, "여성 복압성요실금에 대한 Tension-free Vaginal Tape(TVT) 술식의 장기추적 결과", Korean Journal of Urology, 47(7), pp.729~733 (2006), 대한비뇨기과학회
- C. K. Doo, "Changes in overactive bladder symptoms after discontinuation of a successful treatment with antimuscarinic agent: a prospective trial", Korean J. Urol, 46, pp.713~718 (2006)
- H. S. Han, "The effect of electrical stimulation and biofeedback for female urinary incontinence", Korean J. Urol, 42, pp.1063~1067 (2001)
- T. Yokoyama, "Comparative study of effects of extracorporeal magnetic innervation versus electrical stimulation for urinary incontinence after radical prostatectomy", Urology, 63, pp.264~267 (2004)
- N. Bölükbas, "Effectiveness of functional magnetic versus electrical stimulation in women with urinary incontinence", Eur. Med. Phys, 41, pp.297~301 (2005)
- A. C. Wang, "Single-blind, random- ized trial of pelvic floor muscle training, biofeedback-assisted pelvic floor muscle training, and electrical stimulation in the management of overactive bladder", Urology, 63, pp.61~66 (2004)
- J. M. Holroyd-Leduc, "Management of urinary incontinence in women: scientific review", J. Am. Med. Assoc, 291, pp.986~995 (2004)
- A. T. Barker, "Non-invasive magnetic stimulation of the human motor cortex", Lancet, 1, pp.1106~1107 (1985)
- L. A. Geddes, "History of magnetic stimulation of the nervous system", J. Clin. Neurophysiol, 8, pp.3~9 (1991)
- B. H. Han, "Multi-channel magnetic stimulation system design considering mutual couplings among the stimulation coils", IEEE Trans. Biomed. Eng, 51, pp.812~817 (2004)
- N. T. M. Galloway, "Extracorporeal magnetic innervation therapy for stress urinary incontinence", Urology, 53, pp.1108~1111 (1999)
- J. Ruohonen, "Theory of multichannel magnetic stimulation: Toward functional neuromuscular rehabilitation", IEEE Trans. Biomed. Eng, 46, pp.646~651 (1999)
- B. H. Han, "Multi-channel magnetic nerve stimulation system with small size power supplies", J. Biomed. Eng. Res, 22, pp.139~144 (2001)
- 한병희, "자계 거울 효과를 이용한 신경 자극 코일", 의공학회지, 23(4), pp.287~293 (2002), 대한의용생체공학회
- B. H. Han, "Some technical aspects of magnetic stimulation coil design with the ferroma- gnetic effect", Med. Biol. Eng. Comput, 41, pp.516~518 (2003)
- A. Thielscher, "Electric field properties of two commercial figure-8 coils in TMS: calculation of focality and efficiency", Clin. Neurophysiol, 115, pp.1697~1708 (2004)
- V. W. Lin, "Magnetic coil design considerations for functional magnetic stimulation", IEEE Trans. Biomed. Eng, 47, pp.600~610 (2000)
- T. Wagner, "Intracranial measurement of current densities induced by transcranial magnetic stimulation in the human brain", Neuroscience Letters, 354, pp.91~94 (2004)
- A. T. Barker, "An introduction to the basic principles of magnetic nerve stimulation", J. Clin. Neurophysiol, 8, pp.26~37 (1991)
- K. R. Mills, "Magnetic stimulation of the human nervous system", pp.7~26 (1999), Oxford Univ. Press
- "510(k) summary for Model 1000 Muscle Stimulator System"[URL]
- J. D. Bourland, "Transchest magnetic (eddy-current) stimulation of the dog heart", Med. Biol. Eng. Comput, 28, pp.196~198 (1990)
- L. M. Schradera, "Seizure incidence during single- and paired-pulse transcranial magnetic stimulation (TMS) in individuals with epilepsy", Clin. Neurophysiol, 115, pp.2728~2737 (2004)
- International Electrotechnical Commission (IEC), "Medical electrical equipment - Part 2-33: Particular requirements for the safety of magnetic resonance equipment for medical diagnosis", in IEC 60601-2-33, ~ (2006)
- A. Rotenberg, "Minimal heating of titanium skull plates during 1Hz repetitive transcranial magnetic stimulation", Clin. Neurophysiol, 118, pp.2536~2538 (2007)
- P. Bernardi, "Specific absorption rate and temperature elevation in a subject exposed in the far-field of radio-frequency sources operating in the 10-900-MHz Range", IEEE Trans. Biomed. Eng, 50, pp.295~304 (2003)
- J. G. Webster, "Electrical Impedance Tomography", pp.8~20 (1990), Adam Hilger
- K. Yamazaki, "On the method of investigating human exposure to nonuniform magnetic field", Electr. Eng. Japan, 164, pp.1~11 (2008)
- R. Scorretti, "Modeling of induced current into the human body by low-frequency magnetic field from experimental data", IEEE Trans. Magnetics, 41, pp.1992~1995 (2005)
- J. P. Reilly, "Magnetic field excitation of peripheral nerves and the heart: a comparison of thresholds", Med. Biol. Eng. Comput, 29, pp.571~579 (1991)