Recommendations for the Verification of Quantitative Molecular Hemato-Oncology Tests
2
조회
0
다운로드
1
인용
0
참고문헌
키워드
초록
Quantitative molecular genetic tests are increasingly used for the detection and quantification of target molecules or genetic alterations. When introducing a new assay into clinical laboratories, it is necessary to verify the manufacturers’ claimed performance characteristics within individual laboratories. Appropriate assay verification procedures are essential to ensure the quality of test results in clinical laboratories. This study aimed to provide recommendations for the verification of quantitative molecular genetic testing focused on the hemato-oncology field in clinical genetic laboratories. Based on a literature review, we provide recommendations for the performance verification of quantitative molecular hemato-oncology tests. The performance characteristic elements that comprise the verification procedures are presented and exemplified. These recommendations can assist individual clinical laboratories in verifying quantitative molecular diagnostic assays
Abstract
Quantitative molecular genetic tests are increasingly used for the detection and quantification of target molecules or genetic alterations. When introducing a new assay into clinical laboratories, it is necessary to verify the manufacturers’ claimed performance characteristics within individual laboratories. Appropriate assay verification procedures are essential to ensure the quality of test results in clinical laboratories. This study aimed to provide recommendations for the verification of quantitative molecular genetic testing focused on the hemato-oncology field in clinical genetic laboratories. Based on a literature review, we provide recommendations for the performance verification of quantitative molecular hemato-oncology tests. The performance characteristic elements that comprise the verification procedures are presented and exemplified. These recommendations can assist individual clinical laboratories in verifying quantitative molecular diagnostic assays
참고문헌 (23)
- Heuser M, "2021 Update on MRD in acute myeloid leukemia : a consensus document from the European LeukemiaNet MRD Working Party", Blood, 138, pp.2753~2767 (2021)
- Mazaika E, "Digital droplet PCR : CNV analysis and other applications", Curr Protoc Hum Genet, 82, pp.7.24.1~7.24.13 (2014)
- Halling KC, "Test veriflcation and validation for molecular diagnostic assays", Arch Pathol Lab Med, 136, pp.11~13 (2012)
- Evaluation of detection capability for clinical laboratory measurement procedures; Approved guideline—Second edition, "Clinical and Laboratory Standards Institute", Clinical and Laboratory Standards Institute, ~, pp.CLSI document EP17-A2 (2012)
- "Molecular Diagnostic Assay Validation : Update to the 2009 AMP Molecular Diagnostic Assay Validation White Paper" (2014)[URL]
- Molecular methods for clinical genetics and oncology testing; Approved guideline—Third edition, "Clinical and Laboratory Standards Institute", Clinical and Laboratory Standards Institute, ~, pp.CLSI document MM01-A3 (2012)
- Quality management for molecular genetic testing; Approved guideline, "Clinical and Laboratory Standards Institute", Clinical and Laboratory Standards Institute, ~, pp.CLSI document MM20-A (2012)
- Evaluation of linearity of quantitative measurement procedures, "Clinical and Laboratory Standards Institute", Clinical and Laboratory Standards Institute, ~, pp.CLSI guideline EP06-Ed2 (2020)
- User veriflcation of precision and estimation of bias; Approved guideline—Third edition, "Clinical and Laboratory Standards Institute", Clinical and Laboratory Standards Institute, ~, pp.CLSI document EP15-A3 (2014)
- Deflning, establishing, and verifying reference intervals in the clinical laboratory; Approved guideline—3rd edition, "Clinical and Laboratory Standards Institute", Clinical and Laboratory Standards Institute, ~, pp.CLSI guideline EP28-A3c (2010)
- "Standard: Establishment and veriflcation of performance speciflcations. 42 CFR § 493.1253, Oct. 1, 2015"[URL]
- College of American Pathologists, "Molecular pathology checklist" (2021), College of American Pathologists
- Mattocks CJ, "A standardized framework for the validation and veriflcation of clinical molecular genetic tests", Eur J Hum Genet, 18, pp.1276~1288 (2010)
- Jennings L, "Recommended principles and practices for validating clinical molecular pathology tests", Arch Pathol Lab Med, 133, pp.743~755 (2009)
- Hee-Jung Chung, "Performance Evaluation of the QXDx BCR-ABL %IS Droplet Digital PCR Assay", Annals of Laboratory Medicine, 40(1), pp.72~75 (2020), 대한진단검사의학회[DOI]
- Broeders S, "Guidelines for validation of qualitative real-time PCR methods", Trends Food Sci Technol, 37, pp.115~126 (2014)
- Jennings LJ, "Design and analytic validation of BCR-ABL1 quantitative reverse transcription polymerase chain reaction assay for monitoring minimal residual disease", Arch Pathol Lab Med, 136, pp.33~40 (2012)
- Armbruster DA, "Limit of blank, limit of detection and limit of quantitation", Clin Biochem Rev, 29(Suppl 1), pp.S49~S52 (2008)
- Shrivastava A, "Methods for the determination of limit of detection and limit of quantitation of the analytical methods", Chron Young Sci, 2, pp.21~25 (2011)
- "Rare Mutation Detection Best Practices Guidelines"[URL]
- Yuan D, "Droplet digital PCR for quantiflcation of PML-RARα in acute promyelocytic leukemia : a comprehensive comparison with real-time PCR", Anal Bioanal Chem, 411, pp.895~903 (2019)
- Wolstencroft EC, "Development of a quantitative real-time polymerase chain reaction assay for the detection of the JAK2 V617F mutation", J Mol Diagn, 9, pp.42~46 (2007)
- Schnittger S, "Development and validation of a real-time quantiflcation assay to detect and monitor BRAFV600E mutations in hairy cell leukemia", Blood, 119, pp.3151~3154 (2012)