Performance Evaluation of Cartridge-Type Blood Gas Analyzer: i-Smart 300

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

Background: Blood gas analysis plays a crucial role in critical care settings, and immediate and precise analysis improves clinical outcomes through prompt treatment. We evaluated the performance of a cartridge-type blood gas analyzer, i-Smart 300 (i-SENS, Korea), according to the Clinical and Laboratory Standard Institute (CLSI) guidelines and compared it to a conventional blood gas analyzer. Methods: The precision was evaluated according to CLSI EP5-A3. The i-Smart 300 was compared to the Stat Profile Critical Care Xpress (STP CCX) (Nova CCX; Nova Biomedical, USA) according to CLSI EP9-A3 using the following eight parameters: pH, partial carbon dioxide pressure, partial oxygen pressure, sodium, potassium, chloride, ionized calcium, and hematocrit. Linearity was determined using five levels of control materials according to CLSI EP6-A. Results: Within-run precision and total precision, demonstrated as coefficients of variation, ranged from 0.02 to 2.50% and from 0.05 to 3.46%, respectively. Correlation analysis yielded a correlation coefficient from 0.966 to 0.996 between the i-Smart 300 and the conventional analyzer (Nova CCX). The i-Smart 300 showed excellent linearity at eight parameters with acceptable percent recovery. Conclusions: The i-Smart 300, a portable cartridge-type blood gas analyzer, showed high precision and good correlation with a traditional bench-top blood gas analyzer. It could be useful in critical care settings.

Abstract

Background: Blood gas analysis plays a crucial role in critical care settings, and immediate and precise analysis improves clinical outcomes through prompt treatment. We evaluated the performance of a cartridge-type blood gas analyzer, i-Smart 300 (i-SENS, Korea), according to the Clinical and Laboratory Standard Institute (CLSI) guidelines and compared it to a conventional blood gas analyzer. Methods: The precision was evaluated according to CLSI EP5-A3. The i-Smart 300 was compared to the Stat Profile Critical Care Xpress (STP CCX) (Nova CCX; Nova Biomedical, USA) according to CLSI EP9-A3 using the following eight parameters: pH, partial carbon dioxide pressure, partial oxygen pressure, sodium, potassium, chloride, ionized calcium, and hematocrit. Linearity was determined using five levels of control materials according to CLSI EP6-A. Results: Within-run precision and total precision, demonstrated as coefficients of variation, ranged from 0.02 to 2.50% and from 0.05 to 3.46%, respectively. Correlation analysis yielded a correlation coefficient from 0.966 to 0.996 between the i-Smart 300 and the conventional analyzer (Nova CCX). The i-Smart 300 showed excellent linearity at eight parameters with acceptable percent recovery. Conclusions: The i-Smart 300, a portable cartridge-type blood gas analyzer, showed high precision and good correlation with a traditional bench-top blood gas analyzer. It could be useful in critical care settings.

참고문헌 (17)

  1. Uyanik M, "Comparison of blood gas, electrolyte and metabolite results measured with two different blood gas analyzers and a core laboratory analyzer", Scand J Clin Lab Invest, 75, pp.97~105 (2015)
  2. De Koninck AS, "Analytical performance evaluation of four cartridge-type blood gas analyzers", Clin Chem Lab Med, 50, pp.1083~1091 (2012)
  3. Einstein MH, "Clinical evaluation of the cartridge-based GeneXpert human papillomavirus assay in women referred for colposcopy", J Clin Microbiol, 52, pp.2089~2095 (2014)
  4. Leino A, "Interchangeability of blood gas, electrolyte and metabolite results measured with point-of-care, blood gas and core laboratory analyzers", Clin Chem Lab Med, 49, pp.1187~1191 (2011)
  5. CLSI, "Blood Gas and pH Analysis and Related Measurements; Approved Guideline", Clinical and Laboratory Standards Institute, ~ (2009)
  6. CLSI, "Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline", Clinical and Laboratory Standards Institute, ~ (2014)
  7. CLSI, "Preliminary evaluation of quantitative clinical laboratory measurement procedures; Approved Guideline", Clinical and Laboratory Standards Institute, ~ (2006)
  8. CLSI, "Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline", Clinical and Laboratory Standards Institute, ~ (2003)
  9. CLSI, "Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Approved Guideline" (2013), Clinical and Laboratory Standards Institute
  10. Ricós C, "Current databases on biological variation: pros, cons and progress", Scand J Clin Lab Invest, 59, pp.491~500 (1999)
  11. "Desirable specification for total error, imprecision, and bias, derived from intra-and inter-individual biologic variation" (2014)[URL]
  12. Nichols JH, "Executive summary. The National Academy of Clinical Biochemistry Laboratory Medicine Practice Guideline: evidence-based practice for point-of-care testing", Clin Chim Acta, 379, pp.14~28 (2007)
  13. Fraser CG, "Generation and application of data on biological variation in clinical chemistry", Crit Rev Clin Lab Sci, 27, pp.409~437 (1989)
  14. Cotlove E, "Biological and analytic components of variation in long-term studies of serum constituents in normal subjects. 3. Physiological and medical implications", Clin Chem, 16, pp.1028~1032 (1970)
  15. Harris EK, "Statistical principles underlying analytic goal-setting in clinical chemistry", Am J Clin Pathol, 72(2 Suppl), pp.374~382 (1979)
  16. Fraser CG., "General strategies to set quality specifications for reliability performance characteristics", Scand J Clin Lab Invest, 59, pp.487~490 (1999)
  17. Seok YM, "Evaluation of the i-Smart 30 Point-of-care analyzer for use in clinical laboratory settings", J Lab Med Qual Assur, 33, pp.25~30 (2011)

관련 논문AI추천

이 간행물의 다른 논문 (9건)