음성 간섭을 감소시키도록 개선된 Norudia 당화알부민 시약의 성능평가

Performance Evaluation of a Modified Version of the Norudia Glycated Albumin Assay that Reduces Negative Interference

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

초록

Background: Glycated albumin (GA) is a biomarker of short-term glycemic status. Several cases of extremely low GA levels measured by Norudia GA assay kits (Sekisui Medical Co., Ltd., Tokyo, Japan) were inconsistent with clinical status. This study evaluated the analytical performance of the modified version of the Norudia GA that claims to have resolved the false negative issue. Methods: Precision and linearity were evaluated following the guidance of the Clinical & Laboratory Standards Institute (CLSI) EP05-A3 and EP06-A, respectively. A comparison study was performed against the Lucica GA-L (Asahi Kasei Pharma Corporation, Tokyo, Japan) based on CLSI EP09-A3. The temperature stability of GA was also assessed. The reference interval was verified following CLSI EP28-A3C. Results: Coefficients of variation in the precision analysis were all acceptable. Linearity assessment demonstrated a coefficient of determination (R2) of 0.998. The comparison study showed a high correlation coefficient (r) of 0.973 relative to the Lucica GA-L. Stability analysis revealed GA tended to increase with storage duration. The transference of the reference interval was verified. Negative interference was reduced in the modified version of the Norudia GA. Conclusions: The modified version of the Norudia GA assay showed comparable performance in measuring GA as well as reduced interference in samples that had shown false negative results with the original version of the Norudia GA. The modified assay may avoid negative interference putatively caused by anti-oxidative agents in clinical laboratories using the original version of the Norudia GA assay.

Abstract

Background: Glycated albumin (GA) is a biomarker of short-term glycemic status. Several cases of extremely low GA levels measured by Norudia GA assay kits (Sekisui Medical Co., Ltd., Tokyo, Japan) were inconsistent with clinical status. This study evaluated the analytical performance of the modified version of the Norudia GA that claims to have resolved the false negative issue. Methods: Precision and linearity were evaluated following the guidance of the Clinical & Laboratory Standards Institute (CLSI) EP05-A3 and EP06-A, respectively. A comparison study was performed against the Lucica GA-L (Asahi Kasei Pharma Corporation, Tokyo, Japan) based on CLSI EP09-A3. The temperature stability of GA was also assessed. The reference interval was verified following CLSI EP28-A3C. Results: Coefficients of variation in the precision analysis were all acceptable. Linearity assessment demonstrated a coefficient of determination (R2) of 0.998. The comparison study showed a high correlation coefficient (r) of 0.973 relative to the Lucica GA-L. Stability analysis revealed GA tended to increase with storage duration. The transference of the reference interval was verified. Negative interference was reduced in the modified version of the Norudia GA. Conclusions: The modified version of the Norudia GA assay showed comparable performance in measuring GA as well as reduced interference in samples that had shown false negative results with the original version of the Norudia GA. The modified assay may avoid negative interference putatively caused by anti-oxidative agents in clinical laboratories using the original version of the Norudia GA assay.

참고문헌 (20)

  1. Kisugi R, "Structural and glycation site changes of albumin in diabetic patient with very high glycated albumin", Clin Chim Acta, 382, pp.59~64 (2007)
  2. Koga M, "Glycated albumin; clinical usefulness", Clin Chim Acta, 433, pp.96~104 (2014)
  3. Bellia C, "Clinical usefulness of glycated albumin in the diagnosis of diabetes : results from an Italian study", Clin Biochem, 54, pp.68~72 (2018)
  4. Clinical, Laboratory Standards Institute, "Evaluation of precision of quantitative measurement procedures; Approved guideline-" (2014), Clinical and Laboratory Standards Institute
  5. "Desirable biological variation database specifications"[URL]
  6. Clinical, laboratory standards institute, "Evaluation of the linearity of quantitative measurement procedures: A statistical approach; Approved guideline" (2003), Clinical and Laboratory Standards Institute
  7. Ricós C, "Current databases on biological variation : pros, cons and progress", Scand J Clin Lab Invest, 59, pp.491~500 (1999)
  8. Takei I, "Committee on diabetes mellitus indices of the Japan society of clinical chemistry-recommended reference measurement procedure and reference materials for glycated albumin determination", Ann Clin Biochem, 53, pp.124~132 (2016)
  9. Clinical, Laboratory Standards Institute, "Measurement procedure comparison and bias estimation using patient samples; Approved guideline-" (2013), Clinical and Laboratory Standards Institute
  10. Krouwer JS, "Why Bland-Altman plots should use X, not(Y+X)", 2 when X is a reference method, 2008, pp.27, 778~780
  11. Clinical, Laboratory Standard Institute, "Defining, establishing, and verifying reference intervals in the clinical laboratory; Approved guideline-" (2010), Clinical and Laboratory Standards Institute
  12. Ha C, "Evaluation of the analytical performance of the Norudia GA Glycoalbumin Test", Lab Med Online, 11, pp.55~59 (2021)
  13. Choe W, "Performance of Norudia glycated albumin assay on multiple analytical platforms and comparison to Lucica assay", Clin Lab, 66, pp.2383~2387 (2020)
  14. Berth M, "Protein precipitation as a possible important pitfall in the clinical chemistry analysis of blood samples containing monoclonal immunoglobulins : 2 case reports and a review of the literature", Acta Clin Belg, 59, pp.263~273 (2004)
  15. Dimeski G, "Rare IgM interference with Roche", Hitachi Modular glucose and gamma-glutamyltransferase methods in heparin samples, 2005, pp.51, 2202~2204
  16. Alberti MO, "The pH of chemistry assays plays an important role in monoclonal immunoglobulin interferences", Pract Lab Med, 3, pp.8~16 (2015)
  17. Yukimasa N, "Interference with creatinine assay by IgM-λ monoclonal protein in lithium heparin blood collection tube from a malignant lymphoma patient", Int J Med Pharm Case Reports, 8, pp.1~7 (2016)
  18. Trinder P, "Determination of blood glucose using an oxidase-peroxidase system with a non-carcinogenic chromogen", J Clin Pathol, 22, pp.158~161 (1969)
  19. Ko YK, "The negative role of Vitamin C in total cholesterol and triglycerides tests", Lab Med Online, 13, pp.13~21 (2023)
  20. "Norudia GA package insert sheet"[URL]

관련 논문AI추천

이 간행물의 다른 논문 (11건)