초록

In Korea, all nuclear power plants (NPPs) participate in annual performance tests including in vivo measurements using the FastScan, a stand type whole body counter (WBC), manufactured by Canberra. In 2018, all Korean NPPs satisfied the testing criterion, the root mean square error (RMSE)  0.25, for the whole body configuration, but three NPPs which participated in an additional lung configuration test in the fission and activation product category did not meet the criterion. Due to the low resolution of the FastScan NaI(Tl) detectors, the conventional peak analysis (PA) method of the FastScan did not show sufficient performance to meet the criterion in the presence of interfering radioisotopes (RIs), 134Cs and 137Cs. In this study, we developed an artificial neural network (ANN) to improve the performance of the FastScan in the lung configuration. All of the RMSE values derived by the ANN satisfied the criterion, even though the photopeaks of 134Cs and 137Cs interfered with those of the analytes or the analyte photopeaks were located in a low-energy region below 300 keV. Since the ANN performed better than the PA method, it would be expected to be a promising approach to improve the accuracy and precision of in vivo FastScan measurement for the lung configuration

Abstract

In Korea, all nuclear power plants (NPPs) participate in annual performance tests including in vivo measurements using the FastScan, a stand type whole body counter (WBC), manufactured by Canberra. In 2018, all Korean NPPs satisfied the testing criterion, the root mean square error (RMSE)  0.25, for the whole body configuration, but three NPPs which participated in an additional lung configuration test in the fission and activation product category did not meet the criterion. Due to the low resolution of the FastScan NaI(Tl) detectors, the conventional peak analysis (PA) method of the FastScan did not show sufficient performance to meet the criterion in the presence of interfering radioisotopes (RIs), 134Cs and 137Cs. In this study, we developed an artificial neural network (ANN) to improve the performance of the FastScan in the lung configuration. All of the RMSE values derived by the ANN satisfied the criterion, even though the photopeaks of 134Cs and 137Cs interfered with those of the analytes or the analyte photopeaks were located in a low-energy region below 300 keV. Since the ANN performed better than the PA method, it would be expected to be a promising approach to improve the accuracy and precision of in vivo FastScan measurement for the lung configuration

참고문헌 (13)

  1. "Performance Criteria for Radiobioassay, ANSI/HPS N13.30-2011" (2011)[URL]
  2. vitro Measurement Equipment, "Improvement of the Performance Test Method for In-Vivo", pp.2019, Jung Kwon Son
  3. M. Kamuda, "Automated isotope identification algorithm using artificial neural networks", IEEE Trans. Nucl. Sci, 64, pp.1858~1864 (2017)[DOI]
  4. M. Kamuda, "An automated isotope identification and quantification algorithm for isotope mixtures in low-resolution gamma-ray spectra", Radiat. Phys. Chem, 155, pp.281~286 (2019)[DOI]
  5. J. Kim, "Quantitative analysis of NaI(Tl) gamma-ray spectrometry using an artificial neural network", Nucl. Instruments Methods Phys. Res. Sect. A Accel. Spectrometers, Detect. Assoc. Equip, 944, pp.162549~ (2019)[DOI]
  6. P. Olmos, "A new approach to automatic radiation spectrum analysis", IEEE Trans. Nucl. Sci, 38, pp.971~975 (1991)[DOI]
  7. "Mirion Technol" (2019)[URL]
  8. O. R. Perry, "Calibration of the Accuscan-II In-Vivo System for I-125 Thyroid Counting", Appendix J (2011)
  9. F. Chollet, "Deep Learning with Python", pp.8~97 (2017), Manning Publications Co
  10. S. Alla, "Beginning Anomaly Detection Using Python-Based Deep Learning" (2019), Apress
  11. Genie, "Spectroscopy Software Customization Tools, 9233653J V3.4", Canberra Inc, ~ (2000)
  12. K. Saito, "Monte Carlo calculation of NaI(Tl)detector response functions for low energy gamma rays", Nucl. Instrum. Methods Phys. Res. A, 226, pp.449~454 (1984)[DOI]
  13. J. Plagnard, "Study of scattering effects in low-energy gamma-ray spectrometry", Appl. Radiat. Isot, 66, pp.769~773 (2008)[DOI]

관련 논문AI추천

Bed-type과 Stand-type 상용 전신계수기(Whole Body Counter)의 성능 비교권태은, 김봉기, 김종민, 박민석, 이상경, 이준호, 정규환, 하위호대한방사선과학회 · Journal of Radiological Science and Technology2018 · 제41권 제5호Improvement of internal exposure assessments of the inhalation of fuel-type hot particles during long-term outages김형진, 조문형한국원자력학회 · Nuclear Engineering and Technology2024 · 제56권 제9호Radionuclide identification method for NaI low-count gamma-ray spectra using artificial neural networkHaijun Fan, Hui Zhao, Jinglun Li, Kun Zhang, Shanqiang Wang, Sheng Qi, Xianyun Ai, Ye Chen한국원자력학회 · Nuclear Engineering and Technology2022 · 제54권 제1호방사능재난 시 작업자 선량평가에 신속 활용 가능한 AI 기반 내부피폭선량평가체계 개발김송현, 박민석, 유재룡, 이장희(사)한국방사선산업학회 · 방사선산업학회지2021 · 제15권 제4호전신계측기를 이용한 인체 방사능측정 및 방사능오염 구분에 대한 고찰공태영, 김희근, 이경진(사)한국방사선산업학회 · 방사선산업학회지2018 · 제12권 제2호Effect of non-uniform distribution of radioactivity in human body on measurements using HPGe detector based scanning bed whole body counterAnilkumar S., Chaudhury Probal, Mishra Lokpati, Nadar M.Y., Prathibha P., Sankhla Rajesh, Sawant P.D., Singh I.S.한국원자력학회 · Nuclear Engineering and Technology2025 · 제57권 제9호

이 간행물의 다른 논문 (19건)