Measurement of Skull Size on Computed Tomography Images for Developing a Bone Conduction Headset Suitable for the Korean Standard Head Size

1

조회

0

다운로드

2

인용

0

참고문헌

초록

Background and Objectives: We aimed to measure the head dimensions on computed tomography (CT) images, to compare them to directly measured head dimensions, and to predict a new parameter of bone thickness for aiding bone conduction implant (BCI) placement. Subjects and Methods: We reviewed the facial and mandibular bone CT images of 406 patients. Their head sizes were analyzed using five parameters included in the 6th Size Korea project, and they were divided into age groups (ranging from the 10s to the 80s). We compared the head length, head width, sagittal arc, bitragion arc, and head circumference in the CT and Size Korea groups. We also added the parameter bone thickness for aiding BCI placement. Results: All the head size parameters measured using CT were significantly smaller than those measured directly, with head length showing the smallest difference at 7.85 mm. The differences in the other four parameters between the two groups according to patient age were not statistically significantly different. Bone thickness had the highest value of 4.89±0.93 mm in the 70s and the lowest value of 4.10±0.99 mm in the 10s. Bone thickness also significantly correlated with head width (p=0.038). Conclusions: Our findings suggested that the CT and direct measurements yielded consistent data. Moreover, CT enabled the measurement of bone sizes, including bone thickness, that are impossible to measure directly. CT measurements may complement direct measurements in the Size Korea data when used for developing bone conduction hearing devices (BCIs and headsets) for the Korean population.

Abstract

Background and Objectives: We aimed to measure the head dimensions on computed tomography (CT) images, to compare them to directly measured head dimensions, and to predict a new parameter of bone thickness for aiding bone conduction implant (BCI) placement. Subjects and Methods: We reviewed the facial and mandibular bone CT images of 406 patients. Their head sizes were analyzed using five parameters included in the 6th Size Korea project, and they were divided into age groups (ranging from the 10s to the 80s). We compared the head length, head width, sagittal arc, bitragion arc, and head circumference in the CT and Size Korea groups. We also added the parameter bone thickness for aiding BCI placement. Results: All the head size parameters measured using CT were significantly smaller than those measured directly, with head length showing the smallest difference at 7.85 mm. The differences in the other four parameters between the two groups according to patient age were not statistically significantly different. Bone thickness had the highest value of 4.89±0.93 mm in the 70s and the lowest value of 4.10±0.99 mm in the 10s. Bone thickness also significantly correlated with head width (p=0.038). Conclusions: Our findings suggested that the CT and direct measurements yielded consistent data. Moreover, CT enabled the measurement of bone sizes, including bone thickness, that are impossible to measure directly. CT measurements may complement direct measurements in the Size Korea data when used for developing bone conduction hearing devices (BCIs and headsets) for the Korean population.

참고문헌 (18)

  1. Håkansson B, "Principal design and a psychoacoustical evaluation", Acta Otolaryngol, 100, pp.229~239 (1985)
  2. Tae Hoon Kong, "Evaluation of wireless Bluetooth devices to improve recognition of speech and sentences when using a mobile phone in bone conduction device recipients", European Archives of Oto-Rhino-Laryngology, 276(10), pp.2729~2737 (2019), Springer Science and Business Media LLC[DOI]
  3. Wolfe J, "Evaluation of a wireless audio streaming accessory to improve mobile telephone performance of cochlear implant users", Int J Audiol, 55, pp.75~82 (2016)
  4. Dimitriadis PA, "Single-center experience of over a hundred implantations of a transcutaneous bone conduction device", Otol Neurotol, 38, pp.1301~1307 (2017)
  5. Lee HJ, "Comparison of Korean and Japanese head and face anthropometric characteristics", Hum Biol, 80, pp.313~330 (2008)
  6. Hodgetts WE, "Effects of applied contact force and volume control setting on output force levels of the BAHA Softband", Int J Audiol, 45, pp.301~308 (2006)
  7. Ball R, "A comparison between Chinese and Caucasian head shapes", Appl Ergon, 41, pp.832~839 (2010)
  8. Lindley AA, "The relationship in neonates between clinically measured head circumference and brain volume estimated from head CT-scans", Early Hum Dev, 56, pp.17~29 (1999)
  9. Mudry A, "Historical background of bone conduction hearing devices and bone conduction hearing aids", Adv Otorhinolaryngol, 71, pp.1~9 (2011)
  10. Zhang L, "Bone conduction hearing in congenital aural atresia", Eur Arch Otorhinolaryngol, 273, pp.1697~1703 (2016)
  11. Hougaard DD, "A multicenter study on objective and subjective benefits with a transcutaneous bone-anchored hearing aid device : first Nordic results", Eur Arch Otorhinolaryngol, 274, pp.3011~3019 (2017)
  12. Gawliczek T, "Speech understanding and sound localization with a new nonimplantable wearing option for Baha", Biomed Res Int, 2018, pp.5264124~ (2018)
  13. Verstraeten N, "Comparison of the audiologic results obtained with the bone-anchored hearing aid attached to the headband, the testband, and to the “snap”abutment", Otol Neurotol, 30, pp.70~75 (2009)
  14. Kim G, "Efficacy of boneanchored hearing aids in single-sided deafness : a systematic review", Otol Neurotol, 38, pp.473~483 (2017)
  15. 서영준, "Update on Bone-Conduction Auditory Brainstem Responses: A Review", Journal of Audiology & Otology, 22(2), pp.53~58 (2018), 대한청각학회
  16. Reinfeldt S, "New developments in bone-conduction hearing implants: a review", Med Devices(Auckl), 8, pp.79~93 (2015)
  17. Baker AR, "A retrospective review of temporal bone imaging with respect to bone-anchored hearing aid placement", Otol Neurotol, 38, pp.86~88 (2017)
  18. 장동수, "Baha Attract Implantation Using a Small Incision: Initial Report of Surgical Technique and Surveillance", Clinical and Experimental Otorhinolaryngology, 13(1), pp.15~22 (2020), 대한이비인후과학회[DOI]

관련 논문AI추천

이 간행물의 다른 논문 (7건)