Continuous body temperature monitoring is useful and essential in diverse medical procedures such asinfection onset detection, therapeutic hypothermia, circadian rhythm monitoring, sleep disorder assessment, andgynecological research. However, the existing thermometers are too invasive or intrusive to be applied to long-termbody temperature monitoring. In our previous study, we invented the bi-medium deep body thermometer which cannoninvasively and continuously monitor deep tissue temperature. And the ratio of thermal resistances expressed asK-value should be obtained to estimate body temperature with the thermometer and it can be different under variousmeasurement environments. Although the device was proven to be useful through preliminary simulation test andsmall group of human study, the experimental environment was restrictive in our previous approach. In this study,a finite element simulation was executed to obtain the K-value and evaluate the accuracy of bi-medium thermometerunder various measurement environments. In addition, K-value estimation equation was developed by analyzing theinfluence of 5 measurement environmental factors (medium length, medium height, tissue depth, blood perfusionrate, and ambient temperature) on K-value. The results revealed that the estimation accuracy of bi-medium deepbody thermometer based on computer simulation was very high (RMSE < 0.003oC) in various measurement environments. Also, bi-medium deep body thermometer based on K-value estimation equation showed relatively accurateresults (RMSE < 0.3oC) except for one case. Although the K-value estimation technology should be improved formore accurate body temperature estimation, the results of finite element simulation showed that bi-medium deepbody thermometer could accurately measure various tissue temperatures under diverse environments.

Abstract

Continuous body temperature monitoring is useful and essential in diverse medical procedures such asinfection onset detection, therapeutic hypothermia, circadian rhythm monitoring, sleep disorder assessment, andgynecological research. However, the existing thermometers are too invasive or intrusive to be applied to long-termbody temperature monitoring. In our previous study, we invented the bi-medium deep body thermometer which cannoninvasively and continuously monitor deep tissue temperature. And the ratio of thermal resistances expressed asK-value should be obtained to estimate body temperature with the thermometer and it can be different under variousmeasurement environments. Although the device was proven to be useful through preliminary simulation test andsmall group of human study, the experimental environment was restrictive in our previous approach. In this study,a finite element simulation was executed to obtain the K-value and evaluate the accuracy of bi-medium thermometerunder various measurement environments. In addition, K-value estimation equation was developed by analyzing theinfluence of 5 measurement environmental factors (medium length, medium height, tissue depth, blood perfusionrate, and ambient temperature) on K-value. The results revealed that the estimation accuracy of bi-medium deepbody thermometer based on computer simulation was very high (RMSE < 0.003oC) in various measurement environments. Also, bi-medium deep body thermometer based on K-value estimation equation showed relatively accurateresults (RMSE < 0.3oC) except for one case. Although the K-value estimation technology should be improved formore accurate body temperature estimation, the results of finite element simulation showed that bi-medium deepbody thermometer could accurately measure various tissue temperatures under diverse environments.

참고문헌 (18)

  1. Davidovits P., "Physics in biology and medicine", pp.145~ (2007), Academic press
  2. D. S. Moran., "Evaluation of different levels of hydration using a new physiological strain index", Am. J. Physiol. Regul. Integr. Comp. Physiol., 275(3), pp.R854~R860 (1998)
  3. D. Darwent., "The validity of temperature-sensitive ingestible capsules for measuring core body temperature in laboratory protocols", Chronobiol. Int, 28(8), pp.719~726 (2011)
  4. F. C. Baker., "Sleep and 24 hour body temperatures : a comparison in young men, naturally cycling women and women taking hormonal contraceptives", J. Physiol, 530(3), pp.565~574 (2001)
  5. K. G. Ng., "Evaluation of the Cadi ThermoSENSOR Wireless Skin-Contact Thermometer Against Ear and Axillary Temperatures in Children", J. Pediatr. Nurs, 25(3), pp.176~186 (2010)
  6. H. C. GUNGA., "A new non-invasive device to monitor core temperature on earth and in space", Ann. Kinesiol, 3(1), pp.47~60 (2012)
  7. R. H. Fox., "A new technique for monitoring the deep body temperature in man from the intact skin surface", J. Physiol, 212(2), pp.8~10 (1971)
  8. K. Kitamura., "Development of a new method for the noninvasive measurement of deep body temperature without a heater", Med. Eng. Phys, 32(1), pp.1~6 (2010)
  9. Huang, M., "Theoretical simulation of the dualheat-flux method in deep body temperature measurements", in 32nd Conf. Proc. IEEE Eng. Med. Biol. Soc., Buenos Aires, Argentina, pp.561~564 (2010)
  10. S. Y. Sim., "Simulation study of a newly designed structure for monitoring deep body temperature", in Proc. 9th Intern. Conf. Ubiquitous Healthcare, Gyeongju, Korea (2012)
  11. S. Y. Sim., "Accuracy evaluation of noninvasive and nonintrusive body temperature monitor", in Proc. 47th conference of the KOSOMBE, Gumi, Korea, pp.63~64 (2013)
  12. J.P. Holman, "Heat transfer 6th Edition", pp.27~29 (1986), McGraw
  13. H. H. Pennes, "Analysis of tissue and arterial blood temperatures in the resting human forearm", J. Appl. Physiol, 85(1), pp.5~34 (1998)
  14. "Properties of non-metal"[URL]
  15. M. Masaru., "Non-invasive monitoring of core body temperature rhythms over 72h in 10 bedridden elderly patients with disorders of consciousness in a Japanese hospital : A pilot study", Arch. Gerontol. Geriatr, 57(3), pp.428~432 (2013)
  16. Zhu. L., "Theoretical simulation of temperature distribution in the brain during mild hypothermia treatment for brain injury", Med. Biol. Eng. Comput, 39(6), pp.681~687 (2001)
  17. I. A. Brezovich., "Temperature distributions in hyperthermia by electromagnetic induction : a theoretical model for the thorax", Med. Phys, 10(1), pp.57~65 (1983)
  18. H. L. Staniak., "Subcutaneous tissue thickness is an independent predictor of image noise in cardiac CT", Arq. Bras. Cardiol, 102(1), pp.86~92 (2014)

관련 논문AI추천

이 간행물의 다른 논문 (7건)