Development of a real-time gamma camera for high radiation fields

2

조회

0

다운로드

0

인용

0

참고문헌

키워드

주제분류원자력공학

초록

In high radiation fields, gamma cameras suffer from pulse pile-up, resulting in poor energy resolution, count losses, and image distortion. To overcome this problem, various methods have been introduced to reduce the size of the aperture or pixel, reject the pile-up events, and correct the pile-up events, but these technologies have limitations in terms of mechanical design and real-time processing. The purpose of this study is to develop a realtime gamma camera to evaluate the radioactive contamination in high radiation fields. The gamma camera is composed of a pinhole collimator, NaI(Tl) scintillator, position sensitive photomultiplier (PSPMT), signal processing board, and data acquisition (DAQ). The pulse pile-up is corrected in real-time with a field programmable gate array (FPGA) using the start time correction (STC) method. The STC method corrects the amplitude of the pile-up event by correcting the time at the start point of the pile-up event. The performance of the gamma camera was evaluated using a high dose rate 137Cs source. For pulse pile-up ratios (PPRs) of 0.45 and 0.30, the energy resolution improved by 61.5 and 20.3%, respectively. In addition, the image artifacts in the 137Cs radioisotope image due to pile-up were reduced.

Abstract

In high radiation fields, gamma cameras suffer from pulse pile-up, resulting in poor energy resolution, count losses, and image distortion. To overcome this problem, various methods have been introduced to reduce the size of the aperture or pixel, reject the pile-up events, and correct the pile-up events, but these technologies have limitations in terms of mechanical design and real-time processing. The purpose of this study is to develop a realtime gamma camera to evaluate the radioactive contamination in high radiation fields. The gamma camera is composed of a pinhole collimator, NaI(Tl) scintillator, position sensitive photomultiplier (PSPMT), signal processing board, and data acquisition (DAQ). The pulse pile-up is corrected in real-time with a field programmable gate array (FPGA) using the start time correction (STC) method. The STC method corrects the amplitude of the pile-up event by correcting the time at the start point of the pile-up event. The performance of the gamma camera was evaluated using a high dose rate 137Cs source. For pulse pile-up ratios (PPRs) of 0.45 and 0.30, the energy resolution improved by 61.5 and 20.3%, respectively. In addition, the image artifacts in the 137Cs radioisotope image due to pile-up were reduced.

참고문헌 (26)

  1. International Atomic Energy Agency, "International Nuclear and Radiological Event Scale" (2008)
  2. Teppei J. Yasunari, "Cesium-137deposition and contamination of Japanese soils due to the Fukushima nuclear accident", Proc. Natl. Acad. Sci. U. S. A, 108(49), pp.19530~19534 (2011)
  3. Yuichiro Ueno, "Spectroscopic gamma camera for use in high dose environment", Nucl. Instrum. Methods Phys. Res, 822, pp.48~56 (2016)
  4. Gyuseong Cho, "Gamma-ray detectors for nuclear medical imaging instruments", Nucl. Med. Mol. Imag, 42(2), pp.88~97 (2008)
  5. Simon R. Cherry, "Physics in Nuclear Medicine" (2012), Elsevier
  6. Jerry L. Prince, "Medical Imaging Signals and Systems" (2016)
  7. E. Caroli, "Coded aperture imaging in X-and gamma-ray astronomy", Space Sci. Rev, 45(3), pp.349~403 (1987)
  8. Youngjun Shin, "Estimating the direction and distance of an unknown radiation source using RMC", J. Inst. Electron. Inf. Eng, 53(9), pp.1416~1423 (2016)
  9. Michał J. Cieslak, "Coded-aperture imaging systems: past, present and future development - a review", Radiat. Meas, 92, pp.59~71 (2016)
  10. Ji-Peng Zhang, "Prototype of an array SiPM-based scintillator Compton camera for radioactive materials detection", Radiat. Detect. Technol. Methods, 3(17), pp.1~12 (2019)
  11. Leonid L. Nkuba, "Review of the imaging performance and the current status of the cascade gamma-rays coincidence imagers", Tanzan. J. Eng. Technol, 41(4), pp.229~246 (2022)
  12. Glenn F. Knoll, "Radiation Detection and Measurement" (2010), John Wiley &Sons
  13. H. -S. Kim, "An asynchronous sampling-based 128 X 128 direct photoncounting X-ray image detector with multi-energy discrimination and high spatial resolution", IEEE J. Solid State Circ, 48(2), pp.541~558 (2013)
  14. R. Bellazzini, "Pixie III : a very large area photon-counting CMOS pixel ASIC for sharp X-ray spectral imaging", J. Instrum, 10, pp.1~8 (2015)
  15. Daehee Lee, "A new cross-detection method for improved energy-resolving photon counting under pulse pile-up", Nucl. Instrum. Methods Phys. Res. A, 867, pp.154~162 (2017)
  16. Kun Hu, "An FPGA-based pulse pile-up rejection technique for photon counting imaging detectors", Chin. Phys. Lett, 32(3), pp.1~4 (2015)
  17. Xiaodong Zhong, "A spectrometer with baseline correction and fast pulse pile-up rejection for prompt gamma neutron activation analysis Technology", Rev. Sci. Instrum, 89, pp.1~8 (2018)
  18. Hong Xu, "Counting-loss correction for X-ray spectroscopy using unit impulse pulse shaping", J. Synchrotron Radiat, 25(2), pp.505~513 (2018)
  19. Yi Liu, "Countingloss correction method based on dual-exponential impulse shaping", J. Synchrotron Radiat, 27, pp.1607~1613 (2020)
  20. P. A. B. Scoullar, "Real time pulse pile-up recovery in a high throughput digital pulse processor", AIP Conf. Proc, 1421, pp.270~277 (2011)
  21. X. Wen, "Study on a digital pulse processing algorithm based on templatematching for high-throughput spectroscopy", Nucl. Instrum. Methods Phys. Res. A, 784, pp.269~273 (2015)
  22. W. H. Wong, "Feasibility of a high-speed gamma-camera design using the high-yield-pileup-event-recovery method", J. Nucl. Med, 42, pp.624~632 (2001)
  23. J. Liu, "Real Time Digital Implementation of the High-Yield-Pileup-Event Recover (HYPER) Method", IEEE Nuclear Science Symposium Conference Record, pp.4230~4232 (2007)
  24. X. Wang, "Advantages of digitally sampling scintillation pulses in pileup processing in PET", IEEE Nuclear Science Symposium Conference Record, pp.498~506 (2012)
  25. M. -R. Mohammadian-Behbahani, "Pile-up correction algorithm based on successive integration for high count rate medical imaging and radiation spectroscopy", Nucl. Instrum. Methods Phys. Res. A, 897, pp.1~7 (2018)
  26. Minju Lee, "Pulse pileup correction method for gamma-ray spectroscopy in high radiation fields", Nucl. Eng. Technol, 52, pp.1029~1035 (2020)

관련 논문AI추천

Development and evaluation of a compact gamma camera for radiation monitoring권다은, 김장오, 백철하, 이승재, 이학재, 정경환, 한동희한국원자력학회 · Nuclear Engineering and Technology2023 · 제55권 제8호Development of simultaneous multi-channel data acquisition system for large-area Compton camera (LACC)구영모, 김재현, 김찬형, 엄태현, 이고은, 이준영, 이현수, 최세훈한국원자력학회 · Nuclear Engineering and Technology2023 · 제55권 제10호In-situ estimation of gamma-ray source location at a specific depth based on a coded-aperture gamma camera김덕우, 정만희한국원자력학회 · Nuclear Engineering and Technology2025 · 제57권 제7호Development of a diverging collimator for environmental radiation monitoring in the industrial fields권다은, 김장오, 백철하, 이승재, 이학재, 한동희한국원자력학회 · Nuclear Engineering and Technology2022 · 제54권 제12호Performance evaluation of Yonsei Single-photon Emission Computed Tomography (YSECT) for partial-defect inspection within PWR-type spent nuclear fuel곽성우, 김호직, 민철희, 백희균, 신정기, 정용현, 정윤수, 정희준, 최형주한국원자력학회 · Nuclear Engineering and Technology2024 · 제56권 제11호Transformable geometry for rapid in-situ gamma monitoring: Transformable in-situ gamma monitoring (TRIGAM) system장미, 김영학, 이민선, 임종명한국원자력학회 · Nuclear Engineering and Technology2025 · 제57권 제6호

이 간행물의 다른 논문 (39건)